Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer.

Martin Meyerspeer1, Ewald Unger2, Ivo Rausch3, Alejandra Valladares2, Lalith Kumar Shiyam Sundar2, Thomas Beyer2, Marcus Hacker4.   

Abstract

BACKGROUND: PET/MRI phantom studies are challenged by the need of phantom-specific attenuation templates to account for attenuation properties of the phantom material. We present a PET/MRI phantom built from MRI-visible material for which attenuation correction (AC) can be performed using the standard MRI-based AC.
METHODS: A water-fillable phantom was 3D-printed with a commercially available MRI-visible polymer. The phantom had a cylindrical shape and the fillable compartment consisted of a homogeneous region and a region containing solid rods of different diameters. The phantom was filled with a solution of water and [18F]FDG. A 30 min PET/MRI acquisition including the standard Dixon-based MR-AC method was performed. In addition, a CT scan of the phantom was acquired on a PET/CT system. From the Dixon in-phase, opposed-phase and fat images, a phantom-specific AC map (Phantom MR-AC) was produced by separating the phantom material from the water compartment using a thresholding-based method and assigning fixed attenuation coefficients to the individual compartments. The PET data was reconstructed using the Phantom MR-AC, the original Dixon MR-AC, and an MR-AC just containing the water compartment (NoWall-AC) to estimate the error of ignoring the phantom walls. CT-based AC was employed as the reference standard. Average %-differences in measured activity between the CT corrected PET and the PET corrected with the other AC methods were calculated.
RESULTS: The phantom housing and the liquid compartment were both visible and distinguishable from each other in the Dixon images and allowed the segmentation of a phantom-specific MR-based AC. Compared to the CT-AC PET, average differences in measured activity in the whole water compartment in the phantom of -0.3%, 9.4%, and -24.1% were found for Dixon phantom MR-AC, MR-AC, and NoWall-AC based PET, respectively. Average differences near the phantom wall in the homogeneous region were -0.3%, 6.6%, and -34.3%, respectively. Around the rods, activity differed from the CT-AC PET by 0.7%, 8.9%, and -45.5%, respectively.
CONCLUSION: The presented phantom material is visible using standard MR sequences, and thus, supports the use of standard, phantom-independent MR measurements for MR-AC in PET/MRI phantom studies.

Entities:  

Keywords:  MR visible polymer; PET/MRI; Phantom attenuation correction

Year:  2021        PMID: 33599876     DOI: 10.1186/s40658-021-00364-9

Source DB:  PubMed          Journal:  EJNMMI Phys        ISSN: 2197-7364


  31 in total

Review 1.  Advances in PET/MR instrumentation and image reconstruction.

Authors:  Jorge Cabello; Sibylle I Ziegler
Journal:  Br J Radiol       Date:  2016-07-22       Impact factor: 3.039

2.  Developing and characterizing MR/CT-visible materials used in QA phantoms for MRgRT systems.

Authors:  Angela Steinmann; R Jason Stafford; Gabriel Sawakuchi; Zhifei Wen; Laurence Court; Clifton D Fuller; David Followill
Journal:  Med Phys       Date:  2017-12-21       Impact factor: 4.071

Review 3.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26

4.  Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data.

Authors:  Axel Martinez-Möller; Michael Souvatzoglou; Gaspar Delso; Ralph A Bundschuh; Christophe Chefd'hotel; Sibylle I Ziegler; Nassir Navab; Markus Schwaiger; Stephan G Nekolla
Journal:  J Nucl Med       Date:  2009-03-16       Impact factor: 10.057

5.  MR-based field-of-view extension in MR/PET: B0 homogenization using gradient enhancement (HUGE).

Authors:  Jan O Blumhagen; Ralf Ladebeck; Matthias Fenchel; Klaus Scheffler
Journal:  Magn Reson Med       Date:  2012-11-30       Impact factor: 4.668

6.  Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility.

Authors:  Sune H Keller; Björn Jakoby; Susanne Svalling; Andreas Kjaer; Liselotte Højgaard; Thomas L Klausen
Journal:  EJNMMI Phys       Date:  2016-07-07

7.  A multi-centre evaluation of eleven clinically feasible brain PET/MRI attenuation correction techniques using a large cohort of patients.

Authors:  Claes N Ladefoged; Ian Law; Udunna Anazodo; Keith St Lawrence; David Izquierdo-Garcia; Ciprian Catana; Ninon Burgos; M Jorge Cardoso; Sebastien Ourselin; Brian Hutton; Inés Mérida; Nicolas Costes; Alexander Hammers; Didier Benoit; Søren Holm; Meher Juttukonda; Hongyu An; Jorge Cabello; Mathias Lukas; Stephan Nekolla; Sibylle Ziegler; Matthias Fenchel; Bjoern Jakoby; Michael E Casey; Tammie Benzinger; Liselotte Højgaard; Adam E Hansen; Flemming L Andersen
Journal:  Neuroimage       Date:  2016-12-14       Impact factor: 6.556

8.  Combined PET/MRI: Global Warming-Summary Report of the 6th International Workshop on PET/MRI, March 27-29, 2017, Tübingen, Germany.

Authors:  D L Bailey; B J Pichler; B Gückel; G Antoch; H Barthel; Z M Bhujwalla; S Biskup; S Biswal; M Bitzer; R Boellaard; R F Braren; C Brendle; K Brindle; A Chiti; C la Fougère; R Gillies; V Goh; M Goyen; M Hacker; L Heukamp; G M Knudsen; A M Krackhardt; I Law; J C Morris; K Nikolaou; J Nuyts; A A Ordonez; K Pantel; H H Quick; K Riklund; O Sabri; B Sattler; E G C Troost; M Zaiss; L Zender; Thomas Beyer
Journal:  Mol Imaging Biol       Date:  2018-02       Impact factor: 3.488

9.  SUV-quantification of physiological lung tissue in an integrated PET/MR-system: Impact of lung density and bone tissue.

Authors:  Ferdinand Seith; Holger Schmidt; Sergios Gatidis; Ilja Bezrukov; Christina Schraml; Christina Pfannenberg; Christian la Fougère; Konstantin Nikolaou; Nina Schwenzer
Journal:  PLoS One       Date:  2017-05-31       Impact factor: 3.240

View more
  2 in total

Review 1.  Influences on PET Quantification and Interpretation.

Authors:  Julian M M Rogasch; Frank Hofheinz; Lutz van Heek; Conrad-Amadeus Voltin; Ronald Boellaard; Carsten Kobe
Journal:  Diagnostics (Basel)       Date:  2022-02-10

2.  A Path to Qualification of PET/MRI Scanners for Multicenter Brain Imaging Studies: Evaluation of MRI-Based Attenuation Correction Methods Using a Patient Phantom.

Authors:  Ciprian Catana; Richard Laforest; Hongyu An; Fernando Boada; Tuoyu Cao; David Faul; Bjoern Jakoby; Floris P Jansen; Bradley J Kemp; Paul E Kinahan; Peder Larson; Michael A Levine; Piotr Maniawski; Osama Mawlawi; Jonathan E McConathy; Alan B McMillan; Julie C Price; Abhejit Rajagopal; John Sunderland; Patrick Veit-Haibach; Kristen A Wangerin; Chunwei Ying; Thomas A Hope
Journal:  J Nucl Med       Date:  2021-07-22       Impact factor: 11.082

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.