Computerized method for evaluating diagnostic image quality of calcified plaque images in cardiac CT: validation on a physical dynamic cardiac phantom.

Martin King1, Zachary Rodgers, Maryellen L Giger, Dianna M E Bardo, Amit R Patel.   

Abstract

PURPOSE: In cardiac computed tomography (CT), important clinical indices, such as the coronary calcium score and the percentage of coronary artery stenosis, are often adversely affected by motion artifacts. As a result, the expert observer must decide whether or not to use these indices during image interpretation. Computerized methods potentially can be used to assist in these decisions. In a previous study, an artificial neural network (ANN) regression model provided assessability (image quality) indices of calcified plaque images from the software NCAT phantom that were highly agreeable with those provided by expert observers. The method predicted assessability indices based on computer-extracted features of the plaque. In the current study, the ANN-predicted assessability indices were used to identify calcified plaque images with diagnostic calcium scores (based on mass) from a physical dynamic cardiac phantom. The basic assumption was that better quality images were associated with more accurate calcium scores.
METHODS: A 64-channel CT scanner was used to obtain 500 calcified plaque images from a physical dynamic cardiac phantom at different heart rates, cardiac phases, and plaque locations. Two expert observers independently provided separate sets of assessability indices for each of these images. Separate sets of ANN-predicted assessability indices tailored to each observer were then generated within the framework of a bootstrap resampling scheme. For each resampling iteration, the absolute calcium score error between the calcium scores of the motion-contaminated plaque image and its corresponding stationary image served as the ground truth in terms of indicating images with diagnostic calcium scores. The performances of the ANN-predicted and observer-assigned indices in identifying images with diagnostic calcium scores were then evaluated using ROC analysis.
RESULTS: Assessability indices provided by the first observer and the corresponding ANN performed similarly (AUC(OBS1) = 0.80 [0.73, 0.86] vs AUC(ANN1) = 0.88 [0.82, 0.92]) as that of the second observer and the corresponding ANN (AUC(OBS2) = 0.87 [0.83,0.91] vs. AUC(ANN2) = 0.90 [0.85, 0.94]). Moreover, the ANN-predicted indices were generated in a fraction of the time required to obtain the observer-assigned indices.
CONCLUSIONS: ANN-predicted assessability indices performed similar to observer-assigned assessability indices in identifying images with diagnostic calcium scores from the physical dynamic cardiac phantom. The results of this study demonstrate the potential of using computerized methods for identifying images with diagnostic clinical indices in cardiac CT images.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21158289      PMCID: PMC2973992          DOI: 10.1118/1.3495684

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  31 in total

1.  Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed.

Authors:  Bernd J Wintersperger; Konstantin Nikolaou; Franz von Ziegler; Thorsten Johnson; Carsten Rist; Alexander Leber; Thomas Flohr; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Invest Radiol       Date:  2006-05       Impact factor: 6.016

2.  Motion compensation for interventional navigation on 3D static roadmaps based on an affine model and gating.

Authors:  Holger Timinger; Sascha Krueger; Joern Borgert; Ruediger Grewer
Journal:  Phys Med Biol       Date:  2004-03-07       Impact factor: 3.609

3.  Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time.

Authors:  Xavier L Hamoir; Thomas Flohr; Vincent Hamoir; Laurent Labaki; Jean-Yves Tricquet; Alain Duhamel; Jacques Kirsch
Journal:  Eur Radiol       Date:  2004-11-24       Impact factor: 5.315

4.  Dual-source CT cardiac imaging: initial experience.

Authors:  Thorsten R C Johnson; Konstantin Nikolaou; Bernd J Wintersperger; Alexander W Leber; Franz von Ziegler; Carsten Rist; Sonja Buhmann; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2006-05-13       Impact factor: 5.315

5.  Computerized assessment of motion-contaminated calcified plaques in cardiac multidetector CT.

Authors:  Martin King; Maryellen L Giger; Kenji Suzuki; Dianna M E Bardo; Brent Greenberg; Li Lan; Xiaochuan Pan
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

6.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  The impact of motion artifacts on the reproducibility of repeated coronary artery calcium measurements.

Authors:  Jun Horiguchi; Hiroshi Fukuda; Hideya Yamamoto; Nobuhiko Hirai; Farzana Alam; Hideaki Kakizawa; Masashi Hieda; Toshihiro Tachikake; Kazushi Marukawa; Katsuhide Ito
Journal:  Eur Radiol       Date:  2006-05-23       Impact factor: 5.315

9.  Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction.

Authors:  Dieter Ropers; Ulrich Baum; Karsten Pohle; Katharina Anders; Stefan Ulzheimer; Bernd Ohnesorge; Christian Schlundt; Werner Bautz; Werner G Daniel; Stephan Achenbach
Journal:  Circulation       Date:  2003-02-11       Impact factor: 29.690

10.  Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study.

Authors:  Jaap M Groen; Marcel J W Greuter; R Vliegenthart; C Suess; B Schmidt; F Zijlstra; M Oudkerk
Journal:  Int J Cardiovasc Imaging       Date:  2007-11-23       Impact factor: 2.357

View more
  2 in total

1.  Radial differential interior tomography and its image reconstruction with differentiated backprojection and projection onto convex sets.

Authors:  Shaojie Tang; Xiangyang Tang
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

2.  Diagnostic accuracy of coronary CT angiography combined with dual-energy myocardial perfusion imaging for detection of myocardial infarction.

Authors:  Ruijuan Han; Kai Sun; Bin Lu; Ruiping Zhao; Kuncheng Li; Xinchun Yang
Journal:  Exp Ther Med       Date:  2017-05-22       Impact factor: 2.447

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.