Evaluating performance of a pixel array semiconductor SPECT system for small animal imaging.

Naoki Kubo1, Songji Zhao, Yutaka Fujiki, Akiyoshi Kinda, Nobutoku Motomura, Chietsugu Katoh, Tohru Shiga, Hidekazu Kawashima, Yuji Kuge, Nagara Tamaki.   

Abstract

OBJECTIVES: Small animal imaging has recently been focused on basic nuclear medicine. We have designed and built a small animal SPECT imaging system using a semiconductor camera and a newly designed collimator. We assess the performance of this system for small object imaging.
METHODS: We employed an MGC 1500 (Acrorad Co.) camera including a CdTe semiconductor. The pixel size was 1.4 mm/pixel. We designed and produced a parallel-hole collimator with 20-mm hole length. Our SPECT system consisted of a semiconductor camera with the subject holder set on an electric rotating stage controlled by a computer. We compared this system with a conventional small animal SPECT system comprising a SPECT-2000H scanner with four Anger type cameras and pinhole collimators. The count rate linearity for estimation of the scatter was evaluated for a pie-chart phantom containing different concentrations of 99mTc. We measured the FWHM of the 99mTc SPECT line source along with scatter. The system volume sensitivity was examined using a flood source phantom which was 35 mm long with a 32-mm inside diameter. Additionally, an in vivo myocardial perfusion SPECT study was performed with a rat.
RESULTS: With regards to energy resolution, the semiconductor camera (5.6%) was superior to the conventional Anger type camera (9.8%). In the count rate linearity evaluation, the regression lines of the SPECT values were y = 0.019x + 0.031 (r2 = 0.999) for our system and y = 0.018x + 0.060 (r2 = 0.997) for the conventional system. Thus, the scatter count using the semiconductor camera was less than that using the conventional camera. FWHMs of our system and the conventional system were 2.9 +/- 0.1 and 2.0 +/- 0.1 mm, respectively. Moreover, the system volume sensitivity of our system [0.51 kcps/(MBq/ ml)/cm] was superior to that of the conventional system [0.44 kcps/(MBq/ml)/cm]. Our system provided clear images of the rat myocardium, sufficient for practical use in small animal imaging.
CONCLUSIONS: Our SPECT system, utilizing a semiconductor camera, permits high quantitative analysis by virtue of its low scatter radiation and high sensitivity. Therefore, this system may contribute to molecular imaging of small animals and basic medical research.

Entities:  

Mesh:

Year:  2005        PMID: 16363631     DOI: 10.1007/bf02985059

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  5 in total

1.  Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging.

Authors:  Keiichi Magota; Naoki Kubo; Yuji Kuge; Ken-Ichi Nishijima; Songji Zhao; Nagara Tamaki
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-12-10       Impact factor: 9.236

2.  Hybrid Pixel-Waveform (HPWF) Enabled CdTe Detectors for Small Animal Gamma-Ray Imaging Applications.

Authors:  A Groll; K Kim; H Bhatia; J C Zhang; J H Wang; Z M Shen; L Cai; J Dutta; Q Li; L J Meng
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2016-11-01

3.  Non-invasive near infrared fluorescence imaging of CdHgTe quantum dots in mouse model.

Authors:  Haiyan Chen; Yunqing Wang; Jing Xu; Jinzi Ji; Jian Zhang; Yuzhu Hu; Yueqing Gu
Journal:  J Fluoresc       Date:  2008-01-05       Impact factor: 2.217

Review 4.  Cadmium Telluride Semiconductor Detector for Improved Spatial and Energy Resolution Radioisotopic Imaging.

Authors:  Samira Abbaspour; Babak Mahmoudian; Jalil Pirayesh Islamian
Journal:  World J Nucl Med       Date:  2017 Apr-Jun

5.  A novel iterative modified bicubic interpolation method enables high-contrast and high-resolution image generation for F-18 FDG-PET.

Authors:  Atsutaka Okizaki; Michihiro Nakayama; Kaori Nakajima; Koji Takahashi
Journal:  Medicine (Baltimore)       Date:  2017-12       Impact factor: 1.889

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.