Physical performance evaluation of the Toshiba GCA-9300A triple-headed system.

K Kouris1, G A Clarke, P H Jarritt, C E Townsend, S N Thomas.   

Abstract

The physical performance of the Toshiba GCA-9300A triple-headed SPECT system has been assessed. Using a water-filled cylinder containing 99mTc, the tomographic volume sensitivity was 33.8 and 34.8 kcps/(MBq/ml)/cm for the high-resolution, parallel-hole (HR-PH) collimator and the super high-resolution, lead fanbeam (SHR-FB) collimator, respectively, excluding the rotation time(s) during scanning when data are not acquired. The tomographic spatial resolution at the center, in air, with 132 mm radius of rotation was 10.2 and 7.8 mm FWHM with the HR-PH and SHR-FB collimators, respectively; in water it was 11.0 and 7.8 mm. Reconstructed relative activity concentrations were accurate for both collimator sets if attenuation correction was used. With the SHR-FB collimators, the average peak-to-valley ratio of five-line sources in water improved significantly when an asymmetric energy window was used. Using the three-dimensional Hoffman brain phantom which simulates a 4:1 grey matter-to-white matter ratio and the usual choice of acquisition and processing parameters for brain studies, the reconstructed grey matter-to-white matter ratio was only about 1.7 for total counts typically acquired in a 99mTc-HMPAO study (4.0 M counts) and only 2.3 for 40 M counts. There was a qualitative improvement with an asymmetric energy window.

Entities:  

Mesh:

Year:  1993        PMID: 8410298

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  5 in total

1.  Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging.

Authors:  Keiichi Magota; Naoki Kubo; Yuji Kuge; Ken-Ichi Nishijima; Songji Zhao; Nagara Tamaki
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-12-10       Impact factor: 9.236

2.  Evaluation of SPET quantification of simultaneous emission and transmission imaging of the brain using a multidetector SPET system with the TEW scatter compensation method and fan-beam collimation.

Authors:  T Ichihara; N Motomura; K Ogawa; H Hasegawa; J Hashimoto; A Kubo
Journal:  Eur J Nucl Med       Date:  1996-10

3.  Regional ventilatory evaluation using dynamic SPET imaging of xenon-133 washout in obstructive lung disease: an initial study.

Authors:  K Suga; K Nishigauchi; N Kume; S Koike; K Takano; N Matsunaga
Journal:  Eur J Nucl Med       Date:  1995-03

4.  Evaluation of a variable-aperture full-ring SPECT system using large-area pixelated CZT modules: A simulation study for brain SPECT applications.

Authors:  Yoonsuk Huh; Jaewon Yang; Odera U Dim; Yonggang Cui; Weijie Tao; Qiu Huang; Grant T Gullberg; Youngho Seo
Journal:  Med Phys       Date:  2021-03-30       Impact factor: 4.071

5.  Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators.

Authors:  Keishin Morita; Akira Maebatake; Rina Iwasaki; Yuki Shiotsuki; Kazuhiko Himuro; Shingo Baba; Masayuki Sasaki
Journal:  Asia Ocean J Nucl Med Biol       Date:  2018
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.