The accuracy of frameless stereotactic intracranial radiosurgery.

Wilko F A R Verbakel1, Frank J Lagerwaard, Arjan J E Verduin, Stan Heukelom, Ben J Slotman, Johan P Cuijpers.   

Abstract

PURPOSE: To determine the accuracy of frameless stereotactic radiosurgery using the BrainLAB ExacTrac system and robotic couch by measuring the individual contributions such as the accuracy of the imaging and couch correction system, the linkage between this system and the linac isocenter and the possible intrafraction motion of the patient in the frameless mask.
MATERIALS AND METHODS: An Alderson head phantom with hidden marker was randomly positioned 31 times. Automated 6D couch shifts were performed according to ExacTrac and the deviation with respect to the linac isocenter was measured using the hidden marker. ExacTrac-based set-up was performed for 46 patients undergoing hypofractionated stereotactic radiotherapy for 135 fractions, followed by verification X-rays. Forty-three of these patients received post-treatment X-ray verification for 79 fractions to determine the intrafraction motion.
RESULTS: The hidden target test revealed a systematic error of 1.5 mm in one direction, which was corrected after replacement of the system calibration phantom. The accuracy of the ExacTrac positioning is approximately 0.3 mm in each direction, 1 standard deviation. The intrafraction motion was 0.35±0.21 mm, maximum 1.15 mm.
CONCLUSION: Intrafraction motion in the BrainLAB frameless mask is very small. Users are strongly advised to perform an independent verification of the ExacTrac isocenter in order to avoid systematic deviations.
Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21047692     DOI: 10.1016/j.radonc.2010.06.012

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  25 in total

1.  Accuracy of positional correction for the floor-mounted kV X-ray IGRT system in angled couch positions.

Authors:  Satoshi Koubuchi; Toru Takakura; Mitsuhiro Nakamura; Takashi Mizowaki; Manabu Nakata; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2014-06-13

2.  A Systematic Analysis of Errors in Target Localization and Treatment Delivery for Stereotactic Radiosurgery Using 2D/3D Image Registration.

Authors:  Hao Xu; Stephen Brown; Indrin J Chetty; Ning Wen
Journal:  Technol Cancer Res Treat       Date:  2016-08-31

3.  Intrafractional positioning accuracy during stereotactic brain radiotherapy with a modified thermoplastic mask system, on-board cone-beam CT, and a six-degrees-of-freedom couch.

Authors:  Naohisa Mizukami; Kiyoshi Yoda
Journal:  J Radiosurg SBRT       Date:  2016

4.  Evaluation of additional treatment margins for compensating rotational random errors in linac-based single-isocenter stereotactic radiotherapy for multiple brain metastases.

Authors:  Yasuhisa Yoshida; Maki Soyama; Rieko Azumi
Journal:  J Radiosurg SBRT       Date:  2022

5.  The Continuous Assessment of Cranial Motion in Thermoplastic Masks During CyberKnife Radiosurgery for Trigeminal Neuralgia.

Authors:  Tewfik J Bichay; Alan Mayville
Journal:  Cureus       Date:  2016-05-12

6.  Clinical Assessment of 2D/3D Registration Accuracy in 4 Major Anatomic Sites Using On-Board 2D Kilovoltage Images for 6D Patient Setup.

Authors:  Guang Li; T Jonathan Yang; Hugo Furtado; Wolfgang Birkfellner; Åse Ballangrud; Simon N Powell; James Mechalakos
Journal:  Technol Cancer Res Treat       Date:  2014-09-15

7.  Spatial variations of multiple off-axial targets for a single isocenter SRS treatment in Novalis Tx linac system.

Authors:  Sangroh Kim; Tzu-Chi Tseng; Andrew Morrow
Journal:  J Radiosurg SBRT       Date:  2015

8.  Targeting accuracy at couch kick for a frameless image guided radiosurgery system.

Authors:  Yimei Huang; Bo Zhao; Joshua Kim; Ning Wen; Indrin J Chetty; Salim Siddiqui
Journal:  J Radiosurg SBRT       Date:  2018

9.  Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes.

Authors:  Giuseppe Minniti; Claudia Scaringi; Enrico Clarke; Maurizio Valeriani; Mattia Osti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2011-11-16       Impact factor: 3.481

10.  Technical accuracy of optical and the electromagnetic tracking systems.

Authors:  Tapani Koivukangas; Jani Pa Katisko; John P Koivukangas
Journal:  Springerplus       Date:  2013-03-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.