A Systematic Analysis of Errors in Target Localization and Treatment Delivery for Stereotactic Radiosurgery Using 2D/3D Image Registration.

Hao Xu1, Stephen Brown2, Indrin J Chetty2, Ning Wen2.   

Abstract

PURPOSE: To determine the localization uncertainties associated with 2-dimensional/3-dimensional image registration in comparison to 3-dimensional/3-dimensional image registration in 6 dimensions on a Varian Edge Linac under various imaging conditions.
METHODS: The systematic errors in 6 dimensions were assessed by comparing automatic 2-dimensional/3-dimensional (kV/MV vs computed tomography) with 3-dimensional/3-dimensional (cone beam computed tomography vs computed tomography) image registrations under various conditions encountered in clinical applications. The 2-dimensional/3-dimensional image registration uncertainties for 88 patients with different treatment sites including intracranial and extracranial were evaluated by statistically analyzing 2-dimensional/3-dimensional pretreatment verification shifts of 192 fractions in stereotactic radiosurgery and stereotactic body radiotherapy.
RESULTS: The systematic errors of 2-dimensional/3-dimensional image registration using kV-kV, MV-kV, and MV-MV image pairs were within 0.3 mm and 0.3° for the translational and rotational directions within a 95% confidence interval. No significant difference ( P > .05) in target localization was observed with various computed tomography slice thicknesses (0.8, 1, 2, and 3 mm). Two-dimensional/3-dimensional registration had the best accuracy when pattern intensity and content filter were used. For intracranial sites, means ± standard deviations of translational errors were -0.20 ± 0.70 mm, 0.04 ± 0.50 mm, and 0.10 ± 0.40 mm for the longitudinal, lateral, and vertical directions, respectively. For extracranial sites, means ± standard deviations of translational errors were -0.04 ± 1.00 mm, 0.2 ± 1.0 mm, and 0.1 ± 1.0 mm for the longitudinal, lateral, and vertical directions, respectively. Two-dimensional/3-dimensional image registration for intracranial and extracranial sites had comparable systematic errors that were approximately 0.2 mm in the translational direction and 0.08° in the rotational direction.
CONCLUSION: The standard 2-dimensional/3-dimensional image registration tool available on the Varian Edge radiosurgery device, a state-of-the-art system, is helpful for robust and accurate target positioning for image-guided stereotactic radiosurgery.

Entities:  

Keywords:  2D/3D; 3D/3D; Edge; image guided; image registration; systematic error

Mesh:

Year:  2016        PMID: 27582369      PMCID: PMC5616047          DOI: 10.1177/1533034616664425

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  29 in total

1.  The importance of computed tomography slice thickness in radiographic patient positioning for radiosurgery.

Authors:  M J Murphy
Journal:  Med Phys       Date:  1999-02       Impact factor: 4.071

2.  Setup error in radiotherapy: on-line correction using electronic kilovoltage and megavoltage radiographs.

Authors:  L Pisani; D Lockman; D Jaffray; D Yan; A Martinez; J Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-06-01       Impact factor: 7.038

Review 3.  A review of 3D/2D registration methods for image-guided interventions.

Authors:  P Markelj; D Tomaževič; B Likar; F Pernuš
Journal:  Med Image Anal       Date:  2010-04-13       Impact factor: 8.545

4.  A study of the accuracy of cyberknife spinal radiosurgery using skeletal structure tracking.

Authors:  Anthony K Ho; Dongshan Fu; Cristian Cotrutz; Steven L Hancock; Steven D Chang; Iris C Gibbs; Calvin R Maurer; John R Adler
Journal:  Neurosurgery       Date:  2007-02       Impact factor: 4.654

Review 5.  Advances in image-guided radiation therapy.

Authors:  Laura A Dawson; David A Jaffray
Journal:  J Clin Oncol       Date:  2007-03-10       Impact factor: 44.544

6.  Evaluation of similarity measures for use in the intensity-based rigid 2D-3D registration for patient positioning in radiotherapy.

Authors:  Jian Wu; Minho Kim; Jorg Peters; Heeteak Chung; Sanjiv S Samant
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

7.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

8.  Multimodality image registration by maximization of mutual information.

Authors:  F Maes; A Collignon; D Vandermeulen; G Marchal; P Suetens
Journal:  IEEE Trans Med Imaging       Date:  1997-04       Impact factor: 10.048

9.  Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice.

Authors:  Matthias Guckenberger; Juergen Meyer; Juergen Wilbert; Kurt Baier; Otto Sauer; Michael Flentje
Journal:  Strahlenther Onkol       Date:  2007-06       Impact factor: 3.621

10.  Characteristics of a novel treatment system for linear accelerator-based stereotactic radiosurgery.

Authors:  Ning Wen; Haisen Li; Kwang Song; Karen Chin-Snyder; Yujiao Qin; Jinkoo Kim; Maria Bellon; Misbah Gulam; Stephen Gardner; Anthony Doemer; Suneetha Devpura; James Gordon; Indrin Chetty; Farzan Siddiqui; Munther Ajlouni; Robert Pompa; Zane Hammoud; Michael Simoff; Steven Kalkanis; Benjamin Movsas; M Salim Siddiqui
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

View more
  2 in total

1.  Intra-fractional patient motion when using the Qfix Encompass immobilization system during HyperArc treatment of patients with brain metastases.

Authors:  Riho Komiyama; Shingo Ohira; Hikari Ueda; Naoyuki Kanayama; Akira Masaoka; Masaru Isono; Yoshihiro Ueda; Masayoshi Miyazaki; Teruki Teshima
Journal:  J Appl Clin Med Phys       Date:  2021-03-03       Impact factor: 2.102

2.  Orthogonal image pairs coupled with OSMS for noncoplanar beam angle, intracranial, single-isocenter, SRS treatments with multiple targets on the Varian Edge radiosurgery system.

Authors:  Jasmine A Oliver; Patrick Kelly; Sanford L Meeks; Twyla R Willoughby; Amish P Shah
Journal:  Adv Radiat Oncol       Date:  2017-04-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.