Literature DB >> 27330875

The Continuous Assessment of Cranial Motion in Thermoplastic Masks During CyberKnife Radiosurgery for Trigeminal Neuralgia.

Tewfik J Bichay1, Alan Mayville1.   

Abstract

Stereotactic radiosurgery (SRS) treatment is characterized by high doses per fraction and extremely steep dose gradients. This requires a great degree of accurate localization to the appropriate treatment position, and continuous immobilization during the treatment session. In the case of Trigeminal Neuralgia (TGN) treatment this is especially true as the very small target volume makes positional accuracy critical. In this study we carried out a quantitative analysis of patient motion during the full treatment fraction within a radiosurgery immobilization mask system. Patient cranial movement was assessed by using the image guidance stereo x-ray cameras on a CyberKnife (CK) M6 robotic radiosurgery system (Accuray, Sunnyvale, CA). A total of five patients received treatments for either right or left TGN. The duration of treatment varied from 24-64 minutes. Orthogonal images were taken every 15 seconds during the treatment to assess patient movement. Approximately 60 stereo images were taken per patient and a total of 560 images were analyzed in this study. The mean absolute movement in each of longitudinal, lateral or vertical directions was approximately 0.3 mm for the duration of the treatment; however, on occasion much greater movement was observed during a fraction. The maximum displacement was in the longitudinal direction and reached 2.4 mm compared to the initial setup. Images taken at the end of the treatment session showed that the patients typically return to a position closer to the original setup position than the maximum excursion that occurred. This data suggests that although this mask system appears stable during much of the treatment session; for some patients there may be momentary patient movements that take place. Frequent imaging and correction can help mitigate the effect of this movement. It is important to understand the limitations of non-invasive mask systems when used for very high precision treatment.

Entities:  

Keywords:  radiosurgery; srs; stereotactic radiosurgery; thermoplastic mask; treatment motion; trigeminal neuralgia

Year:  2016        PMID: 27330875      PMCID: PMC4905702          DOI: 10.7759/cureus.607

Source DB:  PubMed          Journal:  Cureus        ISSN: 2168-8184


  8 in total

1.  The stereotaxic method and radiosurgery of the brain.

Authors:  L LEKSELL
Journal:  Acta Chir Scand       Date:  1951-12-13

2.  The accuracy of frameless stereotactic intracranial radiosurgery.

Authors:  Wilko F A R Verbakel; Frank J Lagerwaard; Arjan J E Verduin; Stan Heukelom; Ben J Slotman; Johan P Cuijpers
Journal:  Radiother Oncol       Date:  2010-11-01       Impact factor: 6.280

Review 3.  Advances in technology for intracranial stereotactic radiosurgery.

Authors:  Arjun Sahgal; Lijun Ma; Eric Chang; Almon Shiu; David A Larson; Normand Laperriere; Fang-Fang Yin; May Tsao; Cynthia Menard; Parminder S Basran; Daniel Létourneau; Mostafa Heydarian; David Beachey; Vershalee Shukla; Michael Cusimano; Mojgan Hodaie; Gelareh Zadeh; Mark Bernstein; Michael Schwartz
Journal:  Technol Cancer Res Treat       Date:  2009-08

4.  Point/Counterpoint. Submillimeter accuracy in radiosurgery is not possible.

Authors:  Tewfik Bichay; Sonja Dieterich; Colin G Orton
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

5.  A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions.

Authors:  Naren Ramakrishna; Florin Rosca; Scott Friesen; Evrim Tezcanli; Piotr Zygmanszki; Fred Hacker
Journal:  Radiother Oncol       Date:  2010-01-28       Impact factor: 6.280

6.  Motion monitoring for cranial frameless stereotactic radiosurgery using video-based three-dimensional optical surface imaging.

Authors:  Guang Li; Ase Ballangrud; Li Cheng Kuo; Hyejoo Kang; Assen Kirov; Michael Lovelock; Yoshiya Yamada; James Mechalakos; Howard Amols
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  Inter- and intrafraction patient positioning uncertainties for intracranial radiotherapy: a study of four frameless, thermoplastic mask-based immobilization strategies using daily cone-beam CT.

Authors:  Erik Tryggestad; Matthew Christian; Eric Ford; Carmen Kut; Yi Le; Giuseppe Sanguineti; Danny Y Song; Lawrence Kleinberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-15       Impact factor: 7.038

8.  [Repositioning accuracy of cerebral fractionated stereotactic radiotherapy using CT scanning].

Authors:  D Pasquier; F Dubus; B Castelain; M Delplanque; V Bernier; I Buchheit; C Kerr; I Santoro; A Huchet; N Causse; E Lartigau
Journal:  Cancer Radiother       Date:  2009-07-28       Impact factor: 1.018

  8 in total
  6 in total

1.  Towards frameless maskless SRS through real-time 6DoF robotic motion compensation.

Authors:  Andrew H Belcher; Xinmin Liu; Steven Chmura; Kamil Yenice; Rodney D Wiersma
Journal:  Phys Med Biol       Date:  2017-11-13       Impact factor: 3.609

2.  Personalized tooth-supported digital guide plate used in the treatment of trigeminal neuralgia with balloon compression.

Authors:  Wen-Bin Wei; Yi-Wen Wang; Zi-Xiang Han; Zhi-Yang Liu; Yue-Min Liu; Min-Jie Chen
Journal:  Ann Transl Med       Date:  2022-06

3.  Comparison of Intrafractional Motion in Head and Neck Cancer Between Two Immobilization Methods During Stereotactic Ablative Radiation Therapy by CyberKnife.

Authors:  Chen-Lin Kang; Tsair-Fwu Lee; Shan-Ho Chan; Shyh-Chang Liu; Jui-Chu Wang; Cheng-Hsiang Tsai; Kuan-Cho Liao; Fu-Min Fang; Liyun Chang; Chun-Chieh Huang
Journal:  Cancer Manag Res       Date:  2021-01-05       Impact factor: 3.989

4.  Comparison of Skull Motions in Six Degrees of Freedom Between Two Head Supports During Frameless Radiosurgery by CyberKnife.

Authors:  Chen-Lin Kang; Shyh-Chang Liu; Jui-Chu Wang; Kuan-Cho Liao; Yu-Jie Huang; Fu-Min Fang; Tsung-I Liao; Kuo-Jung Juan; Chun-Chieh Huang
Journal:  Front Oncol       Date:  2018-09-04       Impact factor: 6.244

5.  A review of 3D printed patient specific immobilisation devices in radiotherapy.

Authors:  Amirhossein Asfia; James I Novak; Mazher Iqbal Mohammed; Bernard Rolfe; Tomas Kron
Journal:  Phys Imaging Radiat Oncol       Date:  2020-03-20

6.  Intra-Fraction Motion Management for Radiosurgical Treatments of Trigeminal Neuralgia: Clinical Experience, Imaging Frequency, and Motion Analysis.

Authors:  Nzhde Agazaryan; Stephen Tenn; Nader Pouratian; Tania Kaprealian
Journal:  Cureus       Date:  2021-04-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.