| Literature DB >> 20946342 |
Boel Bengtsson1, Sabina Andersson, Anders Heijl.
Abstract
PURPOSE: To investigate the measures of validity for selective or population screening of the time-domain Stratus and the spectral-domain Cirrus Optical Coherence Tomography (OCT) imaging techniques in a population-based sample and in clinical glaucoma patients at different stages of glaucoma. PATIENTS AND METHODS: A random sample of 307 subjects living in two rural areas in southern Sweden, and a random sample of 394 clinical glaucoma patients were selected. A large battery of examinations, including Stratus and Cirrus OCT was performed. OCT retinal nerve fibre layer (RNFL) thickness analyses for average thickness, quadrant and clock-hour sectors were compared with normative significance limits available in the instruments.Entities:
Mesh:
Year: 2010 PMID: 20946342 PMCID: PMC3440591 DOI: 10.1111/j.1755-3768.2010.01977.x
Source DB: PubMed Journal: Acta Ophthalmol ISSN: 1755-375X Impact factor: 3.761
Fig. 1Mean Deviation (MD) values of visual fields assessed in clinical glaucoma patients (dark bars) and in the nine subjects in the population-based sample who were found to be glaucomatous (white bars). Seven of the latter nine patients had mild visual field defects with MD values better than −6 dB, and two subjects, both previously diagnosed, had MD values of approximately −15 dB.
Diagnostic sensitivity of Optical Coherence Tomography Retinal Nerve Fibre Layer thickness (RNFLT) analyses in a random sample of clinical glaucoma patients
| Sensitivity for average RNFLT | Sensitivity for quadrant sector RNFLT (≥1 quadrant) | Sensitivity for clock-hour sector RNFLT (≥1 clock hour) | |||||
|---|---|---|---|---|---|---|---|
| Cirrus | Stratus | Cirrus | Stratus | Cirrus | Stratus | ||
| Overall ( | Cut-off p < 5% | 0.90 (0.83–0.94) | 0.78 (0.70–0.85) | 0.96 (0.90–0.98) | 0.93 (0.87–0.96) | 0.94 (0.89–0.97) | 0.95 (0.89–0.98) |
| Cut-off p < 1% | 0.73 (0.65–0.80) | 0.68 (0.60–0.76) | 0.91 (0.85–0.95) | 0.83 (0.76–0.89) | 0.93 (0.87–0.96) | 0.93 (0.87–0.96) | |
| MD > −6 dB ( | Cut-off p < 5% | 0.78 (0.61–0.89) | 0.45 (0.30–0.61) | 0.90 (0.75–0.97) | 0.80 (0.64–0.90) | 0.88 (0.72–0.95) | 0.85 (0.69–0.94) |
| Cut-off p < 1% | 0.38 (0.23–0.54) | 0.28 (0.15–0.44) | 0.78 (0.61–0.89) | 0.60 (0.43–0.75) | 0.80 (0.64–0.90) | 0.60 (0.43–0.75) | |
| −6 dB ≥ MD > −12 dB ( | Cut-off p < 5% | 0.88 (0.72–0.96) | 0.82 (0.65–0.93) | 0.94 (0.79–0.99) | 0.94 (0.79–0.99) | 0.97 (0.83–1.00) | 0.97 (0.83–1.00) |
| Cut-off p < 1% | 0.76 (0.58–0.89) | 0.74 (0.55–0.86) | 0.91 (0.75–0.98) | 0.85 (0.68–0.94) | 0.94 (0.79–0.99) | 0.94 (0.79–0.99) | |
| −12 dB ≥ MD > −18 dB ( | Cut-off p < 5% | 0.97 (0.81–1.00) | 0.97 (0.81–1.00) | 1.00 (0.86–1.00) | 1.00 (0.86–1.00) | 1.00 (0.86–1.00) | 1.00 (0.86–1.00) |
| Cut-off p < 1% | 0.93 (0.76–0.99) | 0.90 (0.72–0.97) | 1.00 (0.86–1.00) | 1.00 (0.86–1.00) | 1.00 (0.86–1.00) | 0.97 (0.81–1.00) | |
| MD ≤ −18 dB ( | Cut-off p < 5% | 1.00 (0.87–1.00) | 1.00 (0.87–1.00) | 1.00 (0.87–1.00) | 1.00 (0.87–1.00) | 1.00 (0.87–1.00) | 1.00 (0.87–1.00) |
Diagnostic performance of Optical Coherence Tomography Retinal Nerve Fibre Layer thickness (RNFLT) analyses in a population-based random sample
| Average RNFLT | Quadrant sector RNFLT (≥1 quadrant) | Clock hour sector RNFLT (≥1 clock hour) | |||||
|---|---|---|---|---|---|---|---|
| Cirrus | Stratus | Cirrus | Stratus | Cirrus | Stratus | ||
| AROC (95% CI) | Cut-off p < 5% | 0.96 | 0.93 | 0.94 | 0.97 | ||
| Cut-off p < 1% | 0.99 | 0.81 | 0.97 | 0.82 | |||
| Sensitivity (95% CI) | Cut-off p < 5% | 0.89 (0.68–1.00) | 0.78 (0.51–1.00) | 1.00 (1.00–1.00) | 0.89 (0.68–1.00) | 1.00 (1.00–1.00) | 1.00 (1.00–1.00) |
| Cut-off p < 1% | 0.67 (0.36–0.98) | 0.67 (0.36–0.98) | 1.00 (1.00–1.00) | 0.67 (0.36–0.98) | 1.00 (1.00–1.00) | 0.67 (0.36–0.98) | |
| Specificity (95% CI) | Cut-off p < 5% | 0.95 (0.92–0.99) | 0.99 (0.98–1.00) | 0.81 (0.74–0.88) | 0.93 (0.89–0.97) | 0.65 (0.57–0.73) | 0.81 (0.74–0.87) |
| Cut-off p < 1% | 0.98 (0.95–1.00) | 1.00 (1.00–1.00) | 0.96 (0.93–1.00) | 0.98 (0.95–1.00) | 0.91 (0.86–0.96) | 0.97 (0.94–1.00) | |
| Positive predictive value (95% CI) | Cut-off p < 5% | 0.57 (0.31–0.83) | 0.88 (0.65–1.00) | 0.27 (0.12–0.41) | 0.47 (0.23–0.71) | 0.17 (0.07–0.27) | 0.27 (0.12–0.41) |
| Cut-off p < 1% | 0.67 (0.36–0.98) | 1.00 (1.00–1.00) | 0.64 (0.39–0.89) | 0.67 (0.36–0.98) | 0.43 (0.22–0.64) | 0.60 (0.30–0.90) | |
| Negative predictive value (95% CI) | Cut-off p < 5% | 0.99 (0.98–1.00) | 0.99 (0.96–1.00) | 1.00 (1.00–1.00) | 0.99 (0.98–1.00) | 1.00 (1.00–1.00) | 1.00 (1.00–1.00) |
| Cut-off p < 1% | 0.98 (0.95–1.00) | 0.98 (0.95–1.00) | 1.00 (1.00–1.00) | 0.98 (0.95–1.00) | 1.00 (1.00–1.00) | 0.98 (0.95–1.00) | |
AROC, area under the receiver operating characteristic curve.
95% CI, confidence intervals at the 95% significance level in parenthesis.
AROCs for all quadrant sectors.