Michael Dandel1, Yuguo Weng, Henryk Siniawski, Alexander Stepanenko, Thomas Krabatsch, Evgenij Potapov, Hans B Lehmkuhl, Christoph Knosalla, Roland Hetzer.
Abstract
AIMS: Unloading-promoted reversal of heart failure (HF) allows long-term transplant-free outcome after ventricular assist device (VAD) removal. However, because few patients with chronic cardiomyopathy (CCM) were weaned from VADs (the majority only recently), the reliability of criteria used for weaning decisions to predict long-term post-weaning success is barely known. After 15 years of weaning experience, we assessed this issue. METHODS ANDEntities:
Mesh:
Year: 2010 PMID: 20929978 PMCID: PMC3086897 DOI: 10.1093/eurheartj/ehq353
Source DB: PubMed Journal: Eur Heart J ISSN: 0195-668X Impact factor: 29.983
Main weaning criteria used since March 2002
| Examination | Parameters and parameter-derived measurements obtained during the final pre-explantation off-pump triala |
|---|---|
| Electrocardiography | Sinus rhythm |
| HR <90b.p.m. | |
| Not more than 25% HR increase during off-pump trials | |
| Brachial artery pressure | Mean ≥65 mmHg |
| Echocardiography | LVEDD ≤55 mm |
| LVEF ≥45% | |
| No or maximum grade II mitral and/or aortic valve regurgitation | |
| No RV dilation (RVOT diameter <35 mm, short-/long-axis ratio <0.6) | |
| No or maximum grade II tricuspid or pulmonary valve regurgitation | |
| Right-heart catheterization | Cardiac index >2.6 L/min/m2 |
| Pulmonary artery wedge pressure (mean) <13 mmHg | |
| Right atrial pressure (mean) <10 mmHg |
aMeasurements performed at rest, without any inotropic support.
Weaned patient characteristics at the time of ventricular assist device implantation: comparison of patients who exhibited long-lasting cardiac stability with those showing heart failure recurrence during the first 5 years after ventricular assist device removal
| Main characteristics | All chronic HF patients weaned from VADsa ( | Patients evaluated for long-lasting recovery after VAD removala ( | ||
|---|---|---|---|---|
| Long-lasting cardiac improvement (>5 years) ( | Post-weaning recurrence of HF (in ≤5 years) ( | |||
| Age (years) | 42.4 ± 14.4 | 40.1 ± 12.1 | 47.6 ± 9.7 | 0.03c |
| Sex [ | ||||
| Male | 42 (95.5) | 18 (94.7) | 14 (100) | 1.0 |
| Female | 2 (4.5) | 1 (5.3) | 0 | |
| Diagnosis [ | ||||
| Non-ischaemic chronic CM | 42 (95.5) | 17 (89.5) | 14 (100) | 0.9 |
| Ischaemic chronic CM | 2 (4.5) | 2 (10.5) | 0 | |
| History of HF (years) | 4.1 ± 3.8 | 2.4 ± 1.5 | 6.4 ± 4.5 | 0.01c |
| Patients with LVADs [ | 42 (95.5) | 18 (94.7) | 13 (92.9) | 1 |
| Patients with BVADs [ | 2 (4.5) | 1 (5.3) | 1 (7.1) | |
| Heart rhythm [ | ||||
| Sinus rhythm | 43 (97.7) | 19 (100) | 13 (92.9) | 0.42 |
| Atrial fibrillation | 1 (2.3) | 0 | 1 (7.1) | |
| LVEF (%) | 15.4 ± 3.4 | 15.0 ± 2.6 | 15.3 ± 3.4 | 0.79 |
| LVEDD (mm) | 74.3 ± 7.5 | 73.1 ± 4.3 | 78.4 ± 4.1 | 0.09 |
| Fibrosis (%) | 12.1 ± 9.6 | 9.8 ± 5.2 | 16.0 ± 10.5 | 0.08 |
| Serum A-β1-AABs (LU) | 5.4 ± 2.1 | 5.2 ± 1.7 | 5.4 ± 1.9 | 0.95 |
LU, laboratory units.
aValues are means ± SD.
bOnly 33 of the 47 weaned patients with CCM could be used for comparison. Patients with stable cardiac function who died during the first 5 post-weaning years due to different causes not related to VAD removal and patients who were weaned <5 years before this evaluation were not included.
cStatistically significant.
Patient characteristics at the time of ventricular assist device explantation: comparison of patients who exhibited long-lasting cardiac stability with those showing heart failure recurrence during the first 5 years after ventricular assist device removal
| Patient characteristics | All chronic HF patients weaned from VADsa ( | Patients evaluated for long-lasting recovery after VAD removala ( | ||
|---|---|---|---|---|
| Long-lasting cardiac improvement (>5 years) ( | Post-weaning recurrence of HF (in ≤5 years) ( | |||
| LVEF (%) | 47.2 ± 6.8 | 49.2 ± 2.6 | 42.4 ± 3.4 | <0.01 |
| LVEF change (% of best value) | −7.7 ± 9.6 | −4.2 ± 4.4 | −16.2 ± 19.5 | <0.001 |
| LVEDD (mm) | 52.2 ± 7.5 | 49.2 ± 3.9 | 56.2 ± 5.9 | <0.01 |
| LVEDD change (% of best value) | 9.0 ± 9.6 | 4.7 ± 3.5 | 16.9 ± 6.3 | <0.001 |
| RWT | 0.39 ± 0.07 | 0.42 ± 0.04 | 0.35 ± 0.04 | <0.001 |
| 8.5 ± 1.3 | 9.5 ± 2.2 | 7.0 ± 1.9 | <0.01 | |
| 8.8 ± 1.4 | 9.8 ± 2.6 | 7.3 ± 1.5 | <0.01 | |
| RWT change during off-pump trials (%) | 8.6 ± 6.9 | −5.5 ± 3.5 | −15.8 ± 0.4 | <0.001 |
| LV sphericity Index (short-/long-axis ratio) | 0.66 ± 0.07 | 0.64 ± 0.09 | 0.72 ± 0.07 | 0.05 |
| Duration of LVAD support (months) | 4.9 ± 2.8 | 3.5 ± 2.1 | 6.7 ± 5.2 | 0.02 |
| Time (months) to disappearance of serum A-β1-AABsc | 2.6 ± 0.3 | 2.5 ± 1.7 | 2.6 ± 1.9 | 0.90d |
aValues are means ± SD.
bOnly 33 of the 47 weaned patients with CCM could be used for comparison. The six patients with stable cardiac function who died during the first 5 post-weaning years due to different causes not related to VAD removal and patients who were weaned <5 years before this evaluation were not included.
cAnti-β1-adrenoceptor auto-antibodies.
dStatistically not significant.
Echocardiographic prediction of post-weaning cardiac stability in patients with chronic cardiomyopathy as the underlying cause for ventricular assist device implantation
| Selected parameters and cut-off values | Prediction of freedom from HF recurrence | |||||
|---|---|---|---|---|---|---|
| For ≥5 years | For ≥7 years | |||||
| Sensitivity (%) | Specificity (%) | Predictive value (%) | Sensitivity (%) | Specificity (%) | Predictive value (%) | |
| Finala off-pump LVEF ≥50% | 55.0 | 92.3 | 91.7 | 57.1 | 77.8 | 66.7 |
| at LVEDD ≤55 mm | 55.0 | 100 | 100 | 57.1 | 83.3 | 72.7 |
| plus history of HF ≤5 years | 50.0 | 100 | 100 | 57.1 | 88.9 | 80.00 |
| Finala off-pump LVEF 45% | 95.0 | 61.5 | 79.4 | 100 | 50.00 | 60.9 |
| plus pre-weaning LVEF stabilityb | 90.0 | 76.9 | 85.7 | 92.9 | 61.1 | 65.00 |
| at LVEDD ≤55 mm | 95.0 | 76.9 | 86.4 | 100 | 61.1 | 66.7 |
| at LVEDD ≤55 mm plus pre-weaning LVEDD stabilityb | 80.0 | 92.3 | 94.1 | 78.6 | 72.2 | 68.8 |
| at LVEDD ≤55 mm plus pre-weaning LVEF and LVEDD stabilityb | 80.0 | 92.3 | 94.1 | 78.6 | 71.4 | 73.3 |
| plus history of HF ≤5 years | 90.0 | 76.9 | 85.7 | 100 | 72.2 | 73.7 |
| at LVEDD ≤55 mm with pre-weaning stabilityb plus history of HF ≤5 years | 85.0 | 100 | 100 | 78.6 | 83.3 | 78.6 |
| without alteration during the first post-weaning monthc | 90.0 | 84.6 | 90.0 | 92.9 | 61.1 | 65.0 |
aLast pump-stop before VAD explantation.
bNot more than 10% alteration in comparison with the best of all off-pump values until VAD removal.
cNot more than 10% alteration in comparison with LVEF at VAD removal.
Risk factors for heart failure recurrence after ventricular assist device removal in patients with chronic cardiomyopathy who showed myocardial improvement during ventricular unloading
| Major risk factors | HF recurrence after weaning | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| During first 3 years | During first 5 years | During first 7 years | |||||||
| Events with risk factors | Number of events with risk factors | Predictive value (%) | Events with risk factors | Number of events with risk factors | Predictive value (%) | Events with risk factors | Number of events with risk factors | Predictive value (%) | |
| Off-pump LVEF <45% | 8 | 1 | 88.9 | 8 | 1 | 88.9 | 9 | 0 | 100 |
| Off-pump LVEF <45% plus history of HF >5 years | 7 | 0 | 100 | 7 | 0 | 100 | 7 | 0 | 100 |
| Off-pump LVEF <50 and >10% alteration to the best off-pump value | 7 | 1 | 88.0 | 8 | 0 | 100 | 8 | 0 | 100 |
| Off-pump LVEF ≤50% at LVEDD >55 mm plus history of HF >5 years | 8 | 0 | 100 | 8 | 0 | 100 | 8 | 0 | 100 |
| Off-pump LVEF ≤50% at LVEDD >55 mm | 9 | 1 | 90.0 | 9 | 1 | 90.00 | 10 | 0 | 100 |
| Off-pump LVEF ≤50% plus RWT ≤0.37 | 10 | 2 | 83.3 | 11 | 1 | 91.7 | 12 | 0 | 100 |
| Off-pump LVEF ≤50% with RWT decrease with >10% during off-pump trials | 10 | 2 | 83.3 | 11 | 1 | 91.7 | 11 | 1 | 91.7 |
| Early post-weaning LVEF alteration with >10% | 11 | 2 | 84.6 | 11 | 2 | 91.7 | 11 | 1 | 91.7 |
Relevance of main risk factors for heart failure recurrence
| Selected risk factors for HF recurrence | Odds ratioa | 95% CI | |
|---|---|---|---|
| Final off-pump LVEF | 1.49 | 1.12–1.98 | 0.007 |
| Pre-weaning LVEF stabilityb | 1.18 | 1.04–1.35 | 0.014 |
| Final off-pump LVEDD | 1.26 | 1.06–1.50 | 0.010 |
| Pre-weaning LVEDD stabilityb | 1.19 | 1.05–1.34 | 0.006 |
| Duration of HF before VAD support | 1.62 | 1.09–2.41 | 0.016 |
aRisk for HF with alteration by one unit.
bNot more than 10% alteration in comparison with the best of all off-pump values until VAD removal.