Literature DB >> 8783992

Improved left ventricular function after chronic left ventricular unloading.

O H Frazier1, C R Benedict, B Radovancevic, R J Bick, P Capek, W E Springer, M P Macris, R Delgado, L M Buja.   

Abstract

BACKGROUND: This study assessed the effect of prolonged left ventricular unloading on native ventricular function.
METHODS: We reviewed data from 31 patients (30 men, 1 woman) supported more than 30 days (mean, 137 days; range, 31 to 505 days) with the HeartMate left ventricular assist system. The patients' mean age was 46 years (range, 22 to 64 years); 17 had idiopathic and 14 had ischemic cardiomyopathy. Data (anatomic, physiologic, hemodynamic, histologic, and biochemical) were collected at the time of HeartMate implantation, during support with the device temporarily off, and at the time of device explantation.
RESULTS: Routine chest roentgenogram showed improvement in cardiothoracic ratio (0.62 +/- 0.04 to 0.55 +/- 0.03; p < 0.0001). Echocardiography performed with the pump off showed a significant decrease in left ventricular end-diastolic dimension (6.81 +/- 0.87 cm to 5.39 +/- 1.08 cm; p < 0.0005) and a significant improvement in ejection fraction (0.11 +/- 0.05 to 0.22 +/- 0.17; p < 0.02). Cardiac index increased (1.96 +/- 0.52 L.min-1.m-2 to 2.93 +/- 0.73 L.min-1.m-2; p < 0.0001), mean aortic pressure increased (71.40 +/- 10.63 mm Hg to 76.33 +/- 16.84 mm Hg; p = 0.48), pulmonary capillary wedge pressure decreased (24.18 +/- 6.27 mm Hg to 14.48 +/- 3.01 mm Hg; p < 0.0001), and pulmonary vascular resistance decreased (3.34 +/- 2.00 Wood units to 2.51 +/- 0.88 Wood units; p < 0.05). Comparisons of tissue samples taken at the time of implantation and at the time of transplantation showed a marked reduction in myocytolysis. Calcium uptake, calcium-binding rates, and lipid levels normalized in patients studied. Plasma norepinephrine levels decreased to near normal levels.
CONCLUSION: Prospective studies are now indicated to determine whether device removal without transplantation may be beneficial in selected patients.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8783992     DOI: 10.1016/s0003-4975(96)00437-7

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  51 in total

Review 1.  New developments in the treatment of cardiac failure.

Authors:  S Westaby; O Franklin; M Burch
Journal:  Arch Dis Child       Date:  1999-09       Impact factor: 3.791

Review 2.  Non-transplant surgery for heart failure.

Authors:  S Westaby
Journal:  Heart       Date:  2000-05       Impact factor: 5.994

Review 3.  Pathophysiology of heart failure: a look at the future.

Authors:  R M Delgado; J T Willerson
Journal:  Tex Heart Inst J       Date:  1999

4.  Compact, reliable ventricular assist device as a bridge to recovery or for semipermanent use.

Authors:  K Nishimura; S Kono; T Nishina; K Ueyama; A Ikai; T Ikeda; C Nojiri; T Akamatsu; M Komeda
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2001-11

Review 5.  New surgical treatments for heart failure.

Authors:  Mauricio J Garrido; Mehmet C Oz
Journal:  Curr Cardiol Rep       Date:  2002-05       Impact factor: 2.931

6.  An in vitro fluid dynamic study of pediatric cannulae: the value of animal studies to predict human flow.

Authors:  Tobias C Long; Joseph J Pearson; Andrew C Hankinson; Steven Deutsch; Keefe B Manning
Journal:  J Biomech Eng       Date:  2012-04       Impact factor: 2.097

Review 7.  Novel Left Ventricular Assist System: an electrocardiogram-synchronized LVAS that avoids cardiac cannulation.

Authors:  Domingo Liotta
Journal:  Tex Heart Inst J       Date:  2003

Review 8.  Reverse cardiac remodeling enabled by mechanical unloading of the left ventricle.

Authors:  Konstantinos G Malliaras; John V Terrovitis; Stavros G Drakos; John N Nanas
Journal:  J Cardiovasc Transl Res       Date:  2008-09-30       Impact factor: 4.132

Review 9.  Epidemiology of "Heart Failure with Recovered Ejection Fraction": What do we do After Recovery?

Authors:  Johny S Kuttab; Michael S Kiernan; Amanda R Vest
Journal:  Curr Heart Fail Rep       Date:  2015-12

10.  Change in myocardial oxygen consumption employing continuous-flow LVAD with cardiac beat synchronizing system, in acute ischemic heart failure models.

Authors:  Akihide Umeki; Takashi Nishimura; Yoshiaki Takewa; Masahiko Ando; Mamoru Arakawa; Yuichiro Kishimoto; Tomonori Tsukiya; Toshihide Mizuno; Shunei Kyo; Minoru Ono; Yoshiyuki Taenaka; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2013-01-17       Impact factor: 1.731