Performance evaluation of the FLEX triumph X-PET scanner using the national electrical manufacturers association NU-4 standards.

Rameshwar Prasad1, Osman Ratib, Habib Zaidi.   

Abstract

UNLABELLED: The purpose of this work was to evaluate the performance characteristics of the preclinical X-PET subsystem of the FLEX Triumph PET/CT scanner based on the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA).
METHODS: The performance parameters evaluated include the spatial resolution, scatter fraction, count losses and random coincidences, sensitivity, and image-quality characteristics. The PET detector array consisted of 11,520 individual bismuth germanate crystals arranged in 48 rings and 180 blocks, with an axial field of view (FOV) of 11.6 cm and a inner ring diameter of 16.5 cm. The spatial resolution was measured with a small (22)Na point source (diameter, 0.25 mm) at different radial offsets from the center. Sensitivity was calculated using the same source by stepping the source axially through the axial FOV of the scanner. Scatter fraction and counting-rate performances were determined using a mouse- and rat-sized phantom with an (18)F line source insert. The NEMA image-quality phantom and rodent imaging were also performed to access the overall imaging capabilities of the scanner.
RESULTS: Tangential spatial resolution in terms of full width at half maximum varied between 2.2 mm at the center of the FOV and 2.3 mm at a radial offset of 2.5 cm. The radial spatial resolution varied between 2.0 at the center and 4.4 mm at a radial offset of 2.3 cm. The peak system absolute sensitivity was 5.9% at the center of the FOV. The absolute system sensitivity was 0.67 counts/s/Bq, and the relative total system sensitivity was 73.9%. The scatter fraction for the mouse-sized phantom was 7.9%, with a peak true counting rate of 168 kilocounts per second (kcps) at 0.3 MBq/mL and a peak noise-equivalent counting rate of 106 kcps at 0.17 MBq/mL. The rat-sized phantom had a scatter fraction of 21%, with a peak true counting rate of 93 kcps at 0.034 MBq/mL and a peak noise-equivalent counting rate of 49 kcps at 0.02 MBq/mL. Recovery coefficients for the image-quality phantom ranged from 0.13 to 0.88.
CONCLUSION: The performance of the X-PET scanner based on the NEMA NU 4-2008 standards was fully characterized. The overall performance demonstrates that the X-PET system is suitable for preclinical research.

Entities:  

Mesh:

Year:  2010        PMID: 20847182     DOI: 10.2967/jnumed.110.076125

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  11 in total

1.  Scatter characterization and correction for simultaneous multiple small-animal PET imaging.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2014-04       Impact factor: 3.488

2.  A cone-shaped phantom for assessment of small animal PET scatter fraction and count rate performance.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2012-10       Impact factor: 3.488

3.  (18)F-FDG PET/CT imaging detects therapy efficacy of anti-EMMPRIN antibody and gemcitabine in orthotopic pancreatic tumor xenografts.

Authors:  Nemil Shah; Guihua Zhai; Joseph A Knowles; Cecil R Stockard; William E Grizzle; Naomi Fineberg; Tong Zhou; Kurt R Zinn; Eben L Rosenthal; Hyunki Kim
Journal:  Mol Imaging Biol       Date:  2012-04       Impact factor: 3.488

4.  Small-animal imaging using clinical positron emission tomography/computed tomography and super-resolution.

Authors:  Frank P DiFilippo; Sagar Patel; Kewal Asosingh; Serpil C Erzurum
Journal:  Mol Imaging       Date:  2012-06       Impact factor: 4.488

5.  Salutary Effects of Estrogen Sulfate for Traumatic Brain Injury.

Authors:  Hyunki Kim; Betul Cam-Etoz; Guihua Zhai; William J Hubbard; Kurt R Zinn; Irshad H Chaudry
Journal:  J Neurotrauma       Date:  2015-04-22       Impact factor: 5.269

6.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

7.  Dual combination therapy targeting DR5 and EMMPRIN in pancreatic adenocarcinoma.

Authors:  Hyunki Kim; Guihua Zhai; Sharon L Samuel; Christopher J Rigell; Heidi R Umphrey; Samir Rana; Cecil R Stockard; Naomi S Fineberg; Kurt R Zinn
Journal:  Mol Cancer Ther       Date:  2011-12-27       Impact factor: 6.261

8.  Automated analysis of small animal PET studies through deformable registration to an atlas.

Authors:  Daniel F Gutierrez; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-07-21       Impact factor: 9.236

9.  [¹¹C]acetate and PET/CT assessment of muscle activation in rat studies.

Authors:  Sara Trombella; David García; Didier J Colin; Stéphane Germain; Yann Seimbille; Osman Ratib
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-07-26       Impact factor: 2.924

10.  NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph.

Authors:  Z Gu; R Taschereau; N T Vu; H Wang; D L Prout; R W Silverman; B Bai; D B Stout; M E Phelps; A F Chatziioannou
Journal:  Phys Med Biol       Date:  2013-05-10       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.