Measuring venous oxygenation using the photoplethysmograph waveform.

Zachary D Walton1, Panayiotis A Kyriacou, David G Silverman, Kirk H Shelley.   

Abstract

OBJECTIVE: We investigate the hypothesis that the photoplethysmograph (PPG) waveform can be analyzed to infer regional venous oxygen saturation.
METHODS: Fundamental to the successful isolation of the venous saturation is the identification of PPG characteristics that are unique to the peripheral venous system. Two such characteristics have been identified. First, the peripheral venous waveform tends to reflect atrial contraction. Second, ventilation tends to move venous blood preferentially due to the low pressure and high compliance of the venous system. Red (660 nm) and IR (940 nm) PPG waveforms were collected from 10 cardiac surgery patients using an esophageal PPG probe. These waveforms were analyzed using algorithms written in Mathematica. Four time-domain saturation algorithms (ArtSat, VenSat, ArtInstSat, VenInstSat) and four frequency-domain saturation algorithms (RespDC, RespAC, Cardiac, and Harmonic) were applied to the data set.
RESULTS: Three of the algorithms for calculating venous saturation (VenSat, VenInstSat, and RespDC) demonstrate significant difference from ArtSat (the conventional time-domain algorithm for measuring arterial saturation) using the Wilcoxon signed-rank test with Bonferroni correction (p < 0.0071).
CONCLUSIONS: This work introduces new algorithms for PPG analysis. Three algorithms (VenSat, VenInstSat, and RespDC) succeed in detecting lower saturation blood. The next step is to confirm the accuracy of the measurement by comparing them to a gold standard (i.e., venous blood gas).

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20644985     DOI: 10.1007/s10877-010-9248-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  13 in total

1.  Evaluation of oesophageal pulse oximetry in patients undergoing cardiothoracic surgery.

Authors:  P A Kyriacou; S L Powell; D P Jones; R M Langford
Journal:  Anaesthesia       Date:  2003-05       Impact factor: 6.955

2.  The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in Spo2 calculation.

Authors:  Kirk H Shelley; Doris Tamai; Denis Jablonka; Michael Gesquiere; Robert G Stout; David G Silverman
Journal:  Anesth Analg       Date:  2005-03       Impact factor: 5.108

Review 3.  Pulse oximetry in the oesophagus.

Authors:  P A Kyriacou
Journal:  Physiol Meas       Date:  2005-11-29       Impact factor: 2.833

4.  Variations in arterial blood pressure and photoplethysmography during mechanical ventilation.

Authors:  Giuseppe Natalini; Antonio Rosano; Maria E Franceschetti; Paola Facchetti; Achille Bernardini
Journal:  Anesth Analg       Date:  2006-11       Impact factor: 5.108

Review 5.  Photoplethysmography and its application in clinical physiological measurement.

Authors:  John Allen
Journal:  Physiol Meas       Date:  2007-02-20       Impact factor: 2.833

Review 6.  Peripheral venous pressure waveform.

Authors:  Richa Wardhan; Kirk Shelley
Journal:  Curr Opin Anaesthesiol       Date:  2009-12       Impact factor: 2.706

7.  Measurement of venous oxyhaemoglobin saturation in the adult human forearm by near infrared spectroscopy with venous occlusion.

Authors:  C W Yoxall; A M Weindling
Journal:  Med Biol Eng Comput       Date:  1997-07       Impact factor: 2.602

8.  A system for investigating oesophageal photoplethysmographic signals in anaesthetised patients.

Authors:  P A Kyriacou; A R Moye; A Gregg; D M Choi; R M Langford; D P Jones
Journal:  Med Biol Eng Comput       Date:  1999-09       Impact factor: 2.602

9.  The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph.

Authors:  K H Shelley; M Dickstein; S M Shulman
Journal:  J Clin Monit       Date:  1993-09

10.  Investigation of oesophageal photoplethysmographic signals and blood oxygen saturation measurements in cardiothoracic surgery patients.

Authors:  P A Kyriacou; S Powell; R M Langford; D P Jones
Journal:  Physiol Meas       Date:  2002-08       Impact factor: 2.833

View more
  9 in total

1.  Frequency domain analysis of cerebral near infrared spectroscopy signals during application of an impedance threshold device in spontaneously ventilating volunteers.

Authors:  Douglas A Colquhoun; Kimberly Naden; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2015-06-27       Impact factor: 2.502

2.  Measuring blood oxygen saturation along a capillary vessel in human.

Authors:  Kfir Akons; Eldad J Dann; Dvir Yelin
Journal:  Biomed Opt Express       Date:  2017-10-30       Impact factor: 3.732

3.  Non-invasive estimation of jugular venous oxygen saturation: a comparison between near infrared spectroscopy and transcutaneous venous oximetry.

Authors:  Douglas A Colquhoun; Jason M Tucker-Schwartz; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-04       Impact factor: 2.502

Review 4.  Respiratory physiology and the impact of different modes of ventilation on the photoplethysmographic waveform.

Authors:  Aymen A Alian; Kirk H Shelley
Journal:  Sensors (Basel)       Date:  2012-02-16       Impact factor: 3.576

5.  Comparison of non-invasive peripheral venous saturations with venous blood co-oximetry.

Authors:  A M Belhaj; J P Phillips; P A Kyriacou; R M Langford
Journal:  J Clin Monit Comput       Date:  2016-11-21       Impact factor: 2.502

6.  Proof of concept non-invasive estimation of peripheral venous oxygen saturation.

Authors:  Musabbir Khan; Chris G Pretty; Alexander C Amies; Joel Balmer; Houda E Banna; Geoffrey M Shaw; J Geoffrey Chase
Journal:  Biomed Eng Online       Date:  2017-05-19       Impact factor: 2.819

Review 7.  Sources of Inaccuracy in Photoplethysmography for Continuous Cardiovascular Monitoring.

Authors:  Jesse Fine; Kimberly L Branan; Andres J Rodriguez; Tananant Boonya-Ananta; Jessica C Ramella-Roman; Michael J McShane; Gerard L Coté
Journal:  Biosensors (Basel)       Date:  2021-04-16

Review 8.  Pulse oximetry: fundamentals and technology update.

Authors:  Meir Nitzan; Ayal Romem; Robert Koppel
Journal:  Med Devices (Auckl)       Date:  2014-07-08

Review 9.  The Various Oximetric Techniques Used for the Evaluation of Blood Oxygenation.

Authors:  Meir Nitzan; Itamar Nitzan; Yoel Arieli
Journal:  Sensors (Basel)       Date:  2020-08-27       Impact factor: 3.576

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.