Literature DB >> 29417727

COMP report: CPQR technical quality control guidelines for brachytherapy remote afterloaders.

Normand Frenière1.   

Abstract

The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control testing of brachytherapy remote afterloaders.
© 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  brachytherapy remote afterloaders; quality control guidelines; radiation treatment therapy equipment

Mesh:

Year:  2018        PMID: 29417727      PMCID: PMC5859323          DOI: 10.1002/acm2.12272

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

The Canadian Partnership for Quality Radiotherapy (CPQR) is an alliance amongst the three key national professional organizations involved in the delivery of radiation treatment in Canada: the Canadian Association of Radiation Oncology (CARO), the Canadian Organization of Medical Physicists (COMP), and the Canadian Association of Medical Radiation Technologists (CAMRT). Financial and strategic backing is provided by the federal government through the Canadian Partnership Against Cancer (CPAC), a national resource for advancing cancer prevention and treatment. The mandate of the CPQR is to support the universal availability of high quality and safe radiotherapy for all Canadians through system performance improvement and the development of consensus‐based guidelines and indicators to aid in radiation treatment program development and evaluation. This document contains detailed performance objectives and safety criteria for Brachytherapy Remote Afterloaders. Please refer to the overarching document Technical Quality Control Guidelines for Canadian Radiation Treatment Centres 1 for a programmatic overview of technical quality control, and a description of how the performance objectives and criteria listed in this document should be interpreted. The development of the individual TQC guidelines is spearheaded by expert reviewers and involves broad stakeholder input from the medical physics and radiation oncology community.2 All information contained in this document is intended to be used at the discretion of each individual center to help guide quality and safety program improvement. There are no legal standards supporting this document; specific federal or provincial regulations and licence conditions take precedence over the content of this document.

SYSTEM DESCRIPTION

Brachytherapy is the placement of encapsulated radionuclides or a miniaturized x‐ray tube in, or adjacent to, tissue which has been prescribed a radiation dose.3, 4, 5, 6, 7, 8, 9, 10 This practice offers unique advantages to the management of several treatment sites and has been used to complement or replace external beam radiation therapy since the onset of radiation oncology. Remote afterloading equipment was developed to reduce, and in many cases eliminate, the radiation exposure to members of the staff. With remote afterloading systems the user does not directly handle the radioactive source and the patient is irradiated in a shielded room with staff operating and monitoring the process remotely. High dose rate (HDR) refers to treatment dose rates larger than 20 cGy/min. For all HDR remote afterloaders, a single and small (< 1 mm × 5 mm) radioactive source (mostly iridium‐192, rarely cobalt60), laser‐welded to a metallic cable, is moved out of the safe by a motor‐drive mechanism to step along the prescribed positions (dwell positions) with different irradiation times (dwell times). The user can preselect dwell positions and dwell times at selected positions in a number of applicator lines. The remote afterloader could receive up to two such sources with two independent cables permitting dose delivery in two applicator lines simultaneously. The source strength is approximately 40,000 cGy cm² h−1 (activity of ~370 GBq) on installation of a new iridium‐192 source, while it is of 23,000 cGy cm² h−1 (activity of ~74 GBq) for a new cobalt‐60 source. Because iridium‐192 has a relatively short half‐life (73.8 days), the sources are usually replaced about every 3 months. Cobalt‐60 has a longer half‐life (1 925 days or 5.27 yr), offering less frequent source replacement to every 5 yr. Typical HDR irradiation times are 5 to 30 min, and a treatment course may consist of several fractions. Another form of treatment is pulsed dose rate (PDR) treatment. With a PDR device, irradiations are given in short “pulses” with the total treatment being given in 48 to 72 h. The mechanism for PDR units is very similar to that used in HDR units. PDR remote afterloaders also use a single iridium‐192 source attached to a cable. However, the source strength for these units is typically only 10% of the source strength of an iridium‐192 HDR unit. High‐dose rate miniaturized x‐ray tube remote devices, along with intravascular and cardiovascular brachytherapy remote afterloaders using beta emitting radioactive sources, are beyond the scope of this document. Various recommendations for brachytherapy quality assurance have been reported in the literature.11, 12, 13, 14, 15, 16, 17, 18 “Per treatment” tests must be executed prior to each treatment. “Treatment day” tests must be scheduled before treating the first patient of the day. For PDR remote afterloaders where treatments may last several days, “treatment day” tests should be performed prior to the initiation of the treatment. Per treatment quality control tests Quarterly quality control tests Annual quality control tests

RELATED TECHNICAL QUALITY CONTROL GUIDELINES

In order to comprehensively assess brachytherapy remote afterloader performance, additional guideline tests, as outlined in related CPQR Technical Quality Control (TQC) guidelines, must also be completed and documented, as applicable. Related TQC guidelines, available at http://www.cpqr.ca/programs/technical-quality-control/, include: Safety Systems19 Major Dosimetry Equipment20 Treatment Planning Systems21

TEST TABLES

Notes on per treatment tests

Notes on treatment day (or per treatment for PDR) tests

Notes on quarterly tests

Notes on annual tests

CONFLICT OF INTEREST

No conflicts of interest.

DISCLAIMER

All information contained in this document is intended to be used at the discretion of each individual centre to help guide quality and safety program improvement. There are no legal standards supporting this document; specific federal or provincial regulations and licence conditions take precedence over the content of this document. As a living document, the information contained within this document is subject to change at any time without notice. In no event shall the Canadian Partnership for Quality Radiotherapy (CPQR) or its partner associations, the Canadian Association of Radiation Oncology (CARO), the Canadian Organization of Medical Physicists (COMP), and the Canadian Association of Medical Radiation Technologists (CAMRT), be liable for any damages, losses, expenses, or costs whatsoever arising in connection with the use of this document.
Table 1

Per treatment quality control tests

DesignatorTestPerformance
Action
Per treatment (executed prior to each treatment)
T1Plan data transfer from treatment planning computerData integrity
T2Plan dwell times adjustmentSee note
T3Minimum dwell timesSee note
T4Plan catheters’ connection to remote afterloaderReproducible
T5Complete source retractionFunctional
Treatment day (or per treatment for PDR)
D1Treatment interruptFunctional
D2Console displays (treatment status indicator) and key switchFunctional
D3Date, time, and source strength in treatment unitSee note
D4Source (and dummy) positional accuracy2 mm
D5Dwell time accuracy2%
Table 2

Quarterly quality control tests

DesignatorTestPerformance
Action
Quarterly (or at source replacement)
Q1Mechanical integrity of applicators, guide tubes, connectorsFunctional
Q2Internal battery power supply (power failure recovery)Functional
Q3Source/dummy interlocksFunctional
Q4Dummy wire positional accuracy3 mm (1 mm see note)
Q5Radiological source positional accuracy1 mm
Q6Source strength calibration5%
Q7Source homogeneityReproducible
Q8RecordsComplete
Table 3

Annual quality control tests

DesignatorTestPerformance
Action
Annually
A1Hand crank operationFunctional
A2Leakage radiationReproducible
A3Multi‐channel indexer functionFunctional
A4Dwell time accuracy1%
A5Timer linearity1%
A6Transit time/transit dose reproducibilityReproducible
A7Dosimetric length of applicators and guide tubes1 mm
A8Applicators and templates dimensionsReproducible
A9Shield integrity of shielded applicatorsReproducible
A10X ray marker positional accuracy1 mm
A11Document staff review of emergency response proceduresComplete
A12Independent quality control reviewComplete
A13PDR sequencing (for PDR only)Functional
T1Plan data imported from a treatment planning system into the treatment console should be verified for source strength, dwell positions, and dwell times. In case of many dwell positions, the verification of a sub‐set of positions is acceptable. For a plan already present in the treatment console, the same verification should be made to assure proper plan selection
T2Plan dwell times adjustment by the treatment console for the treatment date should be verified (by an independent calculation: hand calculation, decay factor chart, or software calculation). Action level will depend on the treatment console decay frequency and time resolution; express in percentage of difference and/or in second
T3Minimum dwell time should be verified against the device driven limits. Those limits should take into account both the effect of transit dose and positioning reproducibility. Some remote afterloader systems might have a positioning reproducibility dependence on dwell time. All dwell times should be equal or greater than set limit
T4Catheters/applicators connections to remote afterloader indexer channels must match plan
T5Survey the treatment room and patient to ensure that source has been completely retracted
D1During source exposure, verify that the non‐emergency interrupt button (if equipped) retracts the source to its safe and shielded position
D2On the treatment console, displays should be verified. At minimum, treatment status indicators should be verified by exposing a source. Indicators could be visual and audible. When a treatment key is available, its deactivation should prevent source exposition
D3Remote afterloader console date and time are properly set. Decayed source strength is accurate compared to an independent calculation (hand calculation, decay factor chart or software calculation), taking into account treatment console decay frequency
D4Verify accuracy of source drive mechanism positioning. A visual inspection with a camera is acceptable. Apply also to dummy drive mechanism if used to measure catheter length
D5Comparison of dwell time accuracy with external standard such as a stopwatch. The dwell time used should be sufficiently long such that errors in the measurement of the time (e.g., reaction time of the observer) are less than 1%
Q1Verify the applicators, guide tubes, and connectors are exempt of damage (excessive wear, kinks, etc.)
Q2The configuration of this test will depend on the design of the facility and equipment. Safety is the concern and tests should be designed accordingly. The first objective is to verify that the equipment safely retracts the source wire after a power failure. The second objective is to verify that the equipment properly records treatment delivered before power failure and permits to resume the treatment after power recovery
Q3Verify functionality of remote afterloader interlocks related to source and dummy wires. This includes incorrect connection of applicator to transfer guide tube, incorrect connection of transfer guide tube to remote afterloader, and obstruction
Q4Verify accuracy of dummy drive mechanism positioning. The purpose is to assure proper obstruction detection by assuring no false positive or false negative obstruction. If dummy drive mechanism is used to measure catheter length for treatment planning, then the action level should be lowered to 1 mm
Q5Accuracy of source drive mechanism to be verified. Autoradiographs or ion‐chamber measurements could be used. If visual checks with in‐room cameras are to be used, source positioning in the cable construction should be verified independently
Q6Comparison of measured source strength with manufacturer supplied value. On installation of a new source, source strength must be measured using calibrated re–entrant chamber and electrometer traceable to a national standards laboratory. The re‐entrant chamber and electrometer should have been calibrated within the last 2 yr. Measured source strength should be used for planning and treatment purposes. Discrepancies greater than 5% between the measured and the manufacturer's supplied source strengths must be investigated. This action level could be lowered to 3%22 if the manufacturer supplied source strength offers such precision. Stability of re‐entrant chamber should be verified prior to use. A second qualified medical physicist should perform a check of the calibration
Q7Visual check on film that the radioactive material is evenly distributed in the encapsulated source. Most important for sources composed of multiple source pellets
Q8Documentation relating to the daily quality control checks, preventive maintenance, service calls, and subsequent checks must be complete, legible, and the operator identified
A1Manual emergency hand crank functionality should be verified with manufacturer service engineer. It is desirable that each person responsible to operate the hand crank, in an emergency situation, practices its operation annually when a dummy wire replaces the source wire
A2Monitor leakage radiation to check afterloader's safe integrity. Intensity of leakage radiation must be lower than the value set by manufacturer and local regulations
A3Verify multichannel indexer functions properly. The wire must be sent to the proper programmed channel
A4Comparison of dwell time accuracy with an external standard performed more rigorously than the treatment day test
A5Verification of the linearity of the timer over a clinically relevant range. The action level represents deviations of measured values from those calculated using a linear fit to the measured data
A6Reproducibility of transit time effect or transit dose effect or source speed between dwell positions. Can be verified using autoradiographs, ion‐chamber measurements, or visual checks with in‐room cameras. A fixed and reproducible applicator geometry is required to assure the same wire drive speed
A7Reusable applicators and transfer guide tubes length should be measured to verify dosimetric lengths if used clinically as nominal values. This measurement could also verify that no debris has come into the lumen
A8Verify physical dimensions of reusable applicators and templates (e.g., diameters, angles, shields). They must match dimensions used in the planning process
A9Verify shielding integrity of shielded applicators. Visual and radiographic inspections should be performed
A10Check x‐ray markers positional accuracy if used clinically for source positioning. If markers are only used to draw the applicator path, then only the integrity should be checked
A11The configuration of this test will depend on the design of the facility and equipment, and local regulations. All staff should review the emergency procedures when a source fails to retract properly and remains outside the safe
A12To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually
A13For PDR only, verify pulse sequencing functionality according to manufacturer's recommendations
  11 in total

Review 1.  Inter-society standards for the performance of brachytherapy: a joint report from ABS, ACMP and ACRO.

Authors:  Subir Nag; Ralph Dobelbower; Glenn Glasgow; Gary Gustafson; Nisar Syed; Bruce Thomadsen; Jeffery F Williamson
Journal:  Crit Rev Oncol Hematol       Date:  2003-10       Impact factor: 6.312

2.  A quality assurance tool for high-dose-rate brachytherapy.

Authors:  Daniel W Rickey; David Sasaki; Jeff Bews
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

3.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

Review 4.  Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; J A Meli; A J Olch; J A Stitt; J F Williamson
Journal:  Med Phys       Date:  1997-10       Impact factor: 4.071

5.  High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59.

Authors:  H D Kubo; G P Glasgow; T D Pethel; B R Thomadsen; J F Williamson
Journal:  Med Phys       Date:  1998-04       Impact factor: 4.071

6.  The transit dose component of high dose rate brachytherapy: direct measurements and clinical implications.

Authors:  K T Bastin; M B Podgorsak; B R Thomadsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  1993-07-15       Impact factor: 7.038

7.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40.

Authors:  G J Kutcher; L Coia; M Gillin; W F Hanson; S Leibel; R J Morton; J R Palta; J A Purdy; L E Reinstein; G K Svensson
Journal:  Med Phys       Date:  1994-04       Impact factor: 4.071

8.  The American College of Radiology and the American Brachytherapy Society practice parameter for the performance of radionuclide-based high-dose-rate brachytherapy.

Authors:  Beth A Erickson; Nathan H J Bittner; Manjeet Chadha; Firas Mourtada; D Jeffrey Demanes
Journal:  Brachytherapy       Date:  2017 Jan - Feb       Impact factor: 2.362

9.  A quality assurance test tool for high dose-rate remote afterloading brachytherapy units.

Authors:  Z Li; T P Mitchell; J R Palta; C Liu
Journal:  Med Phys       Date:  1998-02       Impact factor: 4.071

10.  Production, review, and impact of technical quality control guidelines in a national context.

Authors:  Michelle K Nielsen; Kyle E Malkoske; Erika Brown; Kevin Diamond; Normand Frenière; John Grant; Natalie Pomerleau-Dalcourt; Jason Schella; L John Schreiner; Laurent Tantôt; J Eduardo Villareal-Barajas; Jean-Pierre Bissonnette
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.