| Literature DB >> 29417727 |
Abstract
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control testing of brachytherapy remote afterloaders.Entities:
Keywords: brachytherapy remote afterloaders; quality control guidelines; radiation treatment therapy equipment
Mesh:
Year: 2018 PMID: 29417727 PMCID: PMC5859323 DOI: 10.1002/acm2.12272
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Per treatment quality control tests
| Designator | Test | Performance |
|---|---|---|
| Action | ||
| Per treatment (executed prior to each treatment) | ||
| T1 | Plan data transfer from treatment planning computer | Data integrity |
| T2 | Plan dwell times adjustment | See note |
| T3 | Minimum dwell times | See note |
| T4 | Plan catheters’ connection to remote afterloader | Reproducible |
| T5 | Complete source retraction | Functional |
| Treatment day (or per treatment for PDR) | ||
| D1 | Treatment interrupt | Functional |
| D2 | Console displays (treatment status indicator) and key switch | Functional |
| D3 | Date, time, and source strength in treatment unit | See note |
| D4 | Source (and dummy) positional accuracy | 2 mm |
| D5 | Dwell time accuracy | 2% |
Quarterly quality control tests
| Designator | Test | Performance |
|---|---|---|
| Action | ||
| Quarterly (or at source replacement) | ||
| Q1 | Mechanical integrity of applicators, guide tubes, connectors | Functional |
| Q2 | Internal battery power supply (power failure recovery) | Functional |
| Q3 | Source/dummy interlocks | Functional |
| Q4 | Dummy wire positional accuracy | 3 mm (1 mm see note) |
| Q5 | Radiological source positional accuracy | 1 mm |
| Q6 | Source strength calibration | 5% |
| Q7 | Source homogeneity | Reproducible |
| Q8 | Records | Complete |
Annual quality control tests
| Designator | Test | Performance |
|---|---|---|
| Action | ||
| Annually | ||
| A1 | Hand crank operation | Functional |
| A2 | Leakage radiation | Reproducible |
| A3 | Multi‐channel indexer function | Functional |
| A4 | Dwell time accuracy | 1% |
| A5 | Timer linearity | 1% |
| A6 | Transit time/transit dose reproducibility | Reproducible |
| A7 | Dosimetric length of applicators and guide tubes | 1 mm |
| A8 | Applicators and templates dimensions | Reproducible |
| A9 | Shield integrity of shielded applicators | Reproducible |
| A10 | X ray marker positional accuracy | 1 mm |
| A11 | Document staff review of emergency response procedures | Complete |
| A12 | Independent quality control review | Complete |
| A13 | PDR sequencing (for PDR only) | Functional |
| T1 | Plan data imported from a treatment planning system into the treatment console should be verified for source strength, dwell positions, and dwell times. In case of many dwell positions, the verification of a sub‐set of positions is acceptable. For a plan already present in the treatment console, the same verification should be made to assure proper plan selection |
| T2 | Plan dwell times adjustment by the treatment console for the treatment date should be verified (by an independent calculation: hand calculation, decay factor chart, or software calculation). Action level will depend on the treatment console decay frequency and time resolution; express in percentage of difference and/or in second |
| T3 | Minimum dwell time should be verified against the device driven limits. Those limits should take into account both the effect of transit dose and positioning reproducibility. Some remote afterloader systems might have a positioning reproducibility dependence on dwell time. All dwell times should be equal or greater than set limit |
| T4 | Catheters/applicators connections to remote afterloader indexer channels must match plan |
| T5 | Survey the treatment room and patient to ensure that source has been completely retracted |
| D1 | During source exposure, verify that the non‐emergency interrupt button (if equipped) retracts the source to its safe and shielded position |
| D2 | On the treatment console, displays should be verified. At minimum, treatment status indicators should be verified by exposing a source. Indicators could be visual and audible. When a treatment key is available, its deactivation should prevent source exposition |
| D3 | Remote afterloader console date and time are properly set. Decayed source strength is accurate compared to an independent calculation (hand calculation, decay factor chart or software calculation), taking into account treatment console decay frequency |
| D4 | Verify accuracy of source drive mechanism positioning. A visual inspection with a camera is acceptable. Apply also to dummy drive mechanism if used to measure catheter length |
| D5 | Comparison of dwell time accuracy with external standard such as a stopwatch. The dwell time used should be sufficiently long such that errors in the measurement of the time (e.g., reaction time of the observer) are less than 1% |
| Q1 | Verify the applicators, guide tubes, and connectors are exempt of damage (excessive wear, kinks, etc.) |
| Q2 | The configuration of this test will depend on the design of the facility and equipment. Safety is the concern and tests should be designed accordingly. The first objective is to verify that the equipment safely retracts the source wire after a power failure. The second objective is to verify that the equipment properly records treatment delivered before power failure and permits to resume the treatment after power recovery |
| Q3 | Verify functionality of remote afterloader interlocks related to source and dummy wires. This includes incorrect connection of applicator to transfer guide tube, incorrect connection of transfer guide tube to remote afterloader, and obstruction |
| Q4 | Verify accuracy of dummy drive mechanism positioning. The purpose is to assure proper obstruction detection by assuring no false positive or false negative obstruction. If dummy drive mechanism is used to measure catheter length for treatment planning, then the action level should be lowered to 1 mm |
| Q5 | Accuracy of source drive mechanism to be verified. Autoradiographs or ion‐chamber measurements could be used. If visual checks with in‐room cameras are to be used, source positioning in the cable construction should be verified independently |
| Q6 | Comparison of measured source strength with manufacturer supplied value. On installation of a new source, source strength must be measured using calibrated re–entrant chamber and electrometer traceable to a national standards laboratory. The re‐entrant chamber and electrometer should have been calibrated within the last 2 yr. Measured source strength should be used for planning and treatment purposes. Discrepancies greater than 5% between the measured and the manufacturer's supplied source strengths must be investigated. This action level could be lowered to 3% |
| Q7 | Visual check on film that the radioactive material is evenly distributed in the encapsulated source. Most important for sources composed of multiple source pellets |
| Q8 | Documentation relating to the daily quality control checks, preventive maintenance, service calls, and subsequent checks must be complete, legible, and the operator identified |
| A1 | Manual emergency hand crank functionality should be verified with manufacturer service engineer. It is desirable that each person responsible to operate the hand crank, in an emergency situation, practices its operation annually when a dummy wire replaces the source wire |
| A2 | Monitor leakage radiation to check afterloader's safe integrity. Intensity of leakage radiation must be lower than the value set by manufacturer and local regulations |
| A3 | Verify multichannel indexer functions properly. The wire must be sent to the proper programmed channel |
| A4 | Comparison of dwell time accuracy with an external standard performed more rigorously than the treatment day test |
| A5 | Verification of the linearity of the timer over a clinically relevant range. The action level represents deviations of measured values from those calculated using a linear fit to the measured data |
| A6 | Reproducibility of transit time effect or transit dose effect or source speed between dwell positions. Can be verified using autoradiographs, ion‐chamber measurements, or visual checks with in‐room cameras. A fixed and reproducible applicator geometry is required to assure the same wire drive speed |
| A7 | Reusable applicators and transfer guide tubes length should be measured to verify dosimetric lengths if used clinically as nominal values. This measurement could also verify that no debris has come into the lumen |
| A8 | Verify physical dimensions of reusable applicators and templates (e.g., diameters, angles, shields). They must match dimensions used in the planning process |
| A9 | Verify shielding integrity of shielded applicators. Visual and radiographic inspections should be performed |
| A10 | Check x‐ray markers positional accuracy if used clinically for source positioning. If markers are only used to draw the applicator path, then only the integrity should be checked |
| A11 | The configuration of this test will depend on the design of the facility and equipment, and local regulations. All staff should review the emergency procedures when a source fails to retract properly and remains outside the safe |
| A12 | To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually |
| A13 | For PDR only, verify pulse sequencing functionality according to manufacturer's recommendations |