| Literature DB >> 35743696 |
Antonio Herreros1,2, José Pérez-Calatayud3,4, Facundo Ballester5, Jose Barrera-Gómez6, Rosa Abellana1, Joana Melo7, Luis Moutinho7, Luca Tagliaferri8, Ángeles Rovirosa1,2,9.
Abstract
(1) Background: In brachytherapy, there are still many manual procedures that can cause adverse events which can be detected with in vivo dosimetry systems. Plastic scintillator dosimeters (PSD) have interesting properties to achieve this objective such as real-time reading, linearity, repeatability, and small size to fit inside brachytherapy catheters. The purpose of this study was to evaluate the performance of a PSD in postoperative endometrial brachytherapy in terms of source dwell time accuracy. (2)Entities:
Keywords: brachytherapy; endometrial carcinoma; in vivo dosimetry; plastic scintillator dosimeter; treatment verification
Year: 2022 PMID: 35743696 PMCID: PMC9224704 DOI: 10.3390/jpm12060911
Source DB: PubMed Journal: J Pers Med ISSN: 2075-4426
Summary of the results of the repeatability test.
| Detector | Nominal Dwell | Time Average Deviation | Maximum | RSD 1 | RSD 1 | RSD 1 Measured Dwell Time |
|---|---|---|---|---|---|---|
| 10 | 300 | −0.04 | −0.10 | 0.27 | 1.17 | 0.02 |
| 30 | −0.02 | −0.10 | 0.22 | 0.29 | 0.14 | |
| 10 | −0.02 | −0.20 | 0.50 | 1.38 | 0.79 | |
| 20 | 10 | 0.00 | −0.10 | 0.32 | 5.59 | 0.58 |
| 3 | 0.06 | 0.15 | 0.43 | 2.45 | 1.89 | |
| 1 | 0.04 | 0.10 | 1.03 | 9.22 | 4.19 |
1 RSD: relative standard deviation.
Figure 1Absorbed dose rate sensitivity percent deviation as a function of absorbed dose rate. Sensitivity is the quotient between PSD count rate and the absorbed dose rate. The black dots indicate the mean value of three consecutive measurements. The error bars are based on standard deviation.
Figure 2Illustration of the noise in the background and noise in the signal.
Figure 3SNR (signal-to-noise ratio); SNR has no units because signal and noise units are the same.
Figure 4Linear regression of the PSD count rate with the absorbed dose rate.
Figure 5Histogram of the absolute deviations between planned and measured dwell times. The binning is 0.1 s based on the PRO-DOSE acquisition frequency used.
Figure 6(a) An example of six dwell position signals clearly differentiated; (b) an example with two dwell positions signals undistinguished due to their similar count rate.
Figure 7Box plot of the dwell time absolute deviations in each dwell position; #1 is the most distal dwell position, near the tip of the applicator. The step size used in all patients was 2.5 mm, with an inactive source position between active dwell positions: 1-3-5-7-9-11. As only 1 treatment plan used 7 dwell positions (1-3-5-7-9-11-13), the corresponding position number 13 was not included in this box plot. The 6 black dots are outliers: values outside 1.5 times the interquartile range above the upper quartile and below the lower quartile. The average values in each of the 6 dwell positions are represented by the crosses and are from left-to-right: 0.070 s, −0.100 s, −0.003 s, 0.020 s, 0.005 s, and 0.012 s, and the corresponding medians are: 0.100 s, −0.035 s, 0.000 s, 0.000 s, 0.000 s, and 0.060 s.