Effect of Milliamperage Reduction on Pre-surgical Implant Planning Using Cone Beam Computed Tomography by Surgeons of Varying Experience.

N El Sahili1, S David-Tchouda2, S Thoret3, I Nasseh4, A Berberi5, T Fortin6.   

Abstract

BACKGROUND: Differences in CBCT units and the lack of standardization result in exposure to radiation doses beyond what is required for diagnostic purposes, especially when planning the surgical placement of dental implants. AIM: To assess the influence of low- and high-dose milliamperage settings on CBCT images for objective and subjective implant planning among senior specialists (5 years of experience) and juniors (fresh graduates).
MATERIALS AND METHODS: Two dry skulls (4 hemi-maxillary segments of the maxilla and 4 hemi-maxillary segments of the mandible) were scanned under low (2 mA) and high (6.3 mA) dosage settings using the Carestream CS 9300 machine. Cross-sectional slices of both image qualities were evaluated by the 5 seniors and the 5 juniors for subjective image utility for implant planning and for objective linear bone measurements.
RESULTS: There were no significant differences in bone measurements taken on high- or low-dose images by all seniors and by the majority of juniors (p > 0.05). In qualitative image assessments, there was independence between assessment and image quality for almost all observers. For planning posterior mandibular implant placement, increased dosage improved concordance and kappa values between low- and high-dose images for senior observers (from K = 0.287 at low dose to K = 0.718 at high does) but not for juniors (K = 0.661 and K = 0.509 for low and high dose, respectively).
CONCLUSION: Reduction in milliamperage did not affect diagnostic image quality for objective bone measurements and produced sufficient concordance for qualitative assessment. Judicious optimization of milliamperage settings based on individual diagnostic requirements can result in significant dose reduction without compromising diagnostic decision-making.

Entities:  

Keywords:  Cone beam computed tomography; Image resolution; Implant-placement planning; Radiation dose alteration

Year:  2017        PMID: 30344396      PMCID: PMC6181865          DOI: 10.1007/s12663-017-1075-y

Source DB:  PubMed          Journal:  J Maxillofac Oral Surg        ISSN: 0972-8270


  22 in total

1.  Precision of transfer of preoperative planning for oral implants based on cone-beam CT-scan images through a robotic drilling machine.

Authors:  Thomas Fortin; Guillaume Champleboux; Silvio Bianchi; Hervé Buatois; Jean-Loup Coudert
Journal:  Clin Oral Implants Res       Date:  2002-12       Impact factor: 5.977

2.  Effects of tube current on cone-beam computerized tomography image quality for presurgical implant planning in vitro.

Authors:  Jaideep Sur; Kenji Seki; Hiroshi Koizumi; Koh Nakajima; Tomohiro Okano
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2010-07-02

3.  Morphological concordance between CBCT and MDCT: a paranasal sinus-imaging anatomical study.

Authors:  J Demeslay; S Vergez; E Serrano; P Chaynes; P Cantet; B Chaput; G de Bonnecaze
Journal:  Surg Radiol Anat       Date:  2015-08-04       Impact factor: 1.246

4.  Effect of exposure time on the accuracy and reliability of cone beam computed tomography in the assessment of dental implant site dimensions in dry skulls.

Authors:  Asmáa A Al-Ekrish
Journal:  Saudi Dent J       Date:  2012-06-01

5.  Technical assessment of a cone-beam CT scanner for otolaryngology imaging: image quality, dose, and technique protocols.

Authors:  J Xu; D D Reh; J P Carey; M Mahesh; J H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

6.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results.

Authors:  P Mozzo; C Procacci; A Tacconi; P T Martini; I A Andreis
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

7.  Determination of size-specific exposure settings in dental cone-beam CT.

Authors:  Ruben Pauwels; Reinhilde Jacobs; Ria Bogaerts; Hilde Bosmans; Soontra Panmekiate
Journal:  Eur Radiol       Date:  2016-04-23       Impact factor: 5.315

8.  Influence of the milliamperage settings on cone beam computed tomography imaging for implant planning.

Authors:  Taruska Ventorini Vasconcelos; Frederico Sampaio Neves; Deborah Queiroz de Freitas; Paulo Sérgio Flores Campos; Plauto Christopher Aranha Watanabe
Journal:  Int J Oral Maxillofac Implants       Date:  2014-09-26       Impact factor: 2.804

9.  Inter- and intraobserver reproducibility of buccal bone measurements at dental implants with cone beam computed tomography in the esthetic region.

Authors:  Kirsten W Slagter; Gerry M Raghoebar; Arjan Vissink; Henny J A Meijer
Journal:  Int J Implant Dent       Date:  2015-04-18

10.  Cone-beam computed tomography: Time to move from ALARA to ALADA.

Authors:  Prashant P Jaju; Sushma P Jaju
Journal:  Imaging Sci Dent       Date:  2015-12-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.