| Literature DB >> 20586706 |
Truike M Thien1, Johan Kärrholm.
Abstract
BACKGROUND ANDEntities:
Mesh:
Year: 2010 PMID: 20586706 PMCID: PMC2917561 DOI: 10.3109/17453674.2010.501739
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Implant-related parameters for the polished Exeter stem entered into the statistical analysis
| Classification/Name | Number |
|---|---|
| Size | |
| light weight–0 | 2,781 |
| light weight–1 | 8,019 |
| narrow | 6,205 |
| standard | 2,860 |
| heavy duty | 1,013 |
| ultra heavy duty | 130 |
| Offset (mm) | |
| 37.5 | 4,951 |
| 44 | 15,416 |
| 50 | 641 |
| Taper | |
| old | 5,775 |
| V40 | 15,233 |
| Material | |
| standard | 5,751 |
| alumina ceramic | 630 |
| Orthinox | 14,627 |
| Length | |
| short | 3,707 |
| medium | 14,671 |
| long | 2,542 |
| x-long | 88 |
| Diameter (mm) | |
| 22 | 1,567 |
| 26 | 443 |
| 28 | 18,750 |
| 30 | 248 |
Implant-related parameters for the Spectron EF Primary stem entered into the statistical analysis. All implants were supplied with 28-mm heads made of cobalt-chromium alloy
| Classification/Name | Number |
|---|---|
| Size | |
| 1 | 1,101 |
| 2 | 3,501 |
| 3 | 1,987 |
| 4+5 | 551 |
| Offset | |
| standard (s-off) | 4,976 |
| h-offset (h-off) | 2,164 |
| Length | |
| short or standard (s, -3, 0) | 3,847 |
| long (l, +4) | 2,264 |
| x-long (xl, +8) | 822 |
| x-long+ (xxl, +12 or xxxl, +16) | 207 |
| Offset + neck | |
| 1: s-off or h-off + s neck | 2,696 |
| 2: s-off + l neck or h-off + s neck | 2,716 |
| 3: s-off + xl neck or h-off + l neck | 1,275 |
| 4: s-off + xx l neck or h-off + xl neck | 362 |
| 5: s-off + xxxl neck or h-off + xxl neck or xxxl neck | 91 |
Neck length as extracted from the database after assembly of the stem, coded into 5 classes. A higher number means a longer neck.
Relative risks of aseptic stem revision during entire observation period (Lubinus SP, Spectron EF Primary) or within 3 years (Exeter) and 95% CI for implant-related parameters for each stem type. Data from Cox regression adjusted for age, sex, diagnosis, bilaterality, side, and implant-specific parameters (see text). Implant-related parameters with statistically significant influence are shown
| Relative Risk | 95% CI | p-value | ||
|---|---|---|---|---|
| [Exp (B)] | Lower | Upper | ||
| 22 mm | 2.37 | 1.32 | 4.23 | 0.004 |
| 26 mm | 0.02 | 0.00 | 30.03 | 0.3 |
| 28 mm, reference | 1.00 | |||
| 30 mm | 0.79 | 0.19 | 3.28 | 0.7 |
| x-small | 1.72 | 1.22 | 2.44 | 0.002 |
| small | 0.99 | 0.74 | 1.32 | 0.9 |
| medium, reference | 1.00 | |||
| large | 0.94 | 0.70 | 1.26 | 0.7 |
| x and xx-large | 1.01 | 0.69 | 1.47 | 1.0 |
| ceramic | 0.52 | 0.36 | 0.76 | 0.001 |
| cobalt-chromium, reference | 1.00 | |||
| 1 | 0.73 | 0.55 | 0.97 | 0.03 |
| 2 reference | 1.00 | |||
| 3 | 1.11 | 0.86 | 1.43 | 0.4 |
| 4 | 1.72 | 1.09 | 2.71 | 0.02 |
| 1 | 2.23 | 1.25 | 3.97 | 0.006 |
| 2 reference | 1.00 | |||
| 3 | 0.75 | 0.40 | 1.41 | 0.4 |
| 4+5 | 0.55 | 0.17 | 1.84 | 0.3 |
| 1 | 1.27 | 0.68 | 2.37 | 0.4 |
| 2 reference | 1.00 | |||
| 3 | 1.54 | 0.80 | 2.98 | 0.2 |
| 4 | 2.03 | 0.81 | 5.11 | 0.1 |
| 5 | 6.89 | 2.54 | 18.63 | < 0.001 |
Revisions for all the femoral stems analyzed (infections excluded)
| Exeter Polished (All) | Exeter (old) | Exeter V40 | Lubinus SPII | Spectron EF | |
|---|---|---|---|---|---|
| Primary | |||||
| % (n) | % (n) | % (n) | % (n) | % (n) | |
| Aseptic Loosening | 0.4 (96) | 1.1 (71) | 0.2 (25) | 0.3 (114) | 0.8 (57) |
| Dislocation | 0.5 (116) | 0.6 (39) | 0.5 (77) | 0.5 (244) | 0.5 (35) |
| Fracture | 0.2 (43) | 0.3 (18) | 0.2 (25) | 0.0 (17) | 0.0 (3) |
| Implant failure | 0.0 (3) | 0.0 (0) | 0.0 (3) | 0.0 (8) | 0.0 (0) |
| Technical reason | 0.1 (12) | 0.1 (7) | 0.0 (5) | 0.0 (12) | 0.1 (4) |
| Pain only | 0.0 (3) | 0.0 (0) | 0.0 (3) | 0.0 (5) | 0.0 (2) |
| Total | 1.2 (273) | 2.1 (135) | 1.1 (138) | 0.8 (400) | 1.4 (101) |
Number of revisions and causes of revision for Lubinus SPII stems with either cobalt-chromium or aluminium oxide femoral heads (infections excluded)
| Cobalt-chromium | Aluminium oxide | |
|---|---|---|
| Aseptic loosening | 29 (104) | 24 (9) |
| Dislocation | 61 (222) | 53 (20) |
| Fracture | 4.4 (16) | 3 (1) |
| Implant failure | 1.1 (4) | 11 (4) |
| Technical reason | 2.5 (9) | 8 (3) |
| Pain only | 1.1 (4) | 3 (1) |
| Other mechanical reason | 1.1 (4) | 0 (0) |
| Total | 100 (363) | 100 (38) |
Implant-related parameters for the Lubinus SPII stem entered into the statistical analysis. All implants were supplied with 28-mm heads
| Classification/Name | Number |
|---|---|
| Size | |
| x-small | 4,445 |
| small | 11,158 |
| medium | 14,187 |
| large | 9,350 |
| x and xx-large | 3,896 |
| CCD | |
| 117 | 1,388 |
| 126 | 36,179 |
| 135 | 5,469 |
| Offset | |
| standard (s-off) | 41,276 |
| x-offset (x-off) | 1,760 |
| Material | |
| standard | 36,802 |
| alumina ceramic | 6,234 |
| Length | |
| short (s, 46) | 12,538 |
| medium (m, 49) | 19,797 |
| long (l, 53) | 9,833 |
| x-long (xl, 60) | 868 |
| Offset + neck | |
| 1: s-off + s neck | 11,727 |
| 2: s-off + m neck or x-off + s neck | 19,921 |
| 3: s-off + l neck or x-off + m neck | 10,290 |
| 4: s-off + xl neck or x-off + l neck or xl neck | 1,098 |
Neck length as extracted from the database after assembly of the stem, coded into 4 classes. A higher number means a longer neck. Class 4 included 32 implants with combined x-off and xl-neck.