Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry.

Yuehua Li1, Jingming Shi, Xuanchu Duan, Fang Fan.   

Abstract

BACKGROUND: Diaton transpalpebral tonometry is a new version of TGDc-01 transpalpebral tonometry. The aim of this study is to evaluate the reliability of this method, and to compare the results with those of corrected Goldmann applanation tonometry (c-GAT), which is the "gold standard" for the clinical measurement of IOP.
METHODS: IOP was measured in 212 eyes of 129 subjects by means of two different methods. Measurements with two tonometers were performed in random order, and the measurements of central cornea thickness were performed with ultrasound pachymetry.
RESULTS: The average IOP reading was 15.46 ± 5.68 mmHg for c-GAT and 13.84 ± 5.12 mmHg for Diaton tonometer. The mean difference between c-GAT and Diaton tonometer was 1.62 ± 3.60 mmHg (p < 0.05). Compared with c-GAT, 39.6% of IOP readings (84 eyes) by Diaton tonometer was in an interval of ± 1 mmHg; 59.4% (126 eyes) was in an interval of ± 2 mmHg, and 76.4% (162 eyes) was in an interval of ± 3 mmHg. For the subjects whose age was from 20 to 50, the mean difference of IOP readings by two tonometers was 0.53 ± 3.24 mmHg.
CONCLUSIONS: We cannot recommend Diaton as a substitute or alternative method for GAT for diagnosis and follow-up of patients with abnormal IOPs, but it may be helpful as a screening tool, especially for subjects whose age is between 20 and 50 years and for healthy subjects.

Entities:  

Mesh:

Year:  2010        PMID: 20495818     DOI: 10.1007/s00417-009-1243-y

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  11 in total

1.  Applanation tonometry and central corneal thickness.

Authors:  N Ehlers; T Bramsen; S Sperling
Journal:  Acta Ophthalmol (Copenh)       Date:  1975-03

2.  The Advanced Glaucoma Intervention Study (AGIS): 7. The relationship between control of intraocular pressure and visual field deterioration.The AGIS Investigators.

Authors: 
Journal:  Am J Ophthalmol       Date:  2000-10       Impact factor: 5.258

3.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

4.  Measurement of the intraocular pressure with the "transpalpebral tonometer" TGDc-01 in comparison with applanation tonometry.

Authors:  Dirk Sandner; Andreas Böhm; Sandra Kostov; Lutz Pillunat
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-12-22       Impact factor: 3.117

5.  Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry.

Authors:  Markus Kohlhaas; Andreas G Boehm; Eberhard Spoerl; Antje Pürsten; Hans J Grein; Lutz E Pillunat
Journal:  Arch Ophthalmol       Date:  2006-04

6.  Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers.

Authors:  A Troost; S H Yun; K Specht; F Krummenauer; O Schwenn
Journal:  Br J Ophthalmol       Date:  2005-03       Impact factor: 4.638

Review 7.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry.

Authors:  Annette Lösch; Alexander Scheuerle; Volker Rupp; Gerd Auffarth; Matthias Becker
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-10-15       Impact factor: 3.117

10.  Central corneal thickness and Diaton transpalpebral tonometry.

Authors:  Mustafa Ilker Toker; Ayse Vural; Haydar Erdogan; Aysen Topalkara; Mustafa Kemal Arici
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-02-20       Impact factor: 3.117

View more
  6 in total

1.  [Glaucoma attack following Hughes tarsoconjunctival flap: what should be done?].

Authors:  E Bertelmann; N Minko; N Torun
Journal:  Ophthalmologe       Date:  2012-09       Impact factor: 1.059

2.  Comparison of Diaton transpalpebral tonometer with applanation tonometry in keratoconus.

Authors:  Robert Pl Wisse; Natalie Peeters; Saskia M Imhof; Allegonda van der Lelij
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

3.  Change in intraocular pressure during point-of-care ultrasound.

Authors:  Cameron Berg; Stephanie J Doniger; Brita Zaia; Sarah R Williams
Journal:  West J Emerg Med       Date:  2015-03-06

4.  Effects of Scleral Contact Lenses for Keratoconus Management on Visual Quality and Intraocular Pressure.

Authors:  Martina Formisano; Federica Franzone; Ludovico Alisi; Santino Pistella; Leopoldo Spadea
Journal:  Ther Clin Risk Manag       Date:  2021-01-25       Impact factor: 2.423

5.  Comparison of Intraocular Pressure Measurement with Schiotz Tonometer and Tono-Pen Vet Tonometer in Healthy Dogs.

Authors:  Karolina Wrześniewska; Jacek Madany; Dagmara Winiarczyk
Journal:  J Vet Res       Date:  2018-10-24       Impact factor: 1.744

6.  Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear.

Authors:  Jennifer Swingle Fogt; Cherie B Nau; Muriel Schornack; Ellen Shorter; Amy Nau; Jennifer S Harthan
Journal:  Optom Vis Sci       Date:  2020-09       Impact factor: 2.106

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.