Measurement of the intraocular pressure with the "transpalpebral tonometer" TGDc-01 in comparison with applanation tonometry.

Dirk Sandner1, Andreas Böhm, Sandra Kostov, Lutz Pillunat.   

Abstract

BACKGROUND: An irregular corneal surface compromises IOP measurement by Goldmann applanation tonometry. In such cases accurate measurement without corneal contact would be desirable. The new eyelid tonometer TGDc-01 measures IOP without corneal contact through the eyelid. The aim of the study was to evaluate the accuracy of the eyelid tonometer compared with Goldmann applanation tonometry (cornea thickness-corrected values) in subjects without corneal alterations.
METHODS: IOP was measured in 199 eyes of 103 subjects without corneal alterations by means of two different methods. Measurements with the transpalpebral tonometer TGDc-01 and the Goldmann applanation tonometer were performed within 5 min in random order.
RESULTS: The mean difference between lid tonometry and Goldmann applanation tonometry was 0.71 mmHg, SD +/-2.467 mmHg. In the reliability analysis the intraclass correlation coefficient was 0.8620. Compared with Goldmann applanation tonometry 66.4% of the IOP readings measured by lid tonometry were in an interval of +/-2 mmHg, 81.0% in an interval of +/-3 mmHg. The maximum of deviation was -6 mmHg and +6 mmHg, respectively. The Bland and Altman plots are shown.
CONCLUSIONS: Lid tonometry correlates sufficiently with Goldmann applanation tonometry, but in more than 10% of the measurements the IOP readings differed by more than 3 mmHg. The eyelid tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable.

Entities:  

Mesh:

Year:  2004        PMID: 15614540     DOI: 10.1007/s00417-004-1037-1

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  14 in total

1.  The Advanced Glaucoma Intervention Study (AGIS): 7. The relationship between control of intraocular pressure and visual field deterioration.The AGIS Investigators.

Authors: 
Journal:  Am J Ophthalmol       Date:  2000-10       Impact factor: 5.258

2.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

3.  [A new applanation tonometer].

Authors:  H GOLDMANN
Journal:  Bull Mem Soc Fr Ophtalmol       Date:  1954

4.  Blood and plasma viscosity measurements in patients with glaucoma.

Authors:  J H Klaver; E L Greve; H Goslinga; H C Geijssen; J H Heuvelmans
Journal:  Br J Ophthalmol       Date:  1985-10       Impact factor: 4.638

5.  Color Doppler analysis of ocular vessel blood velocity in normal-tension glaucoma.

Authors:  A Harris; R C Sergott; G L Spaeth; J L Katz; J A Shoemaker; B J Martin
Journal:  Am J Ophthalmol       Date:  1994-11-15       Impact factor: 5.258

6.  Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Collaborative Normal-Tension Glaucoma Study Group.

Authors: 
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

7.  The effectiveness of intraocular pressure reduction in the treatment of normal-tension glaucoma. Collaborative Normal-Tension Glaucoma Study Group.

Authors: 
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

8.  Retinal autoregulation in open-angle glaucoma.

Authors:  J E Grunwald; C E Riva; R A Stone; E U Keates; B L Petrig
Journal:  Ophthalmology       Date:  1984-12       Impact factor: 12.079

9.  Pressure compliance of the optic nerve head in low tension glaucoma.

Authors:  L E Pillunat; R Stodtmeister; I Wilmanns
Journal:  Br J Ophthalmol       Date:  1987-03       Impact factor: 4.638

10.  Risk factors for primary open angle glaucoma.

Authors:  J Katz; A Sommer
Journal:  Am J Prev Med       Date:  1988 Mar-Apr       Impact factor: 5.043

View more
  6 in total

1.  Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry.

Authors:  Yuehua Li; Jingming Shi; Xuanchu Duan; Fang Fan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-05-22       Impact factor: 3.117

2.  Evaluation of the impedance tonometers TGDc-01 and iCare according to the international ocular tonometer standards ISO 8612.

Authors:  Peter Christian Ruokonen; Thomas Schwenteck; Jörg Draeger
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-02-21       Impact factor: 3.117

3.  Central corneal thickness and Diaton transpalpebral tonometry.

Authors:  Mustafa Ilker Toker; Ayse Vural; Haydar Erdogan; Aysen Topalkara; Mustafa Kemal Arici
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-02-20       Impact factor: 3.117

4.  Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry.

Authors:  Isabel Cacho; Juan Sanchez-Naves; Laura Batres; Jesús Pintor; Gonzalo Carracedo
Journal:  J Ophthalmol       Date:  2015-06-08       Impact factor: 1.909

5.  Effects of Scleral Contact Lenses for Keratoconus Management on Visual Quality and Intraocular Pressure.

Authors:  Martina Formisano; Federica Franzone; Ludovico Alisi; Santino Pistella; Leopoldo Spadea
Journal:  Ther Clin Risk Manag       Date:  2021-01-25       Impact factor: 2.423

6.  Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear.

Authors:  Jennifer Swingle Fogt; Cherie B Nau; Muriel Schornack; Ellen Shorter; Amy Nau; Jennifer S Harthan
Journal:  Optom Vis Sci       Date:  2020-09       Impact factor: 2.106

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.