Comparison of Diaton transpalpebral tonometer with applanation tonometry in keratoconus.

Robert Pl Wisse1, Natalie Peeters1, Saskia M Imhof1, Allegonda van der Lelij1.   

Abstract

AIM: To investigate the added value of using a Diaton transpalpebral tonometer (DT) to measure IOP in keratoconus. Most type of tonometers use corneal applanation or biomechanical resistance to measure intraocular pressure (IOP); however, these factors can be altered by keratoconus. Specifically, we examined whether DT can detect false-negative low Goldmann applanation tonometry (AT) measurements.
METHODS: Patients with keratoconus were recruited from our tertiary academic treatment center. Measurements included AT and DT (in random order) and Scheimpflug imaging. An age- and gender-matched group of control subjects with no history of corneal disease or glaucoma was also recruited.
RESULTS: In total, 130 eyes from 66 participants were assessed. In the keratoconus group, mean AT was 11.0 ± 2.6, mean DT 11.2±5.5 (P=0.729), and the two measures were correlated significantly (P=0.006, R=0.323). However, a Bland-Altman plot revealed a wide distribution and poor agreement between both measurements. Previous corneal crosslinking, corneal pachymetry, and Krumeich classification had no effect on measured IOP.
CONCLUSION: Measurements obtained using a Diaton tonometer are not affected by corneal biomechanics; however, its poor agreement with Goldmann AT values calls into question the added value of using a Diaton tonometer to measure IOP in keratoconus.

Entities:  

Keywords:  Bland-Altman plot; Diaton; Goldmann applanation tonometry; keratoconus; transpalpebral tonometry

Year:  2016        PMID: 27158609      PMCID: PMC4844034          DOI: 10.18240/ijo.2016.03.12

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  22 in total

1.  Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry.

Authors:  Yuehua Li; Jingming Shi; Xuanchu Duan; Fang Fan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-05-22       Impact factor: 3.117

2.  Combining ocular response analyzer metrics for corneal biomechanical diagnosis.

Authors:  Pablo R Ruiseñor Vázquez; Marianella Delrivo; Fernando Fuentes Bonthoux; Tomás Pförtner; Jeremías Gastón Galletti
Journal:  J Refract Surg       Date:  2013-07-17       Impact factor: 3.573

3.  [Intraocular pressure difference in Goldmann applanation tonometry versus a transpalpebral tonometer TGDc-01"PRA" in glaucoma patients].

Authors:  T Schlote; H Landenberger
Journal:  Klin Monbl Augenheilkd       Date:  2005-02       Impact factor: 0.700

4.  Live-epikeratophakia for keratoconus.

Authors:  J H Krumeich; J Daniel; A Knülle
Journal:  J Cataract Refract Surg       Date:  1998-04       Impact factor: 3.351

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Assessment of intraocular pressure measured by Reichert Ocular Response Analyzer, Goldmann Applanation Tonometry, and Dynamic Contour Tonometry in healthy individuals.

Authors:  Ping-Bo Ouyang; Cong-Yi Li; Xiao-Hua Zhu; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-02-18       Impact factor: 1.779

7.  Second-harmonic imaging microscopy of normal human and keratoconus cornea.

Authors:  Naoyuki Morishige; Andrew J Wahlert; M Cristina Kenney; Donald J Brown; Koji Kawamoto; Tai-Ichiro Chikama; Teruo Nishida; James V Jester
Journal:  Invest Ophthalmol Vis Sci       Date:  2007-03       Impact factor: 4.799

8.  Intraocular pressure measurement precision with the Goldmann applanation, dynamic contour, and ocular response analyzer tonometers.

Authors:  Aachal Kotecha; Edward White; Patricio G Schlottmann; David F Garway-Heath
Journal:  Ophthalmology       Date:  2010-02-01       Impact factor: 12.079

9.  Central corneal thickness and Diaton transpalpebral tonometry.

Authors:  Mustafa Ilker Toker; Ayse Vural; Haydar Erdogan; Aysen Topalkara; Mustafa Kemal Arici
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-02-20       Impact factor: 3.117

10.  [Influence of corneal thickness in keratoconic corneas on IOP measurement with IOPen, iCare, dynamic contour tonometry and Goldmann applanation tonometry].

Authors:  M K J Klamann; A-K B Maier; J Gonnermann; P Ruokonen; E Bertelmann; N Torun
Journal:  Klin Monbl Augenheilkd       Date:  2013-05-13       Impact factor: 0.700

View more
  1 in total

1.  Effects of Scleral Contact Lenses for Keratoconus Management on Visual Quality and Intraocular Pressure.

Authors:  Martina Formisano; Federica Franzone; Ludovico Alisi; Santino Pistella; Leopoldo Spadea
Journal:  Ther Clin Risk Manag       Date:  2021-01-25       Impact factor: 2.423

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.