Literature DB >> 17610285

Tracking planar orientations of active MRI needles.

Shashank Sathyanarayana1, Pelin Aksit, Aravind Arepally, Parag V Karmarkar, Meiyappan Solaiyappan, Ergin Atalar.   

Abstract

PURPOSE: To determine and track the planar orientation of active interventional devices without using localizing RF microcoils.
MATERIALS AND METHODS: An image-based tracking method that determines a device's orientation using projection images was developed. An automated and a manual detection scheme were implemented. The method was demonstrated in an in vivo mesocaval puncture procedure in swine, which required accurate orientation of an active transvascular needle catheter.
RESULTS: The plane of the catheter was determined using two projection images. The scan plane was adjusted automatically to follow the catheter plane, and its orientation with respect to a previously acquired target plane was displayed. The algorithm facilitated navigation for a fast and accurate puncture.
CONCLUSION: Using image-based techniques, with no mechanical design changes, the orientation of an active intravascular probe could be tracked. (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17610285     DOI: 10.1002/jmri.20960

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  3 in total

1.  MRI endoscopy using intrinsically localized probes.

Authors:  Shashank Sathyanarayana; Paul A Bottomley
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

2.  Visualization of active devices and automatic slice repositioning ("SnapTo") for MRI-guided interventions.

Authors:  Ashvin K George; J Andrew Derbyshire; Haris Saybasili; Christina E Saikus; Ozgur Kocaturk; Michael A Guttman; Elliot R McVeigh; Robert J Lederman; Anthony Z Faranesh
Journal:  Magn Reson Med       Date:  2010-04       Impact factor: 4.668

3.  Whole shaft visibility and mechanical performance for active MR catheters using copper-nitinol braided polymer tubes.

Authors:  Ozgur Kocaturk; Christina E Saikus; Michael A Guttman; Anthony Z Faranesh; Kanishka Ratnayaka; Cengizhan Ozturk; Elliot R McVeigh; Robert J Lederman
Journal:  J Cardiovasc Magn Reson       Date:  2009-08-12       Impact factor: 5.364