Development of a traceable calibration methodology for solid (68)Ge/(68)Ga sources used as a calibration surrogate for (18)F in radionuclide activity calibrators.

Brian E Zimmerman1, Jeffrey T Cessna.   

Abstract

UNLABELLED: We have developed a methodology for calibrating (68)Ge radioactivity content in a commercially available calibration source for activity calibrators in a way that is traceable to the national standard. Additionally, the source was cross-calibrated for equivalent (18)F content by direct comparison with the national standard for (18)F in the same geometry.
METHODS: Sources containing standardized (68)GeCl(4) or (18)F-FDG solutions were prepared at the National Institute of Standards and Technology (NIST) with mock syringe blanks used in the construction of a commercially available epoxy-based (68)Ge calibration source. These sources and several NIST-constructed epoxy-based (68)Ge mock syringes were then used as artifact standards to determine calibration factors for NIST-maintained activity calibrators and secondary standard ionization chambers to enable calibration of the actual commercial sources. A direct comparison between the solution-based (68)Ge sources and the (18)F-FDG sources allowed for an empiric determination of the relative response for these radionuclides in several commercial activity calibrators. Potential measurement effects due to differences between the solution composition and the epoxy and theoretic (68)Ge-to-(18)F response ratios were studied by Monte Carlo simulation.
RESULTS: The calibration factors developed in this study enabled NIST to calibrate epoxy-based mock syringe sources with a relative combined standard uncertainty of 0.52%. The direct comparisons of the (68)Ge and (18)F standards in the various ionization chambers allowed the activity to be expressed in terms of equivalent (18)F activity with a relative combined standard uncertainty of about 0.9%.
CONCLUSION: The ability for NIST to calibrate these epoxy-based mock syringes enabled, for the first time to our knowledge, the direct traceability to the national (68)Ge standard to be established for this type of source. Through a direct comparison with the NIST (18)F standard, the determination of the relative response ratios in activity calibrators enabled the equivalent (18)F activity to be determined in a way that was also traceable to the national (18)F activity standard.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20197450     DOI: 10.2967/jnumed.109.070300

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  21 in total

1.  Longitudinal monitoring of reconstructed activity concentration on a clinical time-of-flight PET/CT scanner.

Authors:  Lawrence R MacDonald; Amy E Perkins; Chi-Hua Tung
Journal:  J Med Imaging (Bellingham)       Date:  2016-11-23

2.  Measuring temporal stability of positron emission tomography standardized uptake value bias using long-lived sources in a multicenter network.

Authors:  Darrin Byrd; Rebecca Christopfel; Grae Arabasz; Ciprian Catana; Joel Karp; Martin A Lodge; Charles Laymon; Eduardo G Moros; Mikalai Budzevich; Sadek Nehmeh; Joshua Scheuermann; John Sunderland; Jun Zhang; Paul Kinahan
Journal:  J Med Imaging (Bellingham)       Date:  2018-01-04

3.  An update on 'dose calibrator' settings for nuclides used in nuclear medicine.

Authors:  Denis E Bergeron; Jeffrey T Cessna
Journal:  Nucl Med Commun       Date:  2018-06       Impact factor: 1.690

4.  Quantifying and reducing the effect of calibration error on variability of PET/CT standardized uptake value measurements.

Authors:  Catherine M Lockhart; Lawrence R MacDonald; Adam M Alessio; Wendy A McDougald; Robert K Doot; Paul E Kinahan
Journal:  J Nucl Med       Date:  2011-01-13       Impact factor: 10.057

5.  The next generation of current measurement for ionization chambers.

Authors:  R Fitzgerald; D E Bergeron; S P Giblin; D G Jarrett; S M Judge; C Michotte; H Scherer; N M Zimmerman
Journal:  Appl Radiat Isot       Date:  2020-05-13       Impact factor: 1.513

6.  Factors affecting the repeatability of gamma camera calibration for quantitative imaging applications using a sealed source.

Authors:  N Anizan; H Wang; X C Zhou; R L Wahl; E C Frey
Journal:  Phys Med Biol       Date:  2015-01-16       Impact factor: 3.609

7.  Biases in Multicenter Longitudinal PET Standardized Uptake Value Measurements.

Authors:  Robert K Doot; Larry A Pierce; Darrin Byrd; Brian Elston; Keith C Allberg; Paul E Kinahan
Journal:  Transl Oncol       Date:  2014-02-01       Impact factor: 4.243

8.  Long-term stability of carrier-added (68)Ge standardized solutions.

Authors:  B E Zimmerman; D E Bergeron; R Fitzgerald; J T Cessna
Journal:  Appl Radiat Isot       Date:  2015-11-26       Impact factor: 1.513

9.  Two determinations of the Ge-68 half-life.

Authors:  Denis E Bergeron; Jeffrey T Cessna; Brian E Zimmerman
Journal:  Appl Radiat Isot       Date:  2017-10-20       Impact factor: 1.513

10.  Early experiences in establishing a regional quantitative imaging network for PET/CT clinical trials.

Authors:  Robert K Doot; Tove Thompson; Benjamin E Greer; Keith C Allberg; Hannah M Linden; David A Mankoff; Paul E Kinahan
Journal:  Magn Reson Imaging       Date:  2012-07-15       Impact factor: 2.546

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.