Literature DB >> 20177978

Quality control of ultrasound transducers: analysis of evaluation parameters and results of a survey of 116 transducers in a single hospital.

P Cozzolino1, R Stramare, A Udilano, M Castoro, E Scagliori, V Beltrame, L Rubaltelli.   

Abstract

PURPOSE: The aim of this study was to develop a quality control programme for ultrasound equipment and assess equipment performance at a single hospital.
MATERIALS AND METHODS: We tested 116 transducers by means of a phantom; ultrasound quality control tests were done on parameters from B-mode images, and each parameter was rated according to a three-point score as good, sufficient or poor.
RESULTS: Surface transducers showed worse performance in terms of image uniformity, with 7/55 (13%) transducers rated poor, and depth of penetration, with 24/55 (44%) transducers, rated sufficient or poor. The 3.5-MHz transducers showed worse performance in lateral resolution, with 18/48 (38%) rated poor, and particularly in focusing, with 40/48 (83%) classified as poor. In addition, the 3.5-MHz transducers proved less accurate in vertical distance accuracy than in horizontal distance, with 31/48 (64%) considered sufficient. No significant geometric distortions of masses were identified, although 4/116 (3%) transducers failed to visualise two masses.
CONCLUSIONS: The first goal of an ultrasound quality control programme is to establish specific parameters to evaluate equipment performance and, should results be below recommended action levels, implement corrective measures.

Mesh:

Year:  2010        PMID: 20177978     DOI: 10.1007/s11547-010-0533-0

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  10 in total

1.  [Quality management in cardiovascular echography].

Authors:  Giuseppe Gullace
Journal:  Ital Heart J Suppl       Date:  2002-12

2.  Assessment of the acoustic properties of common tissue-mimicking test phantoms.

Authors:  J E Browne; K V Ramnarine; A J Watson; P R Hoskins
Journal:  Ultrasound Med Biol       Date:  2003-07       Impact factor: 2.998

3.  The effect of speed of sound in ultrasound test objects on lateral resolution.

Authors:  Nicholas J Dudley; Nicholas M Gibson; Mark J Fleckney; Peter D Clark
Journal:  Ultrasound Med Biol       Date:  2002 Nov-Dec       Impact factor: 2.998

4.  Simulation study of effects of speed of sound and attenuation on ultrasound lateral resolution.

Authors:  Quan Chen; James A Zagzebski
Journal:  Ultrasound Med Biol       Date:  2004-10       Impact factor: 2.998

5.  Implementation and validation of three automated methods for measuring ultrasound maximum depth of penetration: application to ultrasound quality control.

Authors:  K R Gorny; D J Tradup; N J Hangiandreou
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

6.  Objective performance testing and quality assurance of medical ultrasound equipment.

Authors:  Johan M Thijssen; Gert Weijers; Chris L de Korte
Journal:  Ultrasound Med Biol       Date:  2007-03       Impact factor: 2.998

7.  Image quality control in breast ultrasound.

Authors:  R Lagalla; M Midiri
Journal:  Eur J Radiol       Date:  1998-05       Impact factor: 3.528

8.  Real-time B-mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1.

Authors:  M M Goodsitt; P L Carson; S Witt; D L Hykes; J M Kofler
Journal:  Med Phys       Date:  1998-08       Impact factor: 4.071

9.  Phantoms and automated system for testing the resolution of ultrasound scanners.

Authors:  J J Rownd; E L Madsen; J A Zagzebski; G R Frank; F Dong
Journal:  Ultrasound Med Biol       Date:  1997       Impact factor: 2.998

10.  Performance testing of medical ultrasound equipment: fundamental vs. harmonic mode.

Authors:  M C van Wijk; J M Thijssen
Journal:  Ultrasonics       Date:  2002-05       Impact factor: 2.890

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.