Phantoms and automated system for testing the resolution of ultrasound scanners.

J J Rownd1, E L Madsen, J A Zagzebski, G R Frank, F Dong.   

Abstract

Tissue-mimicking phantoms and an automated system have been developed for testing the resolution performance of ultrasound scanners by determining detectability of low to higher contrast spherical lesions over the entire depth of field. Axial, lateral and elevational resolutions are accounted for simultaneously and equally. Tissue-mimicking spherical simulated lesions are either 3 or 4 mm in diameter and have one of four different intrinsic material contrasts. For each diameter and contrast, there is a set of 109 lesions in a regular array with coplanar centers extending from 0.5-15.5 cm in depth. With the scan slice superimposed on the spheres, the image is frame-grabbed for automated analysis. A diameter-dependent lesion signal-to-noise ratio is computed for each pixel position in the image, excluding a 5-mm boundary. Two universal thresholds, resulting from maximization of agreement between the automated system and human observers, give rise to a depth range, or "resolution zone", over which detection exists for each type lesion.

Entities:  

Mesh:

Year:  1997        PMID: 9140182     DOI: 10.1016/s0301-5629(96)00205-0

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  4 in total

1.  Mechanical and clinical performance of pulse-inversion tissue harmonic imaging in the superficial region.

Authors:  Chieko Sugawara; Akira Takahashi
Journal:  J Med Ultrason (2001)       Date:  2010-03-18       Impact factor: 1.314

2.  Quality control of ultrasound transducers: analysis of evaluation parameters and results of a survey of 116 transducers in a single hospital.

Authors:  P Cozzolino; R Stramare; A Udilano; M Castoro; E Scagliori; V Beltrame; L Rubaltelli
Journal:  Radiol Med       Date:  2010-02-22       Impact factor: 3.469

3.  Review of tissue simulating phantoms with controllable optical, mechanical and structural properties for use in optical coherence tomography.

Authors:  Guy Lamouche; Brendan F Kennedy; Kelsey M Kennedy; Charles-Etienne Bisaillon; Andrea Curatolo; Gord Campbell; Valérie Pazos; David D Sampson
Journal:  Biomed Opt Express       Date:  2012-05-15       Impact factor: 3.732

4.  Evaluations of UltraiQ software for objective ultrasound image quality assessment using images from a commercial scanner.

Authors:  Zaiyang Long; Donald J Tradup; Scott F Stekel; Krzysztof R Gorny; Nicholas J Hangiandreou
Journal:  J Appl Clin Med Phys       Date:  2018-01-16       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.