Detection of hypoglycemia with the GlucoWatch biographer.

K R Pitzer1, S Desai, T Dunn, S Edelman, Y Jayalakshmi, J Kennedy, J A Tamada, R O Potts.   

Abstract

OBJECTIVE: Hypoglycemia is a common acute complication of diabetes therapy. The GlucoWatch biographer provides frequent and automatic glucose measurements with an adjustable low-glucose alarm. We have analyzed the performance of the biographer low-glucose alarm relative to hypoglycemia as defined by blood glucose < or = 3.9 mmol/l. RESEARCH DESIGN AND METHODS: The analysis was based on 1,091 biographer uses from four clinical trials. which generated 14,487 paired (biographer and blood glucose) readings.
RESULTS: The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose < or = 3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose >3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11347748     DOI: 10.2337/diacare.24.5.881

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  23 in total

1.  Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

2.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

3.  Methodology for hypoglycaemia detection based on the processing, analysis and classification of the electroencephalogram.

Authors:  F Iaione; J L B Marques
Journal:  Med Biol Eng Comput       Date:  2005-07       Impact factor: 2.602

4.  Simultaneous extraction of urea and glucose by reverse iontophoresis in vivo.

Authors:  Anke Sieg; Richard H Guy; M Begoña Delgado-Charro
Journal:  Pharm Res       Date:  2004-10       Impact factor: 4.200

Review 5.  In vitro, in vivo and post explantation testing of glucose-detecting biosensors: current methods and recommendations.

Authors:  Heidi E Koschwanez; William M Reichert
Journal:  Biomaterials       Date:  2007-04-19       Impact factor: 12.479

6.  Evaluation of a minimally invasive system for measuring glucose area under the curve during oral glucose tolerance tests: usefulness of sweat monitoring for precise measurement.

Authors:  Kazuhiko Sakaguchi; Yushi Hirota; Naoko Hashimoto; Wataru Ogawa; Tomoya Hamaguchi; Toshihiro Matsuo; Jun-ichiro Miyagawa; Mitsuyoshi Namba; Toshiyuki Sato; Seiki Okada; Koji Tomita; Munehide Matsuhisa; Hideaki Kaneto; Keisuke Kosugi; Hiroshi Maegawa; Hiromu Nakajima; Atsunori Kashiwagi
Journal:  J Diabetes Sci Technol       Date:  2013-05-01

7.  Reverse iontophoresis of amino acids: identification and separation of stratum corneum and subdermal sources in vitro.

Authors:  Camille C Bouissou; Jean-Philippe Sylvestre; Richard H Guy; M Begoña Delgado-Charro
Journal:  Pharm Res       Date:  2009-10-06       Impact factor: 4.200

Review 8.  Hypo- and Hyperglycemic Alarms: Devices and Algorithms.

Authors:  Daniel Howsmon; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2015-04-30

9.  Single-cell label-free photoacoustic flowoxigraphy in vivo.

Authors:  Lidai Wang; Konstantin Maslov; Lihong V Wang
Journal:  Proc Natl Acad Sci U S A       Date:  2013-03-27       Impact factor: 11.205

10.  Carbon nanotube composites for glucose biosensor incorporated with reverse iontophoresis function for noninvasive glucose monitoring.

Authors:  Tai-Ping Sun; Hsiu-Li Shieh; Congo Tak-Shing Ching; Yan-Dong Yao; Su-Hua Huang; Chia-Ming Liu; Wei-Hao Liu; Chung-Yuan Chen
Journal:  Int J Nanomedicine       Date:  2010-05-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.