Brian J Holland1, Beth F Printz, Wyman W Lai.
Abstract
BACKGROUND: One potential source of error in phase contrast (PC) congenital CMR flow measurements is caused by phase offsets due to local non-compensated eddy currents. Phantom correction of these phase offset errors has been shown to result in more accurate measurements of blood flow in adults with structurally normal hearts. We report the effect of phantom correction on PC flow measurements at a clinical congenital CMR program.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20205725 PMCID: PMC2845128 DOI: 10.1186/1532-429X-12-11
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Figure 1Phase Contrast Images of the Ascending Aorta with Region of Interest.
Figure 2Phase Contrast Images of the Phantom with Region of Interest from Ascending Aorta.
Figure 3Phase Contrast Flow vs. Time Curves. Flow versus time curves from one cardiac cycle for the main pulmonary artery without (blue) and with (red) phantom correction. Flow volumes were calculated by integrating the area under the curves. With phantom correction, the flow in the main pulmonary artery changed from 5.1 to 2.4 L/min/m2, and the regurgitant fraction changed from 0% to 19%.
Definitions for Magnitude of Change
| Variable | Clinically Significant Change | Marked Change |
|---|---|---|
| ≥ 0.5 L/min/m2 | ≥ 1.0 L/min/m2 | |
| ≥ 0.25 L/min/m2 | ≥ 0.5 L/min/m2 | |
| ≥ 0.4 | ≥ 0.8 | |
| =≥10% | ≥ 20% | |
| ≥ 10% | ≥ 20% | |
AAO-Ascending Aorta, AR-Aortic Regurgitation, LPA-Left Pulmonary Artery, MPA-Main Pulmonary Artery, PR-Pulmonary Regurgitation, QP-Pulmonary blood flow, QS-Systemic blood flow, QPR-Right pulmonary blood flow as a percent of total pulmonary blood flow, RPA-Right Pulmonary Artery
Patient Characteristics
| Analyzed Patients | |
|---|---|
| 149 | |
| 17.5 (0.3-69.3) | |
| 1.6 (0.3-2.2) | |
| 36 (25%) | |
| 16 (11%) | |
| 15 (10%) | |
| 13 (9%) | |
| 11 (7%) | |
| 10 (7%) | |
| 28 (19%) | |
| 3 (2%) | |
| 17 (11%) | |
BSA-Body surface area
Overall Effect of Phantom Correction
| Variable | N | Median Absolute Change | Range |
|---|---|---|---|
| 118 | 0.4 | 0.0-5.0 | |
| 144 | 0.4 | 0.0-3.5 | |
| 70 | 0.1 | 0.0-2.6 | |
| 61 | 0.2 | 0.0-2.1 | |
| 113 | 0.2 | 0.0-4.6 | |
| 53 | 6 | 0-27 | |
| 24 | 3 | 1-14 | |
| 57 | 5 | 0-54 | |
AAO-Ascending Aorta, AR-Aortic Regurgitation, LPA-Left Pulmonary Artery, MPA-Main Pulmonary Artery, PR-Pulmonary Regurgitation, QP-Pulmonary blood flow, QS-Systemic blood flow, QPR-Right pulmonary blood flow as a percent of total pulmonary blood flow, RPA-Right Pulmonary Artery
Magnitude of Change with Phantom Correction
| Variable | N | Clinically Significant Change | Marked Change |
|---|---|---|---|
| 118 | 50 (42%) | 21 (18%) | |
| 144 | 51 (35%) | 15 (10%) | |
| 70 | 17 (24%) | 6 (9%) | |
| 61 | 29 (48%) | 15 (25%) | |
| 113 | 20 (18%) | 7 (6%) | |
| 53 | 14 (26%) | 1 (2%) | |
| 24 | 3 (13%) | 0 (0%) | |
| 57 | 13 (23%) | 6 (11%) | |
AAO-Ascending Aorta, AR-Aortic Regurgitation, LPA-Left Pulmonary Artery, MPA-Main Pulmonary Artery, PR-Pulmonary Regurgitation, QP-Pulmonary blood flow, QS-Systemic blood flow, QPR-Right pulmonary blood flow as a percent of total pulmonary blood flow, RPA-Right Pulmonary Artery
Direction of Change with Phantom Correction
| Increase with Phantom Correction (N) | Decrease with Phantom Correction (N) | P value | |
|---|---|---|---|
| 52 | 66 | 0.20 | |
| 109 | 35 | < 0.001 | |
| 32 | 38 | 0.47 | |
| 20 | 41 | < 0.001 | |
| 35 | 78 | < 0.001 | |
| 28 | 25 | 0.68 | |
| 6 | 18 | 0.01 | |
| 33 | 24 | 0.23 | |
AAO-Ascending Aorta, AR-Aortic Regurgitation, LPA-Left Pulmonary Artery, MPA-Main Pulmonary Artery, PR-Pulmonary Regurgitation, QP-Pulmonary blood flow, QS-Systemic blood flow, QPR-Right pulmonary blood flow as a percent of total pulmonary blood flow, RPA-Right Pulmonary Artery
Tetralogy of Fallot Patients: Effect of Phantom Correction
| Variable | N | Clinically | P value | Marked Change | P value |
|---|---|---|---|---|---|
| 32 | 17 (53%) | 0.15 | 6 (19%) | 1.00 | |
| 34 | 15 (44%) | 0.28 | 2 (6%) | 0.55 | |
| 27 | 7 (26%) | 0.64 | 3 (11%) | 0.46 | |
| 24 | 11 (46%) | 0.68 | 5 (21%) | 0.64 | |
| 31 | 7 (23%) | 0.65 | 2 (6%) | 1.00 | |
| 32 | 8 (25%) | 1.00 | 1 (3%) | 1.00 | |
| 3 | 0 (0%) | N/S | 0 (0%) | N/S | |
| 22 | 6 (27%) | 0.61 | 2 (9%) | 1.00 | |
AAO-Ascending Aorta, AR-Aortic Regurgitation, LPA-Left Pulmonary Artery, MPA-Main Pulmonary Artery, PR-Pulmonary Regurgitation, QP-Pulmonary blood flow, QS-Systemic blood flow, QPR-Right pulmonary blood flow as a percent of total pulmonary blood flow, RPA-Right Pulmonary Artery
Figure 4Q. QP-Pulmonary blood flow, QS-Systemic blood flow
Figure 5Flow Correction vs. Blood Vessel Size.