LET dependence of GafChromic films and an ion chamber in low-energy proton dosimetry.

Daniel Kirby1, Stuart Green, Hugo Palmans, Richard Hugtenburg, Cecile Wojnecki, David Parker.   

Abstract

Dosimetry using a PMMA phantom was performed in 15 and 29 MeV proton beams from the Birmingham cyclotron, with a Markus parallel-plate ionization chamber and GafChromic EBT and MD-V2-55 film. Simulations of the depth-dose curves were performed with FLUKA 2008.3 and MCNPX 2.5.0, which agreed almost perfectly with each other in range and only differed by 2% in the Bragg peak (BP) region. FLUKA was also used to calculate k(Q) factors for Markus chamber measurements as an improvement to the IAEA TRS-398 values in low-energy beams. FLUKA depth-dose simulations overestimate the BP height measured by ion chamber by about 10%, where the initial proton energy spread was estimated by fitting to the slope of the measured BP distal edge. Both GafChromic films showed an under-response in the BP compared to ion chamber; however, EBT exhibits this effect at lower energies than MD-V2-55. A possible reason for this is attributed to the shape and arrangement of the monomer particles being different in the active components of EBT and MD-V2-55. Relative effectiveness (RE) of both films is presented as functions of residual range R(res) in water and peak proton energy determined by FLUKA, with considerations for the spatial separation of the two active layers in each film. The proton energies at which RE reduces to 90% of maximum film response are 6.7 and 3.2 MeV for MD-V2-55 and EBT, respectively. Additionally, a beam quality correction factor (g(Q,Q0)) is suggested for both GafChromic films, involving water-to-film stopping power ratios evaluated using ICRU recommendations, and a polymer yield factor G(Q0)/G(Q). RE in this work is equated to the reciprocal of the polymer yield factor. The calculated values of (S(w,film))Q/(S(w,film))Q0 are constant within 2.1% and 1.2% across the proton energy range of 1-300 MeV for EBT and MD-V2-55, respectively, so it is concluded that the polymer yield factor is the dominant factor causing the LET quenching effect.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20019399     DOI: 10.1088/0031-9155/55/2/006

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  17 in total

1.  Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

Authors:  S Reinhardt; M Würl; C Greubel; N Humble; J J Wilkens; M Hillbrand; A Mairani; W Assmann; K Parodi
Journal:  Radiat Environ Biophys       Date:  2015-01-09       Impact factor: 1.925

2.  Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.

Authors:  Maria F Chan; Chin-Cheng Chen; Chengyu Shi; Jingdong Li; Xiaoli Tang; Xiang Li; Dennis Mah
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2017-05-16

3.  LET dependent response of GafChromic films investigated with MeV ion beams.

Authors:  V Grilj; D J Brenner
Journal:  Phys Med Biol       Date:  2018-12-18       Impact factor: 3.609

4.  Measurement of characteristic prompt gamma rays emitted from oxygen and carbon in tissue-equivalent samples during proton beam irradiation.

Authors:  Jerimy C Polf; Rajesh Panthi; Dennis S Mackin; Matt McCleskey; Antti Saastamoinen; Brian T Roeder; Sam Beddar
Journal:  Phys Med Biol       Date:  2013-08-06       Impact factor: 3.609

5.  A New Anthropomorphic Pediatric Spine Phantom for Proton Therapy Clinical Trial Credentialing.

Authors:  Dana J Lewis; Paige A Taylor; David S Followill; Narayan Sahoo; Anita Mahajan; Francesco C Stingo; Stephen F Kry
Journal:  Int J Part Ther       Date:  2018-07-26

6.  Dosimetry with gafchromic films based on a new micro-opto-electro-mechanical system.

Authors:  C Guardiola; A Márquez; M C Jiménez-Ramos; J García López; A Baratto-Roldán; X Muñoz-Berbel
Journal:  Sci Rep       Date:  2021-05-17       Impact factor: 4.379

7.  Radiographic film dosimetry of proton beams for depth-dose constancy check and beam profile measurement.

Authors:  Inhwan J Yeo; Anthony Teran; Abiel Ghebremedhin; Matt Johnson; Baldev Patyal
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

8.  Issues involved in the quantitative 3D imaging of proton doses using optical CT and chemical dosimeters.

Authors:  Simon Doran; Tina Gorjiara; Andrzej Kacperek; John Adamovics; Zdenka Kuncic; Clive Baldock
Journal:  Phys Med Biol       Date:  2015-01-02       Impact factor: 3.609

9.  Evaluation of the dosimetric properties of a diode detector for small field proton radiosurgery.

Authors:  Grant A McAuley; Anthony V Teran; Jerry D Slater; James M Slater; Andrew J Wroe
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

10.  Dose- rather than fluence-averaged LET should be used as a single-parameter descriptor of proton beam quality for radiochromic film dosimetry.

Authors:  Andreas Franz Resch; Paul David Heyes; Hermann Fuchs; Niels Bassler; Dietmar Georg; Hugo Palmans
Journal:  Med Phys       Date:  2020-03-13       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.