[Influence of acetabular cup design on the primary implant stability : an experimental and numerical analysis].

C Zietz1, A Fritsche, D Kluess, W Mittelmeier, R Bader.   

Abstract

BACKGROUND: An adequate primary stability and a subsequent stable osseous fixation (secondary stability) of artificial hip cups are required for long-term implant survival. The aim of this study was to analyse the design of cementless press-fit cups as an influencing factor of primary stability.
MATERIAL AND METHODS: Different hemispherical and conical cup designs were analysed. The fixation stability of the cups was detected experimentally using a spongiosa and a cortical model based on artificial bone as well as a numerical simulation using a spongiosa model by pull-out and lever-out tests. In addition, the stress on the osseous cup cavity was determined in the finite-element analysis.
RESULTS: All tested cup designs revealed higher fixation stability in the cortical bone model compared to the spongiosa model. The experimental tests did not show an increase of fixation stability with the conical cup profile in comparison to hemispherical cup profiles.
CONCLUSION: Therefore, cementless press-fit cups with conical cup profile do not provide a higher primary stability in comparison to hemispherical cups. Moreover, the stress on the bone cavity was lower inserting the hemispherical cup profiles in contrast to the conical profiles.

Entities:  

Mesh:

Year:  2009        PMID: 19636530     DOI: 10.1007/s00132-009-1467-x

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  17 in total

1.  [Primary stability of cement-free press-fit acetabulum cups. In vitro displacement studies].

Authors:  A Kuhn; G Scheller; M Schwarz
Journal:  Biomed Tech (Berl)       Date:  1999-12       Impact factor: 1.411

Review 2.  [Anchoring principles in hip prosthesis implantation. II: Acetabulum components].

Authors:  S J Breusch; P R Aldinger; M Thomsen; M Lukoschek; V Ewerbeck
Journal:  Unfallchirurg       Date:  2000-12       Impact factor: 1.000

3.  Stability of press-fit acetabular cups.

Authors:  E Adler; S A Stuchin; F J Kummer
Journal:  J Arthroplasty       Date:  1992-09       Impact factor: 4.757

4.  The initial stability of uncemented acetabular components.

Authors:  M J Curtis; R H Jinnah; V D Wilson; D S Hungerford
Journal:  J Bone Joint Surg Br       Date:  1992-05

5.  Loss in mechanical contact of cementless acetabular prostheses due to post-operative weight bearing: a biomechanical model.

Authors:  Chiara Maria Bellini; Fabio Galbusera; Roberto Giacometti Ceroni; Manuela Teresa Raimondi
Journal:  Med Eng Phys       Date:  2006-03-29       Impact factor: 2.242

6.  Acetabular strains produced by oversized press fit cups.

Authors:  M D Ries; M Harbaugh
Journal:  Clin Orthop Relat Res       Date:  1997-01       Impact factor: 4.176

7.  [Cementless fixation of the endoprosthesis using trabecular, 3-dimensional interconnected surface structures].

Authors:  W Mittelmeier; I Grunwald; R Schäfer; H Grundei; R Gradinger
Journal:  Orthopade       Date:  1997-02       Impact factor: 1.087

8.  Press-fit acetabular cup fixation: principles and testing.

Authors:  W Macdonald; L V Carlsson; G J Charnley; C M Jacobsson
Journal:  Proc Inst Mech Eng H       Date:  1999       Impact factor: 1.617

9.  [Initial stability of press-fit acetabular cups--an in-vitro study].

Authors:  R Wetzel; M Simnacher; G Scheller
Journal:  Biomed Tech (Berl)       Date:  2005-12       Impact factor: 1.411

10.  Fracture of the acetabulum during insertion of an oversized hemispherical component.

Authors:  Y S Kim; J J Callaghan; P B Ahn; T D Brown
Journal:  J Bone Joint Surg Am       Date:  1995-01       Impact factor: 5.284

View more
  1 in total

1.  Finite element model of the impaction of a press-fitted acetabular cup.

Authors:  Adrien Michel; Vu-Hieu Nguyen; Romain Bosc; Romain Vayron; Philippe Hernigou; Salah Naili; Guillaume Haiat
Journal:  Med Biol Eng Comput       Date:  2016-08-04       Impact factor: 2.602

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.