Loss in mechanical contact of cementless acetabular prostheses due to post-operative weight bearing: a biomechanical model.

Chiara Maria Bellini1, Fabio Galbusera, Roberto Giacometti Ceroni, Manuela Teresa Raimondi.   

Abstract

The primary stability of cementless acetabular components is a prerequisite for their clinical success. The target of the present study was to analyse possible effects of post-operative joint loading on the initial mechanical stability of a press-fitted acetabular prosthesis. For this purpose, a three-dimensional finite element model of the pelvic bone with acetabular reconstruction was set-up. The analysis included two steps: (1) simulation of the prosthesis press-fit implantation and (2) simulation of the instant of peak resultant hip loading during the one-legged stance. The difference between the contact pressures at the bone/implant interface, at the end of the second step and those at the end of the first step was calculated and assumed as an index of variation in mechanical contact due to post-operative weight bearing. The results show that, due to hip loading, contact pressures given by press-fit increase in the postero-superior acetabular region but decrease in the antero-inferior acetabular region. The calculated area in which the contact pressures decrease extend to about 30% of the total contact surface. These results imply that post-operative joint loading significantly reduces the mechanical stability given by press-fit. The decrease in contact pressures at the bone/implant interface may result in a lack of osteointegration, possibly hindering the implant secondary stability. It may also create a route for wear debris, possibly favouring periprosthetic osteolysis, which may lead to further loss in contact and clinical failure of the implant due to loosening.

Entities:  

Mesh:

Year:  2006        PMID: 16569508     DOI: 10.1016/j.medengphy.2006.02.008

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  4 in total

1.  Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins.

Authors:  Wolfgang O Huber; Philip C Noble
Journal:  Int Orthop       Date:  2013-11-22       Impact factor: 3.075

2.  [Influence of acetabular cup design on the primary implant stability : an experimental and numerical analysis].

Authors:  C Zietz; A Fritsche; D Kluess; W Mittelmeier; R Bader
Journal:  Orthopade       Date:  2009-11       Impact factor: 1.087

3.  In vitro assessment of Function Graded (FG) artificial Hip joint stem in terms of bone/cement stresses: 3D Finite Element (FE) study.

Authors:  Fawzi F Al-Jassir; H Fouad; Othaman Y Alothman
Journal:  Biomed Eng Online       Date:  2013-01-16       Impact factor: 2.819

4.  The Biomechanical Effect of Loading Speed on Metal-on-UHMWPE Contact Mechanics.

Authors:  Radovan Zdero; Zahra S Bagheri; Mojtaba Rezaey; Emil H Schemitsch; Habiba Bougherara
Journal:  Open Biomed Eng J       Date:  2014-05-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.