| Literature DB >> 1959276 |
J P Collier1, M B Mayor, J L McNamara, V A Surprenant, R E Jensen.
Abstract
The extent of wear of retrieved, polyethylene tibial components appears to be related to design. To test this observation, 122 tibial inserts were graded for wear, and new components of several designs were tested for contact stress using Fuji film. Finite element analysis provided insight into subsurface stresses. Significant wear was seen in 61.5% of the tibial inserts examined. The presence of unconsolidated polymer powder was seen in 44% of the tibial inserts and was found to be statistically correlated with severe wear of the articulating surface. Contact stress in several noncongruent designs was found to exceed the yield strength of polyethylene. There was a positive correlation between the intensity of wear and the level of contact stress, with noncongruent designs having greater wear than fully congruent geometries. In the noncongruent designs, the thinner polyethylene components showed greater wear than thicker polyethylene inserts of the same design.Entities:
Mesh:
Substances:
Year: 1991 PMID: 1959276
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176