Abstract
In the past few years, high-performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS) has matured to a true alternative to antibody-based immunoassays in routine therapeutic drug monitoring. In transplantation medicine, mass spectrometry-based assessment of immunosuppressant drug levels is considered a gold standard diagnostic procedure. We describe a fast state-of-the-art routine online solid-phase extraction (SPE) HPLC-MS/MS analysis platform that allows monitoring of cyclosporine A, tacrolimus, sirolimus and everolimus from 50-microl aliquots of EDTA whole blood specimens within 3.4 min total analysis time. Sample purification is done by offline protein precipitation followed by two automated chromatographic separation steps. Mass spectrometry-based analyte quantification relies on selected reaction monitoring experiments. The assay underwent complete validation and performance evaluation studies and performs very well in several international proficiency testing schemes. In daily routine, it allows reporting of about 75 patient sample results per work shift with a typical total individual sample turnaround time of less than 3 h.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19325549 DOI: 10.1038/nprot.2009.25
Source DB: PubMed Journal: Nat Protoc ISSN: 1750-2799 Impact factor: 13.491