| Literature DB >> 11092585 |
U Christians1, W Jacobsen, N Serkova, L Z Benet, C Vidal, K F Sewing, M P Manns, G I Kirchner.
Abstract
We developed a universal LC-mass spectrometry assay with automated online extraction (LC/LC-MS) to quantify the immunosuppressants cyclosporine, tacrolimus, sirolimus and SDZ-RAD alone or in combination in whole blood. After protein precipitation, samples were loaded on a C18 extraction column, were washed and, after activation of the column-switching valve, were backflushed onto the C8 analytical column. [M+Na]+ ions were detected in the selected ion mode. For tacrolimus, sirolimus and SDZ-RAD, the assay was linear from 0.25 to 100 microg/l and for cyclosporine from 7.5 to 1250 microg/l (all r2>0.99). Analytical recovery was >85% and, in general, inter-day, intra-day variability for precision and accuracy were <10%.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11092585 DOI: 10.1016/s0378-4347(00)00380-7
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273