Literature DB >> 17702856

Methanol-associated matrix effects in electrospray ionization tandem mass spectrometry.

Thomas M Annesley1.   

Abstract

BACKGROUND: Matrix effects can profoundly reduce the performance of electrospray ionization mass spectrometry. Preliminary observations indicated that the methanol used in the mobile phase could be a source of differential ionization or ion suppression.
METHODS: Drug stability studies, analysis of biological extracts, mixing experiments, and postcolumn infusions were used to test 9 commercial methanols for ionization differences in liquid chromatography-tandem mass spectrometry assays for immunosuppressants. Area responses for the drugs and internal standards were compared for mobile phases prepared with each selected methanol. Postcolumn infusion experiments were performed to confirm the degree of ionization differences occurring at the ion source, and to evaluate the proportions of ammonium, sodium, and potassium adducts.
RESULTS: The decrease in signal for the immunosuppressant drugs was shown to result from differential ionization associated with the selected methanols. Product ion intensity varied by 10-fold among the methanols tested. For sirolimus, tacrolimus, and mycophenolic acid, the percentage change in ionization was the same for the drug and its corresponding internal standard. Postcolumn sirolimus infusion evaluation revealed that a 1000-fold analyte concentration difference did not affect ionization. The proportions of ammonium, sodium, and potassium adducts of sirolimus precursor ions differed in relation to the source of methanol.
CONCLUSIONS: Organic solvents used in mobile phases and extract preparation of biological samples may be associated with ion suppression, affecting adduct formation and assay sensitivity.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17702856     DOI: 10.1373/clinchem.2007.090811

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  9 in total

1.  A method for monitoring and controlling reproducibility of intensity data in complex electrospray mass spectra: a thermometer ion-based strategy.

Authors:  Paolo Lecchi; Jinghua Zhao; Wesley S Wiggins; Tzong-Hao Chen; Ping F Yip; Brian C Mansfield; John M Peltier
Journal:  J Am Soc Mass Spectrom       Date:  2008-11-06       Impact factor: 3.109

2.  Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry.

Authors:  James J Pitt
Journal:  Clin Biochem Rev       Date:  2009-02

Review 3.  The use of mass spectrometry to improve the diagnosis and the management of the HPA axis.

Authors:  Phillip J Monaghan; Brian G Keevil; Peter J Trainer
Journal:  Rev Endocr Metab Disord       Date:  2013-06       Impact factor: 6.514

4.  Acetonitrile Ion Suppression in Atmospheric Pressure Ionization Mass Spectrometry.

Authors:  Kevin Colizza; Keira E Mahoney; Alexander V Yevdokimov; James L Smith; Jimmie C Oxley
Journal:  J Am Soc Mass Spectrom       Date:  2016-08-30       Impact factor: 3.109

5.  A rapid HPLC-MS/MS method for the simultaneous quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human blood samples.

Authors:  Christoph Seger; Karin Tentschert; Wolfgang Stöggl; Andrea Griesmacher; Steven L Ramsay
Journal:  Nat Protoc       Date:  2009       Impact factor: 13.491

6.  Toxic marine puffer fish in Thailand seas and tetrodotoxin they contained.

Authors:  Monrat Chulanetra; Nitat Sookrung; Potjanee Srimanote; Nitaya Indrawattana; Jeeraphong Thanongsaksrikul; Yuwaporn Sakolvaree; Manas Chongsa-Nguan; Hisao Kurazono; Wanpen Chaicumpa
Journal:  Toxins (Basel)       Date:  2011-10-18       Impact factor: 4.546

7.  Validation of an LC-MS/MS method to determine five immunosuppressants with deuterated internal standards including MPA.

Authors:  Armin Buchwald; Karl Winkler; Thomas Epting
Journal:  BMC Clin Pharmacol       Date:  2012-01-11

Review 8.  Review of the Use of Liquid Chromatography-Tandem Mass Spectrometry in Clinical Laboratories: Part II-Operations.

Authors:  Brian A Rappold
Journal:  Ann Lab Med       Date:  2022-09-01       Impact factor: 4.941

9.  Regioisomer-independent quantification of fatty acid oxidation products by HPLC-ESI-MS/MS analysis of sodium adducts.

Authors:  Katelyn W Ahern; Vlad Serbulea; Catherine L Wingrove; Zachary T Palas; Norbert Leitinger; Thurl E Harris
Journal:  Sci Rep       Date:  2019-08-01       Impact factor: 4.379

  9 in total