Local noise estimation in low-dose chest CT images.

J Padgett1, A M Biancardi, C I Henschke, D Yankelevitz, A P Reeves.   

Abstract

PURPOSE: Image noise in computed tomography (CT) images may have significant local variation due to tissue properties, dose, and location of the X-ray source. We developed and tested an automated tissue-based estimator method for estimating local noise in CT images.
METHOD: An automated TBE method for estimating the local noise in CT image in 3 steps was developed: (1) Partition the image into homogeneous and transition regions, (2) For each pixel in the homogeneous regions, compute the standard deviation in a 15 x 15 x 1 voxel local region using only pixels from the same homogeneous region, and (3) Interpolate the noise estimate from the homogeneous regions in the transition regions. Noise-aware fat segmentation was implemented. Experiments were conducted on the anthropomorphic phantom and in vivo low-dose chest CT scans to validate the TBE, characterize the magnitude of local noise variation, and determine the sensitivity of noise estimates to the size of the region in which noise is computed. The TBE was tested on all scans from the Early Lung Cancer Action Program public database. The TBE was evaluated quantitatively on the phantom data and qualitatively on the in vivo data.
RESULTS: The results show that noise can vary locally by over 200 Hounsfield units on low-dose in vivo chest CT scans and that the TBE can characterize these noise variations within 5 %. The new fat segmentation algorithm successfully improved segmentation on all 50 scans tested.
CONCLUSION: The TBE provides a means to estimate noise for image quality monitoring, optimization of denoising algorithms, and improvement of segmentation algorithms. The TBE was shown to accurately characterize the large local noise variations that occur due to changes in material, dose, and X-ray source location.

Entities:  

Mesh:

Year:  2013        PMID: 23877281     DOI: 10.1007/s11548-013-0930-7

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  10 in total

1.  Local noise weighted filtering for emphysema scoring of low-dose CT images.

Authors:  Arnold M R Schilham; Bram van Ginneken; Hester Gietema; Mathias Prokop
Journal:  IEEE Trans Med Imaging       Date:  2006-04       Impact factor: 10.048

Review 2.  CT dose reduction and dose management tools: overview of available options.

Authors:  Cynthia H McCollough; Michael R Bruesewitz; James M Kofler
Journal:  Radiographics       Date:  2006 Mar-Apr       Impact factor: 5.333

3.  Misoperation of CT automatic tube current modulation systems with inappropriate patient centering: phantom studies.

Authors:  Kosuke Matsubara; Kichiro Koshida; Katsuhiro Ichikawa; Masayuki Suzuki; Tadanori Takata; Tomoyuki Yamamoto; Osamu Matsui
Journal:  AJR Am J Roentgenol       Date:  2009-04       Impact factor: 3.959

4.  Adaptive nonlocal means filtering based on local noise level for CT denoising.

Authors:  Zhoubo Li; Lifeng Yu; Joshua D Trzasko; David S Lake; Daniel J Blezek; Joel G Fletcher; Cynthia H McCollough; Armando Manduca
Journal:  Med Phys       Date:  2014-01       Impact factor: 4.071

5.  A public image database to support research in computer aided diagnosis.

Authors:  A P Reeves; A M Biancardi; D Yankelevitz; S Fotin; B M Keller; A Jirapatnakul; J Lee
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009

Review 6.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

7.  Radiation exposure and image quality in chest CT examinations.

Authors:  J G Ravenel; E M Scalzetti; W Huda; W Garrisi
Journal:  AJR Am J Roentgenol       Date:  2001-08       Impact factor: 3.959

8.  Automatic patient centering for MDCT: effect on radiation dose.

Authors:  Jianhai Li; Unni K Udayasankar; Thomas L Toth; John Seamans; William C Small; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2007-02       Impact factor: 3.959

9.  The influence of patient centering on CT dose and image noise.

Authors:  Thomas Toth; Zhanyu Ge; Michael P Daly
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

10.  Coronary artery calcium: a multi-institutional, multimanufacturer international standard for quantification at cardiac CT.

Authors:  Cynthia H McCollough; Stefan Ulzheimer; Sandra S Halliburton; Kaiss Shanneik; Richard D White; Willi A Kalender
Journal:  Radiology       Date:  2007-05       Impact factor: 11.105

  10 in total
  3 in total

1.  Opportunistic Breast Density Assessment in Women Receiving Low-dose Chest Computed Tomography Screening.

Authors:  Jeon-Hor Chen; Siwa Chan; Nan-Han Lu; Yifan Li; Yu Chieh Tsai; Po Yun Huang; Chia-Ju Chang; Min-Ying Su
Journal:  Acad Radiol       Date:  2016-06-06       Impact factor: 3.173

2.  Large-scale image region documentation for fully automated image biomarker algorithm development and evaluation.

Authors:  Anthony P Reeves; Yiting Xie; Shuang Liu
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-07

3.  Classification of moving coronary calcified plaques based on motion artifacts using convolutional neural networks: a robotic simulating study on influential factors.

Authors:  Magdalena Dobrolińska; Niels van der Werf; Marcel Greuter; Beibei Jiang; Riemer Slart; Xueqian Xie
Journal:  BMC Med Imaging       Date:  2021-10-19       Impact factor: 1.930

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.