Improved accuracy of cardiac output estimation by the partial CO2 rebreathing method.

Yoshifumi Kotake1, Takashige Yamada, Hiromasa Nagata, Takeshi Suzuki, Ryohei Serita, Nobuyuki Katori, Junzo Takeda, Hideyuki Shimizu.   

Abstract

OBJECTIVE: This study investigated the accuracy of the NICO monitor equipped with the newer software. Additionally, the effects of the increased dead space produced by the NICO monitor on ventilatory settings were investigated.
METHODS: Forty-two patients undergoing elective aortic reconstruction participated in this prospective, observational study at a university hospital. Cardiac output was continuously monitored using both the NICO monitor and continuous cardiac output (CCO) measured by a pulmonary artery catheter. A NICO monitor equipped with ver. 4.2 software was used for the first 21 patients while a NICO monitor equipped with ver. 5.0 software was used for the rest of the patients. Cardiac output measured by bolus thermodilution (BCO) at 30 min intervals was used as a reference.
RESULTS: The bias +/- precision of the NICO monitor was 0.18 +/- 0.88 l/min with ver. 4.2 software (n = 182) and 0.18 +/- 0.83 l/min with 5.0 software (n = 194). The accuracy of the NICO monitor is comparable to CCO, whose bias +/- precision against BCO is 0.19 +/- 0.81 l/min (n = 376). At the same level of CO(2) production and minute ventilation, PaCO(2) was lower in the patients monitored by NICO with ver. 5.0 software than patients with ver. 4.2 software.
CONCLUSIONS: This study demonstrated the improved performance of the NICO monitor with updated software. The performance of the NICO monitor with ver. 4.2 or later software is similar to CCO. However, the cardiac output measurement did not fulfill the criteria of interchangeability to the cardiac output measurement by bolus thermodilution. Updates to ver. 5.0 attenuated the effects of rebreathing introduced by the NICO monitor without compromising the accuracy of the cardiac output measurement.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19301132     DOI: 10.1007/s10877-009-9172-1

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  21 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Partial CO2 rebreathing cardiac output--operating principles of the NICO system.

Authors:  M B Jaffe
Journal:  J Clin Monit Comput       Date:  1999-08       Impact factor: 2.502

3.  The relationship between cardiac output measured by the thermodilution method and that measured by the carbon dioxide rebreathing technique during laparoscopic surgery.

Authors:  Manzo Suzuki; Syusuke Koda; Yoshihisa Nakamura; Naoki Kawamura; Yoichi Shimada
Journal:  Anesth Analg       Date:  2005-05       Impact factor: 5.108

4.  Lack of agreement between thermodilution and carbon dioxide-rebreathing cardiac output.

Authors:  L B Nilsson; N Eldrup; P G Berthelsen
Journal:  Acta Anaesthesiol Scand       Date:  2001-07       Impact factor: 2.105

5.  Clinical evaluation of a partial CO2 rebreathing technique for cardiac output monitoring in critically ill patients.

Authors:  H Odenstedt; O Stenqvist; S Lundin
Journal:  Acta Anaesthesiol Scand       Date:  2002-02       Impact factor: 2.105

6.  A comparative evaluation of thermodilution and partial CO2 rebreathing techniques for cardiac output assessment in critically ill patients during assisted ventilation.

Authors:  Monica Rocco; Gustavo Spadetta; Andrea Morelli; Donatella Dell'Utri; Patrizia Porzi; Giorgio Conti; Paolo Pietropaoli
Journal:  Intensive Care Med       Date:  2003-12-03       Impact factor: 17.440

7.  Comparison of continuous cardiac output measurements in patients after cardiac surgery.

Authors:  Frank Mielck; Wolfgang Buhre; Gunnar Hanekop; Theodor Tirilomis; Reinhard Hilgers; Hans Sonntag
Journal:  J Cardiothorac Vasc Anesth       Date:  2003-04       Impact factor: 2.628

8.  Multicenter trial of a new thoracic electrical bioimpedance device for cardiac output estimation.

Authors:  W C Shoemaker; C C Wo; M H Bishop; P L Appel; J M Van de Water; G R Harrington; X Wang; R S Patil
Journal:  Crit Care Med       Date:  1994-12       Impact factor: 7.598

9.  Noninvasive cardiac output assessment during heart surgery.

Authors:  Piero Ceriana; Marco Mau Relli; Antonio Braschi; Paola Baiardi; Donatella De Amici
Journal:  J Clin Monit Comput       Date:  2002-02       Impact factor: 2.502

10.  Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present.

Authors:  Kazuya Tachibana; Hideaki Imanaka; Muneyuki Takeuchi; Yuji Takauchi; Hiroshi Miyano; Masaji Nishimura
Journal:  Anesthesiology       Date:  2003-04       Impact factor: 7.892

View more
  6 in total

1.  Can a NICO monitor substitute for thermodilution to measure cardiac output in patients with coexisting tricuspid regurgitation?

Authors:  Naka Imakiire; Takeshi Omae; Akira Matsunaga; Ryuzo Sakata; Yuichi Kanmura
Journal:  J Anesth       Date:  2010-05-19       Impact factor: 2.078

Review 2.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

Review 3.  Noninvasive Monitoring and Potential for Patient Outcome.

Authors:  Susana Vacas; Maxime Cannesson
Journal:  J Cardiothorac Vasc Anesth       Date:  2019-08       Impact factor: 2.628

4.  Agreement between continuous and intermittent pulmonary artery thermodilution for cardiac output measurement in perioperative and intensive care medicine: a systematic review and meta-analysis.

Authors:  Karim Kouz; Frederic Michard; Alina Bergholz; Christina Vokuhl; Luisa Briesenick; Phillip Hoppe; Moritz Flick; Gerhard Schön; Bernd Saugel
Journal:  Crit Care       Date:  2021-03-29       Impact factor: 9.097

5.  Cardiac output assessed by invasive and minimally invasive techniques.

Authors:  Allison J Lee; Jennifer Hochman Cohn; J Sudharma Ranasinghe
Journal:  Anesthesiol Res Pract       Date:  2011-07-06

Review 6.  Non-Invasive Monitoring of Cardiac Output in Critical Care Medicine.

Authors:  Lee S Nguyen; Pierre Squara
Journal:  Front Med (Lausanne)       Date:  2017-11-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.