Comparison of continuous cardiac output measurements in patients after cardiac surgery.

Frank Mielck1, Wolfgang Buhre, Gunnar Hanekop, Theodor Tirilomis, Reinhard Hilgers, Hans Sonntag.   

Abstract

OBJECTIVE: To investigate in a direct comparison accuracy and precision of continuous cardiac output measurements assessed by continuous pulmonary artery thermodilution technique (TDCCO), continuous pulse contour analysis (PCCO), and noninvasive partial CO(2)-rebreathing technique (NICO) in patients after coronary artery bypass grafting (CABG) during the postoperative period.
DESIGN: Prospective, controlled clinical study.
SETTING: University hospital. PARTICIPANTS: Twenty-two patients undergoing elective CABG surgery.
INTERVENTIONS: Hemodynamic measurements were performed after admission to the ICU and in sequence every 2 hours during the subsequent 6-hour period. Simultaneously, cardiac output (CO) was measured using a TDCCO, PCCO, and NICO. After the continuous cardiac output measurements were read, bolus thermodilution-derived cardiac output was obtained from thermodilution curves detected in the pulmonary artery (TDBCO(pa)). Four intermittent consecutive boli consisting of 10 mL of ice-cold saline were randomly injected over the ventilatory cycle.
MEASUREMENTS AND MAIN RESULTS: The comparison between the continuous cardiac output measurement methods TDCCO versus PCCO showed a bias of -0.12 L/min, between TDCCO versus NICO -0.17 L/min, and between PCCO versus NICO -0.44 L/min. The comparison to the reference technique between TDBCO(pa) versus TDCCO revealed a bias of -0.28 L/min, between TDBCO(pa) versus PCCO -0.40 L/min, and between TDBCO(pa) versus NICO -0.64 L/min.
CONCLUSIONS: The results of this clinical investigation show agreement between TDCCO and PCCO to satisfy clinical requirements in a setting of postoperative patients after cardiac surgery. In contrast, the NICO monitor is of very limited use in these patients. Copyright 2003 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12698404     DOI: 10.1053/jcan.2003.49

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  17 in total

Review 1.  [Measurement of cardiac output].

Authors:  D A Reuter; A E Goetz
Journal:  Anaesthesist       Date:  2005-11       Impact factor: 1.041

2.  Correlation between cardiac output measured by the femoral arterial thermodilution technique pulmonary arterial and that measured by contour pulse analysis in a paediatric animal model.

Authors:  Jesús López-Herce; Marta Rupérez; César Sánchez; Cristina García; Elena García
Journal:  J Clin Monit Comput       Date:  2006-02-28       Impact factor: 2.502

Review 3.  [Pulmonary artery catheter in anaesthesiology and intensive care medicine].

Authors:  E E C de Waal; L de Rossi; W Buhre
Journal:  Anaesthesist       Date:  2006-06       Impact factor: 1.041

Review 4.  [Haemodynamic monitoring in the perioperative phase. Available systems, practical application and clinical data].

Authors:  U Wittkowski; C Spies; M Sander; J Erb; A Feldheiser; C von Heymann
Journal:  Anaesthesist       Date:  2009-08       Impact factor: 1.041

Review 5.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

Review 6.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

7.  [Accuracy of pulse contour cardiac index measurements during changes of preload and aortic impedance].

Authors:  T W Felbinger; M S Goepfert; T Goresch; A E Goetz; D A Reuter
Journal:  Anaesthesist       Date:  2005-08       Impact factor: 1.041

8.  Accuracy and precision of transcardiopulmonary thermodilution in patients with cardiogenic shock.

Authors:  Bonaventura Schmid; Katrin Fink; Manfred Olschewski; Stephan Richter; Tilmann Schwab; Michael Brunner; Hans-Joerg Busch
Journal:  J Clin Monit Comput       Date:  2015-10-01       Impact factor: 2.502

9.  Improved accuracy of cardiac output estimation by the partial CO2 rebreathing method.

Authors:  Yoshifumi Kotake; Takashige Yamada; Hiromasa Nagata; Takeshi Suzuki; Ryohei Serita; Nobuyuki Katori; Junzo Takeda; Hideyuki Shimizu
Journal:  J Clin Monit Comput       Date:  2009-03-20       Impact factor: 2.502

10.  Comparison of two methods for enhanced continuous circulatory monitoring in patients with septic shock.

Authors:  F Spöhr; P Hettrich; H Bauer; U Haas; E Martin; B W Böttiger
Journal:  Intensive Care Med       Date:  2007-06-05       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.