Multicenter trial of a new thoracic electrical bioimpedance device for cardiac output estimation.

W C Shoemaker1, C C Wo, M H Bishop, P L Appel, J M Van de Water, G R Harrington, X Wang, R S Patil.   

Abstract

OBJECTIVE: To evaluate the capacity of a new thoracic electric bioimpedance system to estimate cardiac output compared with the conventional thermodilution method.
DESIGN: Prospective, multicenter study.
SETTING: A university-run county hospital, a university-run U.S. Veterans Affairs hospital, and a university-affiliated U.S. military hospital. PATIENTS: A series of 68 critically ill patients whose conditions required pulmonary artery catheter insertion.
MEASUREMENTS AND MAIN RESULTS: A total of 842 simultaneous pairs of cardiac output estimations by conventional thermodilution and a new thoracic electric bioimpedance system that uses an improved signal processing technique based on an all-integer-coefficient filtering technology, using a time-frequency distribution that provides a high signal/noise ratio were evaluated. The r value was .86, r2 = .74, and p < .001 by regression analysis; the mean difference between the two methods relative to their average value was 16.6 +/- 12.9 (SD) %; the precision was 1.4 L/min or 0.8 L/min/m2; the bias was -0.013 L/min. The mean difference between successive pairs of thermodilution measurements was 8.6 +/- 0.6 (SD) %, which was about half the difference between simultaneous pairs of measurement by the two methods. The changes in impedance estimations were close to simultaneously measured changes in thermodilution estimates of cardiac output.
CONCLUSIONS: The new bioimpedance system satisfactorily estimated cardiac output as measured by the thermodilution technique. The difference between the two estimations is more than made up for by the continuous noninvasive capability of the impedance system.

Entities:  

Mesh:

Year:  1994        PMID: 7988125

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  17 in total

1.  Analysis of the O-wave in acute right ventricular apex impedance measurements with a standard pacing lead in animals.

Authors:  K Järverud; S Ollmar; L A Brodin
Journal:  Med Biol Eng Comput       Date:  2002-09       Impact factor: 2.602

2.  A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.

Authors:  L A Critchley; J A Critchley
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

3.  Evaluation of a new advanced thoracic bioimpedance device for estimation of cardiac output.

Authors:  D G Haryadi; D R Westenskow; L A Critchley; S I Schookin; V G Zubenko; K R Beliaev; A A Morozov
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

4.  Comparison of transpulmonary thermodilution and ultrasound dilution technique: novel insights into volumetric parameters from an animal model.

Authors:  Martin Boehne; Florian Schmidt; Lars Witt; Harald Köditz; Michael Sasse; Robert Sümpelmann; Harald Bertram; Armin Wessel; Wilhelm Alexander Osthaus
Journal:  Pediatr Cardiol       Date:  2012-02-14       Impact factor: 1.655

5.  Can a NICO monitor substitute for thermodilution to measure cardiac output in patients with coexisting tricuspid regurgitation?

Authors:  Naka Imakiire; Takeshi Omae; Akira Matsunaga; Ryuzo Sakata; Yuichi Kanmura
Journal:  J Anesth       Date:  2010-05-19       Impact factor: 2.078

6.  Invasive and noninvasive hemodynamic monitoring of patients with cerebrovascular accidents.

Authors:  G C Velmahos; C C Wo; D Demetriades; M H Bishop; W C Shoemaker
Journal:  West J Med       Date:  1998-07

Review 7.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

8.  Impedance cardiography as a noninvasive technique for atrioventricular interval optimization in cardiac resynchronization therapy.

Authors:  Martin U Braun; Andreas Schnabel; Thomas Rauwolf; Matthias Schulze; Ruth H Strasser
Journal:  J Interv Card Electrophysiol       Date:  2005-09       Impact factor: 1.900

9.  Endpoints of resuscitation of critically injured patients: normal or supranormal? A prospective randomized trial.

Authors:  G C Velmahos; D Demetriades; W C Shoemaker; L S Chan; R Tatevossian; C C Wo; P Vassiliu; E E Cornwell; J A Murray; B Roth; H Belzberg; J A Asensio; T V Berne
Journal:  Ann Surg       Date:  2000-09       Impact factor: 12.969

10.  Short-term postural reflexes in diabetic patients with autonomic dysfunction.

Authors:  Y Zhang; L A Critchley; Y H Tam; B Tomlinson
Journal:  Diabetologia       Date:  2003-12-03       Impact factor: 10.122

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.