Noninvasive cardiac output assessment during heart surgery.

Piero Ceriana1, Marco Mau Relli, Antonio Braschi, Paola Baiardi, Donatella De Amici.   

Abstract

OBJECTIVE: To evaluate the reliability of a new noninvasive method for the assessment of cardiac output with the partial carbon dioxide rebreathing technique.
METHODS: This technique was applied to patients undergoing heart surgery. Values of cardiac index obtained with this equipment were compared with the artero-venous CO2 gradient, a reliable index of cardiovascular status. Positive and negative predictive values of the test were assessed.
RESULTS: A total of 21 simultaneous measurement of the cardiac index and of the artero-venous CO2 gradient were obtained. The positive predictive value of the test was 67% while the negative predictive value was 100%, indicating that a normal value of cardiac index recorded with the rebreathing technique predicts with a good reliability a normal cardiovascular state.
CONCLUSIONS: Working through a series of mathematical algorithms, accuracy in the computation of cardiac output can be decreased with this equipment; however, this limitation seems to be outweighed by the simplicity and noninvasive nature of the methods.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12212994     DOI: 10.1023/a:1016373724642

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  5 in total

1.  Noninvasive measurement of cardiac output using partial CO2 rebreathing.

Authors:  J M Capek; R J Roy
Journal:  IEEE Trans Biomed Eng       Date:  1988-09       Impact factor: 4.538

2.  A reassessment of the validity of the iso-shunt graph.

Authors:  P G Lawler; J F Nunn
Journal:  Br J Anaesth       Date:  1984-12       Impact factor: 9.166

3.  Non-invasive cardiac output determination: comparison of a new partial-rebreathing technique with thermodilution.

Authors:  J C Binder; W G Parkin
Journal:  Anaesth Intensive Care       Date:  2001-02       Impact factor: 1.669

4.  Lack of agreement between thermodilution and carbon dioxide-rebreathing cardiac output.

Authors:  L B Nilsson; N Eldrup; P G Berthelsen
Journal:  Acta Anaesthesiol Scand       Date:  2001-07       Impact factor: 2.105

5.  Assessing acid-base status in circulatory failure. Differences between arterial and central venous blood.

Authors:  H J Adrogué; M N Rashad; A B Gorin; J Yacoub; N E Madias
Journal:  N Engl J Med       Date:  1989-05-18       Impact factor: 91.245

  5 in total
  3 in total

1.  [Non-invasive extended hemodynamic monitoring. Reduction of circulatory risk situations].

Authors:  M Bock; T Sturm; J Motsch
Journal:  Anaesthesist       Date:  2007-07       Impact factor: 1.041

2.  An on line technique to detect cardiac output variations and cardiovascular performances during abdominal aortic surgery.

Authors:  Fabrizio Clemente; Emilio D'Avino; Marco Guaragno; Antonio Menichetti
Journal:  J Clin Monit Comput       Date:  2004-04       Impact factor: 2.502

3.  Improved accuracy of cardiac output estimation by the partial CO2 rebreathing method.

Authors:  Yoshifumi Kotake; Takashige Yamada; Hiromasa Nagata; Takeshi Suzuki; Ryohei Serita; Nobuyuki Katori; Junzo Takeda; Hideyuki Shimizu
Journal:  J Clin Monit Comput       Date:  2009-03-20       Impact factor: 2.502