Thermal stress and point-of-care testing performance: suitability of glucose test strips and blood gas cartridges for disaster response.

Richard F Louie1, Stephanie L Sumner, Shaunyé Belcher, Ron Mathew, Nam K Tran, Gerald J Kost.   

Abstract

OBJECTIVE: Point-of-care testing (POCT) devices are deployed in the field for emergency on-site testing under a wide range of environmental conditions. Our objective was to evaluate the performance of glucose meter test strips and handheld blood gas analyzer cartridges following thermal stresses that simulate field conditions.
METHODS: We evaluated electrochemical and spectrophotometric glucose meter systems and a handheld blood gas analyzer. Glucose test strips were cold-stressed (-21 degrees C) and heat-stressed (40 degrees C) for up to 4 weeks. Blood gas cartridges were stressed at -21 degrees C, 2 degrees C, and 40 degrees C for up to 72 hours. Test strip and cartridge performance was evaluated using aqueous quality control solutions. Results were compared with those obtained with unstressed POCT strips and cartridges.
RESULTS: Heated glucose test strips and blood gas cartridges yielded elevated results. Frozen test strips and cooled cartridges yielded depressed glucose and blood gas results, respectively. Frozen cartridges failed.
CONCLUSIONS: The performance of glucose test strips and blood gas cartridges was affected adversely by thermal stresses. Heating generated elevated results, and cooling depressed results. Disaster medical assistance teams and emergency medical responders should be aware of these risks. Field POCT devices must be robust to withstand adverse conditions. We recommend that industry produce POCT devices and reagents suitable for disaster medical assistance teams.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19293739      PMCID: PMC2717087          DOI: 10.1097/DMP.0b013e3181979a06

Source DB:  PubMed          Journal:  Disaster Med Public Health Prep        ISSN: 1935-7893            Impact factor:   1.385


  14 in total

1.  Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.

Authors:  Z Tang; X Du; R F Louie; G J Kost
Journal:  Am J Clin Pathol       Date:  2000-01       Impact factor: 2.493

2.  Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing.

Authors:  Z Tang; J H Lee; R F Louie; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-08       Impact factor: 5.534

3.  Thermal inactivation of glucose oxidase. Mechanism and stabilization using additives.

Authors:  Mudeppa Devaraja Gouda; Sridevi Annapurna Singh; A G Appu Rao; Munna Singh Thakur; Naikankatte Ganesh Karanth
Journal:  J Biol Chem       Date:  2003-04-26       Impact factor: 5.157

4.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

5.  Oxygen effects on glucose measurements with a reference analyzer and three handheld meters.

Authors:  Z Tang; R F Louie; M Payes; K C Chang; G J Kost
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

6.  Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.

Authors:  Z Tang; R F Louie; J H Lee; D M Lee; E E Miller; G J Kost
Journal:  Crit Care Med       Date:  2001-05       Impact factor: 7.598

7.  Effect of high altitude on blood glucose meter performance.

Authors:  Kenneth S Fink; Dale B Christensen; Allan Ellsworth
Journal:  Diabetes Technol Ther       Date:  2002       Impact factor: 6.118

8.  Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada.

Authors:  G J Kost; H T Vu; J H Lee; P Bourgeois; F L Kiechle; C Martin; S S Miller; A O Okorodudu; J J Podczasy; R Webster; K J Whitlow
Journal:  Crit Care Med       Date:  1998-03       Impact factor: 7.598

9.  Effect of adverse storage conditions on performance of glucometer test strips.

Authors:  Richard Bamberg; Kathleen Schulman; Melissa MacKenzie; Joanna Moore; Sarah Olchesky
Journal:  Clin Lab Sci       Date:  2005

10.  Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curves.

Authors:  Gerald J Kost; Nam K Tran; Victor J Abad; Richard F Louie
Journal:  Clin Chim Acta       Date:  2007-12-03       Impact factor: 3.786

View more
  9 in total

1.  The Effects of Temperature and Relative Humidity on Point-of-Care Glucose Measurements in Hospital Practice in a Tropical Clinical Setting.

Authors:  Busadee Pratumvinit; Nattakom Charoenkoop; Soamsiri Niwattisaiwong; Gerald J Kost; Panutsaya Tientadakul
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

2.  Hematocrit interference of blood glucose meters for patient self-measurement.

Authors:  Sanja Ramljak; John Paul Lock; Christina Schipper; Petra B Musholt; Thomas Forst; Martha Lyon; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

3.  Evidence-Based Point-of-Care Device Design for Emergency and Disaster Care.

Authors:  Daniel M Mecozzi; T Keith Brock; Nam K Tran; Kristin N Hale; Gerald J Kost
Journal:  Point Care       Date:  2010-06-01

4.  Evidence-based point-of-care tests and device designs for disaster preparedness.

Authors:  T Keith Brock; Daniel M Mecozzi; Stephanie Sumner; Gerald J Kost
Journal:  Am J Disaster Med       Date:  2010 Sep-Oct

5.  Enhancing crisis standards of care using innovative point-of-care testing.

Authors:  Gerald J Kost; Ann Sakaguchi; Corbin Curtis; Nam K Tran; Pratheep Katip; Richard F Louie
Journal:  Am J Disaster Med       Date:  2011 Nov-Dec

6.  Future Connectivity for Disaster and Emergency Point of Care.

Authors:  Jimmy N Yu; Terry Keith Brock; Daniel M Mecozzi; Nam K Tran; Gerald J Kost
Journal:  Point Care       Date:  2010-12

7.  Short-Term Thermal-Humidity Shock Affects Point-of-Care Glucose Testing: Implications for Health Professionals and Patients.

Authors:  Mandy Lam; Richard F Louie; Corbin M Curtis; William J Ferguson; John H Vy; Anh-Thu Truong; Stephanie L Sumner; Gerald J Kost
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

Review 8.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

9.  Point-of-care testing for disasters: needs assessment, strategic planning, and future design.

Authors:  Gerald J Kost; Kristin N Hale; T Keith Brock; Richard F Louie; Nicole L Gentile; Tyler K Kitano; Nam K Tran
Journal:  Clin Lab Med       Date:  2009-09       Impact factor: 1.935

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.