Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing.

Z Tang1, J H Lee, R F Louie, G J Kost.   

Abstract

OBJECTIVES: To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors.
DESIGN: Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared.
SETTING: Six hand-held glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL).
RESULTS: Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration.
CONCLUSIONS: We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10923072     DOI: 10.5858/2000-124-1135-EODHLO

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  50 in total

1.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

Review 2.  Challenges of inpatient blood glucose monitoring: standards, methods, and devices to measure blood glucose.

Authors:  Kathie L Hermayer; Aundrea S Loftley; Sumana Reddy; Satya Nandana Narla; Nina A Epps; Yusheng Zhu
Journal:  Curr Diab Rep       Date:  2015-03       Impact factor: 4.810

Review 3.  Glycemic control in the burn intensive care unit: focus on the role of anemia in glucose measurement.

Authors:  Elizabeth A Mann; Alejandra G Mora; Heather F Pidcoke; Steven E Wolf; Charles E Wade
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

4.  Hematocrit interference of blood glucose meters for patient self-measurement.

Authors:  Sanja Ramljak; John Paul Lock; Christina Schipper; Petra B Musholt; Thomas Forst; Martha Lyon; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

5.  Performance variability of seven commonly used self-monitoring of blood glucose systems: clinical considerations for patients and providers.

Authors:  Ronald L Brazg; Leslie J Klaff; Christopher G Parkin
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

6.  Determination of hematocrit interference in blood samples derived from patients with different blood glucose concentrations.

Authors:  Andreas Pfützner; Christina Schipper; Sanja Ramljak; Frank Flacke; Jochen Sieber; Thomas Forst; Petra B Musholt
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

7.  Evaluation of hematocrit interference with MyStar extra and seven competitive devices.

Authors:  Filiz Demircik; Sanja Ramljak; Iris Hermanns; Anke Pfützner; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2014-12-30

8.  Comparative Accuracy of 17 Point-of-Care Glucose Meters.

Authors:  Laya Ekhlaspour; Debbie Mondesir; Norman Lautsch; Courtney Balliro; Mallory Hillard; Kendra Magyar; Laura Goergen Radocchia; Aryan Esmaeili; Manasi Sinha; Steven J Russell
Journal:  J Diabetes Sci Technol       Date:  2016-10-03

9.  Thermal stress and point-of-care testing performance: suitability of glucose test strips and blood gas cartridges for disaster response.

Authors:  Richard F Louie; Stephanie L Sumner; Shaunyé Belcher; Ron Mathew; Nam K Tran; Gerald J Kost
Journal:  Disaster Med Public Health Prep       Date:  2009-03       Impact factor: 1.385

10.  Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curves.

Authors:  Gerald J Kost; Nam K Tran; Victor J Abad; Richard F Louie
Journal:  Clin Chim Acta       Date:  2007-12-03       Impact factor: 3.786

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.