Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.

Z Tang1, X Du, R F Louie, G J Kost.   

Abstract

Thirty drugs used primarily in critical care and hospital settings were tested in vitro to observe interference on glucose measurements with 6 hand-held glucose meters and a portable glucose analyzer. Paired differences of glucose measurements between drug-spiked samples and unspiked control samples were calculated to determine bias. A criterion of +/- 6 mg/dL was used as the cutoff for interference. Ascorbic acid interfered with the measurements on all glucose devices evaluated. Acetaminophen, dopamine, and mannitol interfered with glucose measurements on some devices. Dose-response relationships help assessment of drug interference in clinical use. High dosages of these drugs may be given to critically ill patients or self-administered by patients without medical supervision. Package inserts for the glucose devices may not provide adequate warning information. Hence, we recommend that clinicians choose glucose devices carefully and interpret results cautiously when glucose measurements are performed during or after drug interventions.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10631860     DOI: 10.1309/QAW1-X5XW-BVRQ-5LKQ

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  46 in total

1.  Glucose information for tight glycemic control: different methods with different challenges.

Authors:  Christian Weber; Kurt Neeser
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

3.  Impact of High-Dose Intravenous Vitamin C for Treatment of Sepsis on Point-of-Care Blood Glucose Readings.

Authors:  Anna Peyton Howell; Jenna L Parrett; Daniel R Malcom
Journal:  J Diabetes Sci Technol       Date:  2019-11-25

4.  No effect of acidification or freezing on urinary metanephrine levels.

Authors:  L C Pinto; C K Kramer; J L Camargo; L H Canani; J L Gross; C B Leitão
Journal:  J Endocrinol Invest       Date:  2019-06-21       Impact factor: 4.256

Review 5.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

6.  Hematocrit interference of blood glucose meters for patient self-measurement.

Authors:  Sanja Ramljak; John Paul Lock; Christina Schipper; Petra B Musholt; Thomas Forst; Martha Lyon; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

7.  Accuracy of bedside capillary blood glucose measurements in critically ill patients.

Authors:  C Dana Critchell; Vincent Savarese; Amy Callahan; Christine Aboud; Serge Jabbour; Paul Marik
Journal:  Intensive Care Med       Date:  2007-09-01       Impact factor: 17.440

8.  In Vitro Sugar Interference Testing With Amperometric Glucose Oxidase Sensors.

Authors:  Ryan Boehm; John Donovan; Disha Sheth; Andrew Durfor; Jason Roberts; Irada Isayeva
Journal:  J Diabetes Sci Technol       Date:  2018-08-03

9.  Thermal stress and point-of-care testing performance: suitability of glucose test strips and blood gas cartridges for disaster response.

Authors:  Richard F Louie; Stephanie L Sumner; Shaunyé Belcher; Ron Mathew; Nam K Tran; Gerald J Kost
Journal:  Disaster Med Public Health Prep       Date:  2009-03       Impact factor: 1.385

10.  GlucoMen Day continuous glucose monitoring system: a screening for enzymatic and electrochemical interferents.

Authors:  Fausto Lucarelli; Francesco Ricci; Felice Caprio; Francesco Valgimigli; Cosimo Scuffi; Danila Moscone; Giuseppe Palleschi
Journal:  J Diabetes Sci Technol       Date:  2012-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.