| Literature DB >> 19223841 |
Guanghua Yan1,2, Chihray Liu1, Thomas A Simon1,2, Lee-Cheng Peng1,2, Christopher Fox1, Jonathan G Li1.
Abstract
Accurate multileaf collimator (MLC) leaf positioning plays an essential role in the effective implementation of intensity modulated radiation therapy (IMRT). This work evaluates the sensitivity of current patient-specific IMRT quality assurance (QA) procedures to minor MLC leaf positioning errors. Random errors of up to 2 mm and systematic errors of +/-1 mm and +/-2 mm in MLC leaf positions were introduced into 8 clinical IMRT patient plans (totaling 53 fields). Planar dose distributions calculated with modified plans were compared to dose distributions measured with both radiochromic films and a diode matrix. The agreement between calculation and measurement was evaluated using both absolute distance-to-agreement (DTA) analysis and gamma index with 2%/2 mm and 3%/3 mm criteria. It was found that both the radiochromic film and the diode matrix could only detect systematic errors on the order of 2 mm or above. The diode array had larger sensitivity than film due to its excellent detector response (such as small variation, linear response, etc.). No difference was found between DTA analysis and gamma index in terms of the sensitivity to MLC positioning errors. Higher sensitivity was observed with 2%/2 mm than with 3%/3 mm in general. When using the diode array and 2%/2 mm criterion, the IMRT QA procedure showed strongest sensitivity to MLC position errors and, at the same time, achieved clinically acceptable passing rates. More accurate dose calculation and measurement would further enhance the sensitivity of patient-specific IMRT QA to MLC positioning errors. However, considering the significant dosimetric effect such MLC errors could cause, patient-specific IMRT QA should be combined with a periodic MLC QA program in order to guarantee the accuracy of IMRT delivery.Entities:
Mesh:
Year: 2009 PMID: 19223841 PMCID: PMC5720508 DOI: 10.1120/jacmp.v10i1.2915
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Average passing rates when comparing Pinnacle calculated planar dose distribution and Gafchromic Ebt film measurement using and criterion with (a) Dta analysis and (b) the γ index for a total of 10 Imrt fields. The x‐axis, from left to right, stands for no MLC positioning error, random errors and systematic errors of and , respectively. The error bar indicates one standard deviation.
Figure 2Average passing rates when comparing Pinnacle calculated planar dose distribution and MapCHECK measurement using and criterion with (a) DTA analysis and (b) the γ index for a total of 53 IMRT fields. The x‐axis, from left to right, stands for no MLC positioning error, random errors and systematic errors of and , respectively. The error bar indicates one standard deviation.
Figure 3Drop of average passing rates with respect to the clinical plan (no MLC positioning errors) with the introduction of random and systematic errors. The comparison was between Pinnacle calculated planar dose distribution and (a) film or (b) MapCHECK measurement using DTA and as well as γ and criterion. The x‐axis, from left to right, stands for no MLC positioning error, random errors and systematic errors of and , respectively.
P‐values from two‐tailed Wilcoxon rank‐sum test. The test compared the distribution of passing rates before and after the introduction of Mlc position errors.
|
|
|
|
|
| |||
|---|---|---|---|---|---|---|---|
| EBT Film | DTA |
| 3.28E‐01 | 9.59E‐01 | 1.05E‐01 | 2.81E‐02 | 6.22E‐04 |
|
| 5.56E‐01 | 7.98E‐01 | 1.61E‐01 | 7.74E‐02 | 1.10E‐03 | ||
| Gamma |
| 5.05E‐01 | 9.81E‐01 | 1.30E‐01 | 2.81E‐02 | 6.22E‐04 | |
|
| 6.29E‐01 | 1.00E+00 | 3.14E‐01 | 9.95E‐02 | 1.90E‐03 | ||
| MapCHECK | DTA |
| 1.10E‐02 | 1.29E‐05 | 2.60E‐03 | 2.35E‐11 | 6.86E‐10 |
|
| 3.20E‐02 | 1.90E‐05 | 3.60E‐03 | 2.84E‐11 | 1.62E‐08 | ||
| Gamma |
| 1.76E‐02 | 1.08E‐05 | 3.45E‐04 | 3.97E‐11 | 3.28E‐10 | |
|
| 7.59E‐05 | 3.85E‐06 | 3.57E‐04 | 4.33E‐12 | 9.20E‐10 |
Figure 4Average passing rates when comparing error free plans (re‐sampled according to MapCHECK detector positions) to plans with and systematic MLC errors using DTA criterion. The error bar stands for one standard deviation.