Wanyong Shin1, Hong Gu, Yihong Yang.
Abstract
A fast T(1) measurement sequence using inversion recovery Look-Locker echo-planar imaging at steady state (IR LL-EPI SS) is presented. Delay time for a full magnetization recovery is not required in the sequence, saving acquisition time significantly for high-resolution T(1) mapping. Imaging parameters of the IR LL-EPI SS sequence were optimized to minimize the bias from the excitation pulses imperfection and to maximize the accuracy and reliability of T(1) measurements, which are critical for its applications. Compared with the conventional inversion recovery Look-Locker echo-planar imaging (IR LL-EPI) sequence, IR LL-EPI SS method preserves similar accuracy and reliability, while saving 20% in acquisition time. Optimized IR LL-EPI SS provided quantitative T(1) mapping with 1 x 1 x 4 mm(3) resolution and whole-brain coverage (28 slices) in approximately 4 min.Entities:
Mesh:
Year: 2009 PMID: 19195021 PMCID: PMC2663603 DOI: 10.1002/mrm.21836
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668