J Haselgrove1, J Moore, Z Wang, E Traipe, L Bilaniuk.
Abstract
We have developed a multislice protocol for quantitative T1 measurements in which the processing time and the acquisition time are under 2 minutes each for a complete brain study of 15 slices. An echoplanar, inversion-recovery image sequence is designed to collect data suitable for analysis using a linear regression algorithm. The precision is approximately twice the noise to signal ratio of the images. The accuracy of the protocol is better than 1% for T1 in the range 0-2 seconds and deviates slightly for longer T1 values. The protocol is insensitive to B1 field values. If needed, the data can be postprocessed using a slow, nonlinear algorithm to give an accuracy of less than 1% and a precision of approximately the noise to signal ratio throughout a range of T1 values from 0 to 4 seconds. Copyright 2000 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2000 PMID: 10767064 DOI: 10.1002/(sici)1522-2586(200004)11:4<360::aid-jmri3>3.0.co;2-g
Source DB: PubMed Journal: J Magn Reson Imaging ISSN: 1053-1807 Impact factor: 4.813