Quality and accuracy of cone beam computed tomography gated by active breathing control.

Bria P Thompson1, Geoffrey D Hugo.   

Abstract

The purpose of this study was to evaluate the quality and accuracy of cone beam computed tomography (CBCT) gated by active breathing control (ABC), which may be useful for image guidance in the presence of respiration. Comparisons were made between conventional ABC-CBCT (stop and go), fast ABC-CBCT (a method to speed up the acquisition by slowing the gantry instead of stopping during free breathing), and free breathing respiration correlated CBCT. Image quality was assessed in phantom. Accuracy of reconstructed voxel intensity, uniformity, and root mean square error were evaluated. Registration accuracy (bony and soft tissue) was quantified with both an anthropomorphic and a quality assurance phantom. Gantry angle accuracy was measured with respect to gantry speed modulation. Conventional ABC-CBCT scan time ranged from 2.3 to 5.8 min. Fast ABC-CBCT scan time ranged from 1.4 to 1.8 min, and respiratory correlated CBCT scans took 2.1 min to complete. Voxel intensity value for ABC gated scans was accurate relative to a normal clinical scan with all projections. Uniformity and root mean square error performance degraded as the number of projections used in the reconstruction of the fast ABC-CBCT scans decreased (shortest breath hold, longest free breathing segment). Registration accuracy for small, large, and rotational corrections was within 1 mm and 1 degrees. Gantry angle accuracy was within 1 degrees for all scans. For high-contrast targets, performance for image-guidance purposes was similar for fast and conventional ABC-CBCT scans and respiration correlated CBCT.

Mesh:

Year:  2008        PMID: 19175117      PMCID: PMC2673601          DOI: 10.1118/1.3013568

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  Optimizing 4D cone-beam CT acquisition protocol for external beam radiotherapy.

Authors:  Tianfang Li; Lei Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-12-29       Impact factor: 7.038

2.  Iterative sorting for four-dimensional CT images based on internal anatomy motion.

Authors:  Rongping Zeng; Jeffrey A Fessler; James M Balter; Peter A Balter
Journal:  Med Phys       Date:  2008-03       Impact factor: 4.071

3.  Enhanced 4D cone-beam CT with inter-phase motion model.

Authors:  Tianfang Li; Albert Koong; Lei Xing
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

4.  Active breathing control (ABC): determination and reduction of breathing-induced organ motion in the chest.

Authors:  Bernd Gagel; Cengiz Demirel; Aline Kientopf; Michael Pinkawa; Marc Piroth; Sven Stanzel; Christian Breuer; Branka Asadpour; Thomas Jansen; Richard Holy; Joachim E Wildberger; Michael J Eble
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-12-29       Impact factor: 7.038

5.  Reduction of organ motion in lung tumors with respiratory gating.

Authors:  Philippe Giraud; Ellen Yorke; Eric C Ford; Raquel Wagman; Gig S Mageras; Howard Amols; Clifton C Ling; Kenneth E Rosenzweig
Journal:  Lung Cancer       Date:  2005-09-28       Impact factor: 5.705

6.  Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position.

Authors:  Thomas G Purdie; Jean-Pierre Bissonnette; Kevin Franks; Andrea Bezjak; David Payne; Fanny Sie; Michael B Sharpe; David A Jaffray
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-02-27       Impact factor: 7.038

7.  Short-term and long-term reproducibility of lung tumor position using active breathing control (ABC).

Authors:  Rojano Kashani; Rojano Koshani; James M Balter; James A Hayman; George T Henning; Marcel van Herk
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-08-01       Impact factor: 7.038

8.  Influence of retrospective sorting on image quality in respiratory correlated computed tomography.

Authors:  Matthias Guckenberger; Markus Weininger; Juergen Wilbert; Anne Richter; Kurt Baier; Thomas Krieger; Buelent Polat; Michael Flentje
Journal:  Radiother Oncol       Date:  2007-09-12       Impact factor: 6.280

9.  Variability of four-dimensional computed tomography patient models.

Authors:  Jan-Jakob Sonke; Joos Lebesque; Marcel van Herk
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-26       Impact factor: 7.038

10.  Observation of interfractional variations in lung tumor position using respiratory gated and ungated megavoltage cone-beam computed tomography.

Authors:  Jenghwa Chang; Gig S Mageras; Ellen Yorke; Fernando De Arruda; Jussi Sillanpaa; Kenneth E Rosenzweig; Agung Hertanto; Hai Pham; Edward Seppi; Alex Pevsner; C Clifton Ling; Howard Amols
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-04-01       Impact factor: 7.038

View more
  5 in total

1.  Digital tomosynthesis for respiratory gated liver treatment: clinical feasibility for daily image guidance.

Authors:  Q Jackie Wu; Jeffrey Meyer; Jessica Fuller; Devon Godfrey; Zhiheng Wang; Junan Zhang; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-06-18       Impact factor: 7.038

Review 2.  Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy?

Authors:  Anna O Simeonova; Katharina Fleckenstein; Hansjörg Wertz; Anian Frauenfeld; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz
Journal:  Transl Lung Cancer Res       Date:  2012-03

Review 3.  Improving radiotherapy planning, delivery accuracy, and normal tissue sparing using cutting edge technologies.

Authors:  Carri K Glide-Hurst; Indrin J Chetty
Journal:  J Thorac Dis       Date:  2014-04       Impact factor: 2.895

4.  Anatomic and pathologic variability during radiotherapy for a hybrid active breath-hold gating technique.

Authors:  Carri K Glide-Hurst; Ellen Gopan; Geoffrey D Hugo
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-07-01       Impact factor: 7.038

5.  Extracting respiratory signals from thoracic cone beam CT projections.

Authors:  Hao Yan; Xiaoyu Wang; Wotao Yin; Tinsu Pan; Moiz Ahmad; Xuanqin Mou; Laura Cerviño; Xun Jia; Steve B Jiang
Journal:  Phys Med Biol       Date:  2013-02-11       Impact factor: 3.609

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.