Optimizing 4D cone-beam CT acquisition protocol for external beam radiotherapy.

Tianfang Li1, Lei Xing.   

Abstract

PURPOSE: Four-dimensional cone-beam computed tomography (4D-CBCT) imaging is sensitive to parameters such as gantry rotation speed, number of gantry rotations, X-ray pulse rate, and tube current, as well as a patient's breathing pattern. The aim of this study is to optimize the image acquisition on a patient-specific basis while minimizing the scan time and the radiation dose. METHODS AND MATERIALS: More than 60 sets of 4D-CBCT images, each with a temporal resolution of 10 phases, were acquired using multiple-gantry rotation and slow-gantry rotation techniques. The image quality was quantified with a relative root mean-square error (RE) and correlated with various acquisition settings; specifically, varying gantry rotation speed, varying both the rotation speed and the number of rotations, and varying both the rotation speed and tube current to keep the radiation exposure constant. These experiments were repeated for three different respiratory periods.
RESULTS: With similar radiation dose, 4D-CBCT images acquired with low current and low rotation speed have better quality over images obtained with high current and high rotation speed. In general, a one-rotation low-speed scan is superior to a two-rotation double-speed scan, even though they provide the same number of projections. Furthermore, it is found that the image quality behaves monotonically with the relative speed as defined by the gantry rotation speed and the patient respiratory period.
CONCLUSIONS: The RE curves established in this work can be used to predict the 4D-CBCT image quality before a scan. This allows the acquisition protocol to be optimized individually to balance the desired quality with the associated scanning time and patient radiation dose.

Entities:  

Mesh:

Year:  2006        PMID: 17197125     DOI: 10.1016/j.ijrobp.2006.10.024

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  32 in total

1.  In-treatment 4D cone-beam CT with image-based respiratory phase recognition.

Authors:  Satoshi Kida; Yoshitaka Masutani; Hideomi Yamashita; Toshikazu Imae; Taeko Matsuura; Naoya Saotome; Kuni Ohtomo; Keiichi Nakagawa; Akihiro Haga
Journal:  Radiol Phys Technol       Date:  2012-02-25

2.  Extraction of tumor motion trajectories using PICCS-4DCBCT: a validation study.

Authors:  Zhihua Qi; Guang-Hong Chen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  Iterative image reconstruction for CBCT using edge-preserving prior.

Authors:  Jing Wang; Tianfang Li; Lei Xing
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

4.  Intrafractional motion of the prostate during hypofractionated radiotherapy.

Authors:  Yaoqin Xie; David Djajaputra; Christopher R King; Sabbir Hossain; Lijun Ma; Lei Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-09-01       Impact factor: 7.038

5.  Quality and accuracy of cone beam computed tomography gated by active breathing control.

Authors:  Bria P Thompson; Geoffrey D Hugo
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

6.  Autoadaptive phase-correlated (AAPC) reconstruction for 4D CBCT.

Authors:  Frank Bergner; Timo Berkus; Markus Oelhafen; Patrik Kunz; Tinsu Pan; Marc Kachelriess
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

7.  Prior Image Constrained Compressed Sensing (PICCS).

Authors:  Guang-Hong Chen; Jie Tang; Shuai Leng
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008-03-03

8.  Target-specific optimization of four-dimensional cone beam computed tomography.

Authors:  Moiz Ahmad; Tinsu Pan
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

9.  Cine cone beam CT reconstruction using low-rank matrix factorization: algorithm and a proof-of-principle study.

Authors:  Jian-Feng Cai; Xun Jia; Hao Gao; Steve B Jiang; Zuowei Shen; Hongkai Zhao
Journal:  IEEE Trans Med Imaging       Date:  2014-04-21       Impact factor: 10.048

10.  High temporal resolution and streak-free four-dimensional cone-beam computed tomography.

Authors:  Shuai Leng; Jie Tang; Joseph Zambelli; Brian Nett; Ranjini Tolakanahalli; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2008-09-24       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.