| Literature DB >> 19033411 |
Satish K Garg1, Mary K Voelmle, Peter Gottlieb.
Abstract
OBJECTIVE: The purpose of this pilot study was to evaluate the feasibility of 10-day use of a transcutaneous, real-time, continuous glucose-monitoring (CGM) system. All previous reports using different CGM systems were for 3-, 5-, or 7-day use. RESEARCH DESIGN AND METHODS: On day 1, subjects received the CGM device (SEVEN System) and underwent training on proper use. Subjects returned to the clinic on days 2, 7, and 10 for in-clinic sessions. On days 2 and 7, half the subjects performed fingersticks every 15 min and the other half had Yellow Springs Instruments (YSI) samples drawn every 15 min. On day 10, all subjects participated in an 8-h in-clinic session with YSI and fingerstick testing.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19033411 PMCID: PMC2646024 DOI: 10.2337/dc08-1745
Source DB: PubMed Journal: Diabetes Care ISSN: 0149-5992 Impact factor: 19.112
Differences in glucose values (mg/dl) from CGM in reference to YSI and SMBG
| No. of paired data points | Absolute difference (mg/dl) | Relative difference (%) | Absolute relative difference (%) | ||||
|---|---|---|---|---|---|---|---|
| Mean ± SD | Median | Mean ± SD | Median | Mean ± SD | Median | ||
| YSI in-clinic days 2, 7, and 10 | 5.0 ± 27.3 | 1.4 | |||||
| Overall | 905 | 25.2 ± 25.4 | 17.3 | 17.9 ± 21.2 | 12.9 | ||
| Day | −0.1 ± 18.3 | −1.2 | |||||
| 2 | 271 | 23.6 ± 21.1 | 16.5 | 4.5 ± 20.5 | 1.7 | 14.4 ± 11.3 | 12.6 |
| 7 | 227 | 21.2 ± 21.0 | 16.0 | 8.7 ± 34.3 | 1.6 | 15.7 ± 14.0 | 11.3 |
| 10 | 407 | 28.4 ± 29.7 | 18.8 | 5.0 ± 27.3 | 1.4 | 21.6 ± 27.9 | 14.5 |
| | 0.08 | 0.02 | 0.63 | ||||
| SMBG in-clinic days 2, 7, and 10 | |||||||
| Overall | 1,130 | −0.9 ± 26.3 | −4.9 | 17.7 ± 19.5 | 13.3 | ||
| Day | 24.8 ± 25.3 | 17.0 | |||||
| 2 | 379 | 23.0 ± 22.5 | 16.0 | 0.2 ± 21.4 | −1.9 | 15.6 ± 14.5 | 12.8 |
| 7 | 317 | 20.9 ± 25.1 | 15.5 | −2.7 ± 21.3 | −6.4 | 15.7 ± 14.6 | 12.5 |
| 10 | 434 | 20.9 ± 25.1 | 15.5 | ||||
| | 0.03 | 0.002 | 0.002 | ||||
Data are mean ± SD or median unless otherwise indicated.
Calculated as absolute value of (sensor − YSI) where, for each paired point, sensor = time-matched continuously measured glucose value and YSI = SMBG values.
Calculated as (sensor − YSI)/YSI.
‡Calculated as the absolute value of (sensor − YSI)/YSI.
From the time of sensor insertion (in 24-h increments). P values calculated by the χ2 Kruskal-Wallis test from median values.
Percent of points within 20% of reference = 70.8%; percent of points within 30% of reference = 87.6%.
Percent of points within 20% of reference = 73.4%; percent of points within 30% of reference = 88.7%.