Abstract

BACKGROUND: A new continuous glucose monitoring system (CGMS Datalogger, Medtronic MiniMed, Northridge, CA) potentiates extended sensor use by eliminating the cable connection to a monitor and by being waterproof. We evaluated the performance, safety, and patient tolerance of using the CGMS for 7 continuous days in children with type 1 diabetes mellitus who were encouraged to participate fully in their usual sports and activities in their home environment.
METHODS: Twenty pediatric subjects (12.2 +/- 4.6 years old [mean +/- SD] and glycosylated hemoglobin of 8.06 +/- 1.22%) wore two CGMS devices simultaneously for 7 days. Sensor function was assessed by paired sensor-meter values obtained from the CGMS and their Paradigm Link (Medtronic Minimed) home glucose meter and a daily patient log of sensor and Datalogger sites.
RESULTS: Subjects were wearing 90% of the sensors at the end of 7 days. The devices were well tolerated except for pruritus at the adhesive sites in 29% of subjects, and one sensor site (3%) became infected. Once a correction was made to the connection between the cable and Datalogger, 89% of the 18 sensors that initialized were functional at the end of 5 days [r = 0.91; percent mean absolute relative difference (%MARD) = 12.4%], and 78% were functioning at the end of 7 days (r = 0.91; %MARD s 15.4%). Patient comfort while wearing the device decreased after 5 days of sensor wear.
CONCLUSIONS: This study demonstrates that the life of the glucose sensor can be extended well beyond the current labeling of 72 h. Once the cable connection was corrected, there was no statistically significant change in sensor performance over 7 days. Patients preferred to wear the device for a maximum of 5-6 days.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16734544     DOI: 10.1089/dia.2006.8.139

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  6 in total

Review 1.  A systems approach to clinical oncology uses deep phenotyping to deliver personalized care.

Authors:  James T Yurkovich; Qiang Tian; Nathan D Price; Leroy Hood
Journal:  Nat Rev Clin Oncol       Date:  2019-10-16       Impact factor: 66.675

2.  Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical.

Authors:  Irene Mamkin; Svetlana Ten; Sonal Bhandari; Neesha Ramchandani
Journal:  J Diabetes Sci Technol       Date:  2008-09

3.  Noninvasive ultrasonic glucose sensing with large pigs (approximately 200 pounds) using a lightweight cymbal transducer array and biosensors.

Authors:  Eun-Joo Park; Jacob Werner; Joshua Beebe; Samantha Chan; Nadine Barrie Smith
Journal:  J Diabetes Sci Technol       Date:  2009-05-01

4.  Cutaneous Complications With Continuous or Flash Glucose Monitoring Use: Systematic Review of Trials and Observational Studies.

Authors:  Nurul A Mohd Asarani; Andrew N Reynolds; Sara E Boucher; Martin de Bock; Benjamin J Wheeler
Journal:  J Diabetes Sci Technol       Date:  2019-08-27

5.  The n-of-1 clinical trial: the ultimate strategy for individualizing medicine?

Authors:  Elizabeth O Lillie; Bradley Patay; Joel Diamant; Brian Issell; Eric J Topol; Nicholas J Schork
Journal:  Per Med       Date:  2011-03       Impact factor: 2.512

6.  Feasibility of 10-day use of a continuous glucose-monitoring system in adults with type 1 diabetes.

Authors:  Satish K Garg; Mary K Voelmle; Peter Gottlieb
Journal:  Diabetes Care       Date:  2008-11-25       Impact factor: 19.112

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.