| Literature DB >> 19016814 |
S Kajander1, H Ukkonen, H Sipilä, M Teräs, J Knuuti.
Abstract
OBJECTIVES: To test the image quality and feasibility of a sequential low radiation dose protocol for hybrid cardiac PET/CT angiography (CTA).Entities:
Mesh:
Substances:
Year: 2008 PMID: 19016814 PMCID: PMC2701563 DOI: 10.1111/j.1475-097X.2008.00838.x
Source DB: PubMed Journal: Clin Physiol Funct Imaging ISSN: 1475-0961 Impact factor: 2.273
Figure 1Principle of prospectively triggered step-and-shoot cardiac CT. Top: Heart rate trend graph; Bottom: Patient ECG during step-and-shoot CT angiography acquisition. White segments denote the time periods with X-ray tube on. White vertical line: 75% of the R-R-Interval. Vertical lines with diamonds: ECG triggering points.
Figure 2Combined cardiac CT angiography and PET perfusion protocol.
Typical radiation doses of diagnostic cardiac imaging modalities.
| Method | Dose, mSv |
|---|---|
| 99mTc sestamibi rest-stress | 11·3 |
| 201Tl stress-reinjection | 31·4 |
| 15O-water | 2·5 |
| 82Rb | 13·5 |
| 13N-ammonia | 2·4 |
| CTA (64 MDCT | 8·4–21·4 |
| Dual Source CT | <10 |
| Coronary Angiography | 2·3–22·7 |
Einstein .
Achenbach .
Patient data and results of image quality, diagnostic confidence and radiation dose of spiral and sequential CTA (n = 20).
| PT # | Gender | Protocol: 1 = SPIR, 2 = SEQ | BMI | HR Prior to acquisition | MG Metoprolol | Nitrate (Y/N) | HR During acquisition | Best ECG phase | Overall image quality 1–5 (1 = Poor, 5 = Excellent) | Artefact type | DG Confidence 1–5 (1 = Poor 5 = Excellent) | DLP (mGy*cm) | Calc.Eff.Dose (MsV) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | F | 1 | 20·4 | 63 | 25 | Y | 61 | 80 | 4 | STAIR | 5 | 866 | 14·7 |
| 2 | F | 1 | 22·4 | 80 | 25 | Y | 55 | 80 | 5 | NONE | 5 | 1064 | 18·1 |
| 3 | F | 1 | 25·7 | 65 | 10 | Y | 50 | 75 | 5 | NONE | 5 | 1076 | 18·3 |
| 4 | F | 1 | 28·3 | 64 | 10 | Y | 67 | 80LCA 40RCA | 3 | MOTION | 4 | 1074 | 18·3 |
| 5 | M | 1 | 32·1 | 58 | 5 | Y | 63 | 80 | 3 | NOISE | 4 | 1369 | 23·3 |
| 6 | F | 1 | 43·0 | 75 | 20 | Y | 98 | 40 | 1 | NOISE | 2 | 1538 | 26·1 |
| 7 | F | 1 | 24·5 | 86 | 30 | Y | 77 | 40 | 3 | MOTION | 3 | 1053 | 17·9 |
| 8 | F | 1 | 20·4 | 59 | 5 | Y | 55 | 75 | 5 | NONE | 5 | 1005 | 17·1 |
| 9 | F | 1 | 28·7 | 64 | 15 | Y | 55 | 75 | 3 | NOISE, STAIR | 3 | 1364 | 23·2 |
| 10 | M | 1 | 23·7 | 78 | 25 | Y | 64 | 70LCA40RCA | 4 | NOISE | 4 | 961 | 16·3 |
| Mean ± SD | 26·9 ± 6·8 | 69·2 ± 9·7 | 17 ± 9·2 | 64·5 ± 14·1 | 3·6 ± 1·3 | 4·0 ± 1·1 | 1137 ± 213 | 19·3 ± 3·6 | |||||
| 11 | M | 2 | 27·6 | 68 | 15 | N | 61 | 75 | 4 | NOISE | 4 | 470 | 8·0 |
| 12 | F | 2 | 20·9 | 75 | 25 | Y | 61 | 75 | 4 | STAIR | 5 | 437 | 7·4 |
| 13 | F | 2 | 32·0 | 68 | 20 | Y | 61 | 75 | 3 | NOISE | 4 | 371 | 6·3 |
| 14 | F | 2 | 39·2 | 50 | 0 | N | 62 | 75 | 3 | NOISE | 4 | 495 | 8·4 |
| 15 | M | 2 | 26·9 | 50 | 0 | Y | 51 | 80 | 4 | NONE | 4 | 494 | 8·4 |
| 16 | M | 2 | 30·1 | 52 | 0 | Y | 52 | 75 | 4 | STAIR | 5 | 465 | 7·9 |
| 17 | M | 2 | 24·8 | 48 | 0 | Y | 44 | 80 | 5 | NONE | 5 | 491 | 8·3 |
| 18 | F | 2 | 24·1 | 64 | 15 | Y | 60 | 75 | 5 | NONE | 5 | 436 | 7·4 |
| 19 | M | 2 | 30·3 | 74 | 15 | Y | 56 | 75 | 2 | NOISE, MOTION | 3 | 472 | 8·0 |
| 20 | M | 2 | 31·1 | 73 | 30 | Y | 58 | 75 | 3 | NOISE | 4 | 366 | 6·2 |
| Mean ± SD | 28·7 ± 5·1 | 62·2 ± 11·0 | 12·0 ± 11·4 | 56·6 ± 5·9 | 3·7 ± 0·9 | 4·3 ± 0·7 | 450 ± 48 | 7·6 ± 0·8 |
Figure 3Example of sequential CT angiography with normal result. Radiation dose was 5.5 mSv.
Radiation dose (mSv) of spiral and sequential hybrid CT/PET.
| Scan Type | CTA + calcium score | 15O-water PET | combined protocol |
|---|---|---|---|
| Spiral ( | 19·9 ± 3·2 | 2·1 ± 0·1 | 22·0 ± 3·2 |
| Sequential ( | 7·6 ± 1·1 | 2·0 ± 0·2 | 9·5 ± 1·5 |
Figure 4Example of hybrid PET perfusion and CT angiography. a: CT angiography (spiral mode) shows multi-vessel disease with occluded left anterior descending artery (LAD) and poorly visualized mid right coronary artery (RCA). Moderate disease in left circumflex artery LCX. b: Hybrid display during adenosine stress demonstrates severe perfusion reduction in LAD associated region (anterior view) while perfusion was best preserved in LCX related region (posterior view). Perfusion was colour scaled so that red colour denotes 3.5 ml/min/g. Resting perfusion was normal (not shown). Total radiation dose of the hybrid study was 22.3 mSv.
Figure 5Example of hybrid PET perfusion and CT angiography. a: CT angiography (step-and-shoot mode) shows multi-vessel disease with massive coronary calcifications and severe stenoses in all major coronary arteries. b: Hybrid display during adenosine stress demonstrates quite preserved perfusion in anterolateral wall but moderate reduction in septal wall (anterior view). In RCA related region the perfusion was the most severely reduced (posterior view) indicating culprit stenosis in this vessel. Perfusion was colour scaled so that red colour denotes to 3.5 ml/min/g. Resting perfusion was normal (not shown). Total radiation dose of the hybrid study w as 11.3 mSv.