The accuracy of the Guardian RT continuous glucose monitor in children with type 1 diabetes.

.   

Abstract

OBJECTIVE: The aim of this study was to determine the accuracy of the Guardian RT system (Medtronic Minimed, Northridge, CA) in young children and adolescents with type 1 diabetes (T1D) during different scenarios of glucose levels and sensor age.
METHODS: At five clinical centers, 30 subjects between 4 and 17 years old with T1D were recruited. All subjects had a glycosylated hemoglobin level of <or=10.0% and were using an insulin pump. Subjects initially used a Guardian RT for approximately 1 week at home. Each subject was then hospitalized overnight for about 18 h in a clinical research center, during which time insulin-induced hypoglycemia occurred, along with frequently sampled glucose.
RESULTS: There were 1,511 laboratory glucose measurements paired with glucose measurements from 48 Guardian RT sensors. Overall, the median absolute difference (AD) was 21 mg/dL, and the median relative AD (RAD) was 14%, with 64% of sensor values meeting International Organization for Standardization home glucose meter criteria. The median AD was 27 mg/dL for reference glucose values<or=60 mg/dL and 25 mg/dL for reference glucose values<or=70 mg/dL. The median RAD was 19% for reference glucose values 71-120 mg/dL, 14% for reference glucose values 121-180 mg/dL, and 10% for reference glucose values>180 mg/dL.
CONCLUSIONS: The Guardian RT appears to perform as well in children with T1D as it has been reported to perform in adults with diabetes. The Guardian RT has an accuracy similar to that of other available continuous glucose monitors and can give important and useful clinical information.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18828242      PMCID: PMC2587171          DOI: 10.1089/dia.2007.0279

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  6 in total

1.  A multicenter study of the accuracy of the One Touch Ultra home glucose meter in children with type 1 diabetes.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

2.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

3.  The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes.

Authors:  Darrell M Wilson; Roy W Beck; William V Tamborlane; Mariya J Dontchev; Craig Kollman; Peter Chase; Larry A Fox; Katrina J Ruedy; Eva Tsalikian; Stuart A Weinzimer
Journal:  Diabetes Care       Date:  2007-01       Impact factor: 19.112

4.  Evaluation and comparison of 10 glucose methods and the reference method recommended in the proposed product class standard (1974).

Authors:  R B Passey; R L Gillum; J B Fuller; F M Urry; M L Giles
Journal:  Clin Chem       Date:  1977-01       Impact factor: 8.327

5.  Real-time continuous glucose monitoring in pediatric patients during and after cardiac surgery.

Authors:  Hannah G Piper; Jamin L Alexander; Avinash Shukla; Frank Pigula; John M Costello; Peter C Laussen; Tom Jaksic; Michael S D Agus
Journal:  Pediatrics       Date:  2006-09       Impact factor: 7.124

6.  Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the guardian continuous monitoring system.

Authors:  Bruce Bode; Kenneth Gross; Nancy Rikalo; Sherwyn Schwartz; Timothy Wahl; Casey Page; Todd Gross; John Mastrototaro
Journal:  Diabetes Technol Ther       Date:  2004-04       Impact factor: 6.118

  6 in total
  27 in total

Review 1.  Can technological solutions for diabetes replace islet cell function?

Authors:  Justin M Gregory; Daniel J Moore
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

Review 2.  Essential elements of the native glucoregulatory system, which, if appreciated, may help improve the function of glucose controllers in the intensive care unit setting.

Authors:  Leon DeJournett
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

3.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

4.  Evaluation of stochastic adjustment for glucose sensor bias during closed-loop insulin delivery.

Authors:  Craig Kollman; Peter Calhoun; John Lum; Werner Sauer; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2013-11-15       Impact factor: 6.118

5.  Analysis of the Accuracy and Performance of a Continuous Glucose Monitoring Sensor Prototype: An In-Silico Study Using the UVA/PADOVA Type 1 Diabetes Simulator.

Authors:  Marc D Breton; Rolf Hinzmann; Enrique Campos-Nañez; Susan Riddle; Michael Schoemaker; Guenther Schmelzeisen-Redeker
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

Review 6.  New-generation diabetes management: glucose sensor-augmented insulin pump therapy.

Authors:  Eda Cengiz; Jennifer L Sherr; Stuart A Weinzimer; William V Tamborlane
Journal:  Expert Rev Med Devices       Date:  2011-07       Impact factor: 3.166

7.  Use of continuous glucose monitoring as an outcome measure in clinical trials.

Authors:  Roy W Beck; Peter Calhoun; Craig Kollman
Journal:  Diabetes Technol Ther       Date:  2012-10       Impact factor: 6.118

8.  Practical aspects of real-time continuous glucose monitors: the experience of the Yale Children's Diabetes Program.

Authors:  Brett Ives; Kristin Sikes; Andrea Urban; Kerry Stephenson; William V Tamborlane
Journal:  Diabetes Educ       Date:  2009-11-23       Impact factor: 2.140

9.  Characterizing accuracy and precision of glucose sensors and meters.

Authors:  David Rodbard
Journal:  J Diabetes Sci Technol       Date:  2014-07-18

10.  Performance evaluation of a continuous glucose monitoring system under conditions similar to daily life.

Authors:  Stefan Pleus; Christina Schmid; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-07-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.