A multicenter study of the accuracy of the One Touch Ultra home glucose meter in children with type 1 diabetes.

.   

Abstract

Data are not readily available on the accuracy of one of the most commonly used home blood glucose meters, the One Touch Ultra (LifeScan, Milpitas, California). The purpose of this report is to provide information on the accuracy of this home glucose meter in children with type 1 diabetes. During a 24-h clinical research center stay, the accuracy of the Ultra meter was assessed in 91 children, 3-17 years old, with type 1 diabetes by comparing the Ultra glucose values with concurrent reference serum glucose values measured in a central laboratory. The Pearson correlation between the 2,068 paired Ultra and reference values was 0.97, with the median relative absolute difference being 6%. Ninety-four percent of all Ultra values (96% of venous and 84% of capillary samples) met the proposed International Organisation for Standardisation (ISO) standard for instruments used for self-monitoring of glucose when compared with venous reference values. Ninety-nine percent of values were in zones A + B of the Modified Error Grid. A high degree of accuracy was seen across the full range of glucose values. For 353 data points during an insulin-induced hypoglycemia test, the Ultra meter was found to have accuracy that was comparable to concurrently used benchmark instruments (Beckman, YSI, or i-STAT); 95% and 96% of readings from the Ultra meter and the benchmark instruments met the proposed ISO criteria, respectively. These results confirm that the One Touch Ultra meter provides accurate glucose measurements for both hypoglycemia and hyperglycemia in children with type 1 diabetes.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14709195     DOI: 10.1089/152091503322640971

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  20 in total

1.  Response to nocturnal alarms using a real-time glucose sensor.

Authors:  Bruce Buckingham; Jen Block; Jonathan Burdick; Andrea Kalajian; Craig Kollman; Michael Choy; Darrell M Wilson; Peter Chase
Journal:  Diabetes Technol Ther       Date:  2005-06       Impact factor: 6.118

2.  The effects of aerobic exercise on glucose and counterregulatory hormone concentrations in children with type 1 diabetes.

Authors:  Michael J Tansey; Eva Tsalikian; Roy W Beck; Nelly Mauras; Bruce A Buckingham; Stuart A Weinzimer; Kathleen F Janz; Craig Kollman; Dongyuan Xing; Katrina J Ruedy; Michael W Steffes; Timothy M Borland; Ravinder J Singh; William V Tamborlane
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

Review 3.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

4.  Preventing hypoglycaemia: what is the appropriate glucose alert value?

Authors:  P E Cryer
Journal:  Diabetologia       Date:  2008-11-19       Impact factor: 10.122

5.  Hypoglycaemia in a 94-year-old man without diabetes.

Authors:  Dustin James Mullens; Jay H Shubrook
Journal:  BMJ Case Rep       Date:  2014-05-20

6.  Home blood glucose prediction: clinical feasibility and validation in islet cell transplantation candidates.

Authors:  A M Albisser; D Baidal; R Alejandro; C Ricordi
Journal:  Diabetologia       Date:  2005-06-03       Impact factor: 10.122

7.  Accuracy of the modified Continuous Glucose Monitoring System (CGMS) sensor in an outpatient setting: results from a diabetes research in children network (DirecNet) study.

Authors:  Michael J Tansey; Roy W Beck; Bruce A Buckingham; Nelly Mauras; Rosanna Fiallo-Scharer; Dongyuan Xing; Craig Killman; William V Tamborlane; Katrina J Ruedy
Journal:  Diabetes Technol Ther       Date:  2005-02       Impact factor: 6.118

8.  Accuracy of newer-generation home blood glucose meters in a Diabetes Research in Children Network (DirecNet) inpatient exercise study.

Authors:  Stuart A Weinzimer; Roy W Beck; H Peter Chase; Larry A Fox; Bruce A Buckingham; William V Tamborlane; Craig Kollman; Julie Coffey; Dongyuan Xing; Katrina J Ruedy
Journal:  Diabetes Technol Ther       Date:  2005-10       Impact factor: 6.118

9.  The accuracy of the Guardian RT continuous glucose monitor in children with type 1 diabetes.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2008-08       Impact factor: 6.118

10.  Lack of accuracy of continuous glucose sensors in healthy, nondiabetic children: results of the Diabetes Research in Children Network (DirecNet) accuracy study.

Authors:  Nelly Mauras; Roy W Beck; Katrina J Ruedy; Craig Kollman; William V Tamborlane; H Peter Chase; Bruce A Buckingham; Eva Tsalikian; Stuart Weinzimer; Andrea D Booth; Dongyuan Xing
Journal:  J Pediatr       Date:  2004-06       Impact factor: 4.406

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.